K992084 · Cynosure, Inc. · GEX · Jul 12, 1999 · General, Plastic Surgery
Device Facts
Record ID
K992084
Device Name
CYNOSURE SURESCAN WITH THE CO3 ER: YAG LASER
Applicant
Cynosure, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 12, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Skin resurfacing and for the incision, excision ablation or vaporization of soft bodily tissues.
Device Story
Cynosure SureScan with CO3 Er:YAG Laser is a medical laser system comprising a power supply, water cooling system, and optical bench. It delivers laser energy to target soft bodily tissues for resurfacing, incision, excision, ablation, or vaporization. Operated by physicians in clinical settings, the device provides precise tissue interaction via Er:YAG laser technology. The system allows clinicians to perform surgical procedures across multiple specialties, including dermatology and plastic surgery, by vaporizing or cutting tissue. The device benefits patients by providing a controlled method for skin resurfacing and soft tissue surgery.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Medical laser system consisting of power supply, water cooling system, and optical bench. Utilizes Er:YAG laser technology for tissue ablation/vaporization. No specific materials or software algorithm details provided.
Indications for Use
Indicated for skin resurfacing and the incision, excision, ablation, or vaporization of soft bodily tissues in clinical specialties including dermatology, plastic surgery, urology, gastroenterology, neurosurgery, gynecology, arthroscopy, general surgery, ENT, and ophthalmology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Cynosure CO3 Er:YAG Laser
Submission Summary (Full Text)
{0}------------------------------------------------
K992084
## 510(K) Summary
| Submitter: | Cynosure, Inc.<br>10 Elizabeth drive<br>Chelmsford, MA 01824 |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | George Cho<br>Senior Vice President of Medical Technology |
| Date Summary Prepared: | June 18, 1999 |
| Device Trade name: | Cynosure SureScan with the CO3 Er:YAG Laser |
| Common Name: | Medical Laser System |
| Classification Name: | Instrument, surgical, powered, laser<br>79-GEX<br>21 CFR 878.48 |
| Equivalent Device | Cynosure CO3 Er:YAG Laser |
| Device description: | The SureScan with the CO3 Er:YAG Laser<br>consists of three interconnected sections: the<br>power supply, the water cooling system and the<br>optical bench. |
| Intended Use: | Skin resurfacing and for the incision, excision<br>ablation or vaporization of soft bodily tissues. |
| Comparison: | The SureScan with the CO3 Er:YAG Laser is<br>substantially equivalent to the Cynosure CO3<br>Er:YAG Laser in terms of treatment wavelength,<br>pulse duration, pulse energy, and biological<br>effects. |
| Nonclinical Performance Data: | None |
| Clinical Performance Data: | None |
| Conclusion: | The Cynosure SureScan with the CO3 Er:YAG<br>Laser is another safe and effective laser for skin<br>resurfacing and for the incision, excision, ablation<br>or vaporization of soft bodily tissues. |
| Additional Information: | None requested at this time. |
{1}------------------------------------------------
Public Health Service
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 17 1999
Mr. George Cho Senior Vice President Cynosure, Inc. 10 Elizabeth Drive 10 Limsford, Massachusetts 01824
K992084 K992084
Trade Name: Cynosure SureScan with the CO3 Er: YAG Laser
Trade Name: Cynosure SEV Re: Regulatory Class: GEX Product Code: II Dated: June 18, 1999 Received: June 21, 1999
Dear Mr. Cho:
1941 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 We have reviewed your Section 510(k) notification of manter and above and we have determined the devices markets in interstate commerce prior to use stated in the enclosure) to devices marketed in increasing to to devices that May 28, 1976, the enacment date of the Medical Device Finishaleral Food, Drug, and have been reclassified in accordance with the provisions of the general controls
Cosmetic Act (Act). You may, therefore, market the device, subject to the generals for Cosmetic Act (Act). You may, therefore, market the Act include requirements for
provisions of the Act. The general controls provisions of the Act include requirements for provisions of the Act. The general controls provisions of the Active, labeling, and
annual registration, listing of devices, good manufacturing practice, labeling, and annual registration, fisting of active, or active, and adulteration.
Prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III
n and the success of the chiest to such additional controls. Existing major If your device is classified (see above) into entrols. Existing major
(Premarket Approval), it may be subject to such additional controls. Existing major (Premarket Approval), it may be subject to such additions. The Code of Federal Regultions, Title 21.
regulations affecting your device can be found in the Code of Federal in regulations affecting your device can be found in the coon assumes compliance with the
Parts 800 to 895. A substantially equivalent determination assumes county System Parts 800 to 895. A substantially equirement, as set forth in the Quality System
current Good Manufacturing Practice requirement, as set forth in the Quality System current Good Manufacturing Practice requirences: Gencral regulation (21 CFR Part 820) and that,
Regulation (QS) for Medical Devices: Gencral regulation (21 CFR Parts of the Regulation (QS) for Medical Devices: Gentelar (viri tration (FDA) will verify
through periodic (QS) inspections, the Food and Drug Administration (FDA) in regulatory through periodic (QS) inspections, the Food and Drag result in regulation may result in regulatory such assumptions. Failure to comply with in Christ concerning your device in the action. In addition, FDA may publish turne and notification submission does
Federal Register. Please note: this response to your premarket notification of the Act for Federal Register. Please note: this response to your promations of the Act for
not affect any obligation you might have moder sections. Or other Federal law not affect any obligation you might have under sections 531 through 3 12 67 and 11
devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 - Mr. George Cho
This letter will allow you to begin marketing your device as described in your 510(k) This letter will anow you to begal finding of substantial equivalence of your device to a premarket notification. "The PDA missification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 If you desire speoffic acries its from diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. advertising of your do regulation entitled, "Misbranding by reference to prematiket notification" (21 CFR 807.97). Other general information on your responsibilities under the nonication (21 OF IC 0071) - 3 211-5 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page ol
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: Cynosure SurcScan with the CO3 Er:YAG Laser
Indications For Use:
The Cynosure SureScan with the CO3 Er: YAG Laser is used for skin resurfacing, The Cynosure SureScan with the COS Et. ITTS Lace. In the Costs of bodily tissues.
and for the incision, excision, ablation or vaporization of soft bodily tissues.
Typical applications include dermatology, plastic surgery, urology, Typical applications include definatology, placio surgery, and of .
gastroenterology, neurosurgery, gynecology, arthroscopy, general surgery, ENT and opthalmology.
(PLEASE DO NOT WRITE BELOW TIJIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter Use*_*
(Optional Format 1-2-96)
(Division Sign-Off)
Division of General Restorative Ducts R992084
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.