← Product Code [FPA](/productcode/FPA) · K991979

# SOF-SET MICRO QR INFUSION SET, MODELS 320 AND 321 (K991979)

_Medtronic Minimed · FPA · Sep 3, 1999 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K991979

## Device Facts

- **Applicant:** Medtronic Minimed
- **Product Code:** [FPA](/productcode/FPA.md)
- **Decision Date:** Sep 3, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The MiniMed Sof-set Micro QR is intended for the subcutaneous infusion of medicine, including insulin, from an external infusion pump. The set is not intended nor indicated for use with blood or blood products.

## Device Story

Infusion administration set for subcutaneous delivery of insulin/medication from external infusion pump; connects proximally to medication reservoir via female Luer connector; distal end features indwelling 24-gauge FEP catheter inserted via 26-gauge 304 stainless steel introducer needle; tubing composed of Polyfin (PVC with polyolefin liner); hub includes winged configuration with adhesive patch for stability; adhesive dressing secures catheter/line to user; used in clinical or home settings by pump users; facilitates continuous medication delivery; modifications from predicate include shorter catheter (6mm vs 9mm), shortened introducer needle, and enlarged adhesive patch with modified wing radius.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Indwelling 24-gauge FEP catheter; 26-gauge 304 stainless steel introducer needle; Polyfin (PVC with polyolefin liner) tubing; female Luer connector; winged hub with adhesive patch; manual insertion mechanism.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- MiniMed Sof-set Ultimate™ QR infusion sets, models 315 and 316

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for MiniMed. Above the name MiniMed is the text K991979. To the left of the name MiniMed is a black square with white lines inside.

## SEP 3 1999

## Part C. 510(k) Summary

Submitter: MiniMed® Inc. 12744 San Fernando Rd., Sylmar, California 91342

Contact: Don Selvey, Regulatory Affairs (818) 362-5958, 3011; (480) 704-8070 (v/f)

Name of Device: MiniMed Sof-set Micro QR® infusion sets, models 320 and 321

Predicate Device: MiniMed Sof-set Ultimate™ QR infusion sets, models 315 and 316

Description of the New Device: The MiniMed Sof-set Micro QR infusion sets, models 320 and 321. are infusion administration sets, connected to a medication reservoir proximally and inserted in the subcutaneous tissue of a user distally by means of an introducer needle. The reservoir to which the infusion set attaches proximally is inserted into an external infusion pump, such as a MiniMed infusion pump.

The administration set attaches to the reservoir by means of a female Luer connector, and subcutaneously in the user through an indwelling catheter made of Fluorinated Ethylene Propylene (FEP). The tubing is made of polyvinyl chloride (PVC) with a polyolefin liner. This configuration of PVC and polyolefin has been trademarked by MiniMed as Polyfin®.

The 24 gauge indwelling catheter is introduced into the subcutaneous tissue by a removable 26 gauge introducer needle made of 304 stainless steel. The indwelling catheter and tubing share a common hub through which the introducer needle fits. The hub incorporates a winged configuration with an adhesive patch to facilitate handling of the administration set during insertion and stability following insertion. An adhesive dressing covers the wings and hub of the administration set, securing the subcutaneous catheter and infusion line to the user.

Intended Use of the New Device: The MiniMed Sof-set Micro QR is intended for the subcutaneous infusion of medicine, including insulin, from an external infusion pump. The set is not intended nor indicated for use with blood or blood products.

Comparison of the Technological Features of the New Device and Predicate Device: The modified device is substantially similar to the lawfully marketed predicate device. Both are intended for subcutaneous delivery of insulin or other appropriately labeled medication from an infusion pump to the pump user. Both sets attach proximally to a reservoir by means of a Luer connector, and insert distally into the device user's subcutaneous tissue with a removable stainless steel needle and flexible catheter. The materials and manufacturing processes are the same for the modified and predicate devices.

Image /page/0/Picture/14 description: The image shows the number 000004 in a bold, sans-serif font. The numbers are black against a white background. The numbers are evenly spaced and aligned horizontally.

<sup>@</sup>MiniMed, Polyfin, Sof-set, and QR are registered trademarks of MiniMed Inc. TMUltimate and Micro are Trademarks of MiniMed Inc.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for MiniMed. The logo consists of a black square with white curved lines inside, followed by the word "MiniMed" in a bold, serif font. There is a trademark symbol to the upper right of the word "Med".

Several differences exist between the new and predicate device:

- 1) The new device has a 6 mm long catheter instead of the current 9 mm version.
- 2) The introducer needle has been shortened by 3 mm.
- 3) The adhesive patch beneath the stabilizing wings has been increased in size to cover the entire underside of the wings.
- 4) The underside of the wing has been modified with a smoother radius to facilitate attachment of the larger adhesive patch.

These modifications do not affect the safety or effectiveness of the device.

Signed,

Dan Solvou 6-9-99
date

Don Selvey date Senior Regulatory Affairs Specialist Department of Clinical and Regulatory Affairs MiniMed Inc.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 3 1999

Mr. Don Selvey Director, Regulatory Affairs Specialist Department of Clinical and Regulatory Affairs MiniMed Technologies, Incorporated 12744 San Fernando Road Sylmar, California 91342

K991979 Re : Trade Name: Sof-Set Mirco QR Infusion Sets, Models 320 and 321 Requlatory Class: II Product Code: FPA Dated: June 9, 1999 Received: June 14, 1999

Dear Mr. Selvey:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of

{3}------------------------------------------------

Page 2 - Mr. Selvey

the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/16 description: The image shows the logo for MiniMed. The logo consists of a black square on the left with white wavy lines inside. To the right of the square is the word "MiniMed" in a serif font, with a registered trademark symbol above the "d".

## INDICATIONS FOR USE

Ka991979

510(k) Number:

Device Name:

MiniMed Sof-set Micro QR infusion sets, models 320 and 321.

Indications for Use:

The MiniMed Sof-set Micro QR infusion set is indicated for the subcutaneous infusion of medicine, including insulin, from an external infusion pump.

Concurrence of CIDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number K991979

Over-the Counter Use

**Prescription Use** √
(Per 21 CFR 801.109)

ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ

01.

Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________

0000008

---

**Source:** [https://fda-staging.innolitics.com/device/K991979](https://fda-staging.innolitics.com/device/K991979)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).