PURILENS SYSTEM FOR CLEANING AND DISINFECTING CONTACT LENSES
K991965 · Purilens, Inc. · LYL · Oct 1, 1999 · Ophthalmic
Device Facts
Record ID
K991965
Device Name
PURILENS SYSTEM FOR CLEANING AND DISINFECTING CONTACT LENSES
Applicant
Purilens, Inc.
Product Code
LYL · Ophthalmic
Decision Date
Oct 1, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Purilense System is indicated for the cleaning and disinfection of soft (hydrophilic) contact lenses by means of subsonic agitation to remove lens deposits and microorganisms, and ultraviolet irradiation of the surrounding storage solution for disinfection. The lenses may be stored in the disinfected solution for a period of 24 hours following the cleaning and disinfection cycle.
Device Story
Purilens System cleans/disinfects soft contact lenses using mechanical and radiation-based processes. Device consists of unit with upper housing (germicidal UV light) and lower housing (cleaning chamber with lens holder). User places lenses in basket; fills chamber with preservative-free saline. Electronic control activates magnetic pedal in chamber, oscillating at 50-120 cycles/second to create subsonic turbulence/whirlpool effect; this removes debris/microorganisms from lens surfaces into saline. Simultaneously, UV light disinfects solution. 15-minute cycle ends automatically. Electrical interlock prevents operation if housing is open. Healthcare provider/patient uses device to eliminate need for digital rubbing or chemical disinfectants. Benefits include effective cleaning/disinfection without chemical preservatives, reducing potential for ocular irritation.
Clinical Evidence
Crossover clinical study compared Purilens System to multipurpose solution control over 3 months. Results indicated Purilens performed equally to control. Pre-clinical testing included toxicology (MEM Elution, systemic toxicity, ocular irritation), lens compatibility (180 cycles on Groups I-IV lenses), and microbiological efficacy (D-value, multi-item challenge, biofilm study). Computer image processing quantification of light reflections confirmed successful removal of proteins, lipids, and mucins.
Technological Characteristics
System includes cleaning unit and preservative-free saline. Features magnetic pedal agitator (50-120 Hz) and germicidal UV light source. Electrical: 110 VAC to 14 VAC transformer. Safety: electrical interlock. Materials: plastic components tested for biocompatibility. Cleaning mechanism: subsonic turbulence. Disinfection: UV irradiation. Standalone device.
Indications for Use
Indicated for cleaning and disinfection of soft (hydrophilic) contact lenses. Removes lens deposits and microorganisms via subsonic agitation and UV irradiation. Allows 24-hour storage in disinfected solution.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo for "Purilens COMPLETE CONTACT LENS CARE". The logo features the word "Purilens" in a bold, sans-serif font, with the words "COMPLETE CONTACT LENS CARE" in a smaller font below it. Above the word "Purilens" is a graphic of a circle.
# PuriLens, Inc.
12170 Race Track Rd. Tampa, FL 33626
K991965
510(k) Summary
Tel: 813 814-4141 Fax: 813 814-4545
#### SUBMITTER:
Submitted on behalf of:
Company Name: Address:
Phone: Fax:
Purilens, Inc.. 12170 Race Track Road Tampa, Florida 33626 (813) 814 4141 (813) 814 4545
CONTACT PERSON:
Richard E. Lippman, O.D., F.A.A.O. Official Representative and Correspondent C.L. McIntosh & Associates, Inc. 12300 Twinbrook Parkway Suite 625 Rockville, MD 20852
### DATE SUMMARY PREPARED: August 26, 1999
| TRADE NAME: | PurilensR System |
|--------------|-----------------------------------------------|
| COMMON NAME: | contact lens cleaning and disinfection device |
### DESCRIPTION of the DEVICE:
The device is a cleaning and disinfection unit for contact lenses. It consists of the Purilens Unit and Purilens Solution. The Purilens Unit is compact, with an upper housing containing a germicidal UV light, and a lower housing which contains a removable, clear cleaning chamber with a lens holder for the contact lenses. A mechanism in the lower portion of the cleaning chamber creates the cleaning turbulence activity. Within the lower housing unit is an electronic control unit which operates the UV light and turbulence vibrating mechanism. The lower unit contains an internal automatic timing device which shuts the unit after its 15 minute cycle.
The lens holder consists of a removable basket which houses the contact lenses in separate chambers and is suspended within the cleaning chamber in the lower housing. The removable basket assembly is shielded from the UV light. The cleaning chamber contains an unpreserved saline medium as a storage solution for the contact lenses.
{1}------------------------------------------------
The Purilens Solution is a preservative free saline solution which is used to fill the cleaning chamber. The turbulence mechanism is a magnetic pedal suspended on the end of a flexible spring located in the bottom of the cleaning chamber. When activated by the electronic connection, the pedal osillates in a subsonic range of 50 - 120 cycles per second creating a whirlpool effect that swirls the Purilens Solution surrounding the lens baskets and removes debris and microorganisms from the lenses during the 15 minute cycle.
UV radiation energy from the upper housing disinfects loosened microorganisms in the solution. Disinfection is accomplished without heating, preservatives, or chemicals. The unit cannot function unless the electrical interlock is complete between the upper and lower housings.
An electrical transformer is located on the power cord. The transformer converts house current of 110 VAC to 14 VAC for the purpose of operating the device.
## INDICATIONS FOR USE:
The Purilense System is indicated for the cleaning and disinfection of soft (hydrophilic) contact lenses by means of subsonic agitation to remove lens deposits and microorganisms, and ultraviolet irradiation of the surrounding storage solution for disinfection. The lenses may be stored in the disinfected solution for a period of 24 hours following the cleaning and disinfection cycle.
### SUBSTANTIAL EQUIVALENCE:
The Purilense System for Cleaning and Disinfecting Contact Lenses is substantially equivalent to multipurpose contact lens solution accessory products (ReNu Multipurpose Solution P860023/S6, approved January 24, 1994; and similarly approved multipurpose products) which include ingredients for cleaning and disinfecting contact lenses.
The Purilens System is comprised of a two step system. The cleaning step is accomplished by subsonic turbulence created by an agitator wand which is activated by a magnet in close proximity to the wand. This physical agitation of lens solution which aids in removal of lens deposits and microorganisms is substantially equivalent to contact lens ultrasonic cleaners and hand-driven mechanical agitation units (The Lens Comfort, Inc. Ultrasonic Contact Lens Care Accessory, K962112, cleared November 29, 1996; the Sola/Barnes -Hind Soft Mate Automatic Cleaning Unit, K852386, cleared 9/4/85; and the Clensatron 700CL. K884414 cleared 12/29/88). The turbulence created by this action loosens lens deposits and microorganisms from the lens surfaces and places those deposits and microorganisms into the surrounding storage solution, a non-preserved sterile saline. The disinfection step is
{2}------------------------------------------------
accomplished by action of ultraviolet irradiation. This activity is created by the presence of an ultraviolet light source in the device's upper housing. As the UV irradiation energy is emitted, it acts on the loosened microorganisms to reduce the organisms viability.
This device is substantially equivalent to ReNu Multipurpose Solution by Bausch & Lomb, approved under Premarket Approval P860023/S1 on July 28, 1988, and other multipurpose soft contact lens solutions with similar indications. ReNu Multipurpose Solution includes in its formulation, ingredients which clean contact lenses with the aid of digital rubbing, and other ingredients which serve as disinfecting agents when lenses are stored for a minimum of four hours or longer.
The equivalence determination is made by the comparison of actions of the relative devices within their indicated uses. The principal difference between the Purilens® System and ReNu Multipurpose Solution for cleaning is the cleaning step for the Purilens® System does not require digital rubbing, but acts by shear forces from subsonic action of the saline solution surrounding the lenses, created by a vibrating agitator. The agitator wand vibrates with sufficient intensity to remove deposits and microorganisms. The cleaning action is equivalent to a series of mechanical cleaning products including The Lens Comfort Contact Lens Care Accessory: the Sola/Barnes-Hind Soft Mate Automatic Cleaning Unit, and the Clensatron, all of which clean lenses by mechanical agitation, either by ultrasonic wave form action of the storage solution, or by mechanical agitation of storage solution in a saline held chamber.
The rinse process for the devices are similar in that the Purilense System rinse is accomplished by the continued agitation of the solution during the complete cycle process of the device. This action acts as a continued rinse as the lens is continually "washed" by the action of the subsonic vibration of the storage medium (saline). The equivalent action of rinsing "washing" lenses occurs with the comparable products using mechanical action including the aforementioned devices. This rinse step is further enhanced by a steady stream of saline rinse prior to setting the lenses in the unit to complete the agitation process.
The disinfection step for the Purilens System is accomplished by ultraviolet irradiation from the UV bulb housed in the upper housing of the device. The energy emitted by the UV bulb acts on the loosened microorganisms as they are circulating about in the saline storage solution in the carrying case chamber due to continued vibrating subsonic action on the solution medium. The effectiveness of disinfection is complete after 15 minutes in that no microorganisms are recovered after the disinfection cycle is complete on the lenses or in the chamber for up to 24 hours after the cycle is completed. The disinfection step for ReNu Multipurpose Solution is a minimum four hour soak after the lenses are stored in a carrying case with the solution added as a storage medium.
{3}------------------------------------------------
#### COMPATIBILITY OF THE DEVICE:
A series of pre-clinical studies were performed on the Purilens® System device in which toxicology studies were performed on the plastic components of the device and determined to be non-toxic and safe for use with contact lenses. The tests conducted included an evaluation of toxicity of residual solution after cycling; a toxicity evaluation by the MEM Elution Method; systemic toxicity in mice; ocular irritation study of extracts in rabbitrs. A cytotoxicity study was also conducted on the plastic coated agitator found at the base of the portable lens carrier. Solution compatibility studies were performed after 180 cycles of the device on Groups I, III, III, and IV soft contact lenses to determine the stability of lens parameters after cycling and found to have no negative affect on the lens stability or compatibility with the storage solution. Microbiological studies were performed on the lenses, storage solution and lens case to determine whether the device performed the disinfection phase of the cycle, and were determined to have performed all tests successfully. Specific tests which were performed included: disinfection efficacy or "D" value test; multi-item challenge test; evaluation of pH and temperature of the Purilens System chamber solution after cycle completion: and an evaluation of microbial accumulation within the unit (biofilm study).
A lens cleaning analysis using computer image processing quantification of visible light reflections of hydrogel contact lens deposits was performed on all lenses worn in the clinical study. The study documented that the lenses were successfully cleaned and free of deposits including proteins, lipids, and mucins.
#### CLINICAL ANALYSIS
A clinical study was performed with the device and a control with a comparable multipurpose contact lens cleaning and disinfection system. A crossover design study was employed in which all subjects experienced both the Purilens® System and the multipurpose disinfection control for three months each. The results of the clinical study indicate that the Purilens® System performs equally as a multipurpose contact lens solution for the same indication. The effect of the Purilense System indicates that the performance is superior to a comparable multipurpose cleaning and disinfection system.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 1999
Purilens. Inc. c/o Richard E. Lippman, O.D., F.A.A.O. C.L. McIntosh & Associates, Inc. 12300 Twinbrook Parkway, Suite 625 Rockville, MD 20852
Re: K991965
. ……
Trade Name: Purilens® System (consisting of the Purilens Unit and Purilens Solution, a preservative free saline)
Regulatory Class: II Product Codes: LYL and LPN Dated: August 26, 1999 Received: August 27, 1999
Dear Dr. Lippman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Alvarez L. Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications Statement
510(k) Number (if known) K991965_
Device Name: _ Purilens® System
. .
#### Indications for Use: (Revised: 8/26/99)
The Purilense System is indicated for the cleaning and disinfection of soft (hydrophilic) contact lenses by means of subsonic agitation to remove lens deposits and microorganisms, and ultraviolet irradiation of the surrounding storage solution for disinfection. The lenses may be stored in the disinfected solution for a period of 24 hours following the cleaning and disinfection cycle.
Additional Claims:
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Presription Use
OR
Over-The-Counter Use
(Oprtional Format 1-2-96)
Karin Warburton
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K991965
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.