# MODIFICATION TO ENGINEERED MEDICAL SYSTEMS DISPOSABLE MANUAL RESUSCITATOR WITH CO2 DETECTION (K991953)

_Engineered Medical Systems · BTM · Jul 1, 1999 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K991953

## Device Facts

- **Applicant:** Engineered Medical Systems
- **Product Code:** [BTM](/productcode/BTM.md)
- **Decision Date:** Jul 1, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5915
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

Disposable manual resuscitator with CO2 detection for use with patients requiring manual ventilatory support. The CO2 detector is located on the expiratory port. It may be supplied separately and attached to any resuscitator expiratory port. The CO2 detector assists in verification of the endotracheal tube placing during endotracheal or nasotracheal intubation. The CO2 detector detects approximate ranges of CO2 by color comparison. It is intended for use by qualified healthcare professionals in any environment where pulmonary support resuscitation is indicated such as in the hospital, transport, mobile or home settings.

## Device Story

Device is a bag-valve-mask manual resuscitator with integrated CO2 detection. Input is exhaled gas from patient airway; CO2 detector on expiratory port uses colorimetric chemical indicator to estimate CO2 ranges. Output is visual color change indicating CO2 presence/level. Used by qualified healthcare professionals in hospital, transport, mobile, or home settings. Assists clinicians in verifying endotracheal tube placement during intubation and monitoring ventilatory support. Benefits patient by providing immediate feedback on airway placement and ventilation efficacy.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on design, material, and functional comparison to legally marketed predicate devices.

## Technological Characteristics

Manual bag-valve-mask resuscitator with integrated colorimetric CO2 detector. CO2 detection via chemical color comparison media. Single-patient use. Provided clean, non-sterile. No electronic components or software algorithms.

## Regulatory Identification

A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.

## Predicate Devices

- EMS Disposable Manual Resuscitator (K912203B, K924610A)
- Nellcor Puritan Bennett - DMR2 Plus Manual Resuscitator with CO2 Detection ([K973419](/device/K973419.md))
- Nellcor Puritan Bennett - EasyCap CO2 Detector ([K894053](/device/K894053.md), [K944400](/device/K944400.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 1953

| EMS Engineered Medical Systems                                                      |                                                                                         | Attribute                                                                        | EMS DMR<br>with CO₂<br>(Modification) | EMS DMR<br>K912203B<br>K924610A | NPB DMR<br>Plus with CO₂<br>K973419 |
|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|---------------------------------------|---------------------------------|-------------------------------------|
| 2055 EXECUTIVE DRIVE • INDIANAPOLIS, IN 46241 • (317) 246-5500 • FAX (317) 246-5501 |                                                                                         | Use                                                                              |                                       |                                 |                                     |
| Non-Confidential Summary of Safety and Effectiveness                                |                                                                                         | Intended to provide manual<br>ventilatory support                                | Yes                                   | Yes                             | Yes                                 |
| page 1 of 3                                                                         |                                                                                         | Intended to assist in verification of<br>tube placement by expired CO₂ detection | Yes                                   | --                              | Yes                                 |
| May 28, 1999                                                                        |                                                                                         | Used in hospitals, home, transport<br>mobile settings                            | Yes                                   | Yes                             | Yes                                 |
| Engineered Medical Systems, Inc.<br>2055 Executive Dr.<br>Indianapolis, IN 46241    | Tel - (317) 246-5500                                                                    | Single Patient Use                                                               | Yes                                   | Yes                             | Yes                                 |
|                                                                                     | Fax - (317) 246-5501                                                                    | Design                                                                           |                                       |                                 |                                     |
| Official Contact:                                                                   | Bonnie Holly - Quality Manager                                                          | CO₂ detector placed on expiratory<br>port of manual resuscitator                 | Yes                                   | --                              | Yes                                 |
| Proprietary or Trade Name:                                                          | EMS Disposable Manual Resuscitator with CO2 Detection                                   | CO₂ detection by color comparison                                                | Yes                                   | --                              | Yes                                 |
| Common/Usual Name:                                                                  | Disposable Manual Resuscitator with CO₂ Detection                                       | CO₂ detector good for up to 2 hours                                              | Yes                                   | --                              | Yes                                 |
| Classification Name:                                                                | Emergency Manual Ventilator Resuscitator                                                | Can be replaced if needed                                                        | Yes                                   | --                              | Yes                                 |
| Device:                                                                             | EMS Disposable Manual Resuscitator with CO₂ Detection                                   | Materials                                                                        |                                       |                                 |                                     |
| Predicate Devices:                                                                  | EMS Disposable Manual Resuscitator with CO₂ Detection -<br>K912203B and K924610A        | Materials in CO₂ detection media the<br>exactly the same as predicate            | Yes                                   | --                              | Yes                                 |
|                                                                                     | Nellcor Puritan Bennett - DMR2 Plus Manual Resuscitator with CO2<br>Detection - K973419 | All other materials exactly the<br>same as predicate                             | Yes                                   | Yes                             | --                                  |
|                                                                                     | Nellcor Puritan Bennett - EasyCap CO₂ Detector - K894053 and<br>K944400                 | Packaging                                                                        |                                       |                                 |                                     |
| Provided clean, non-sterile                                                         | Yes                                                                                     | Yes                                                                              | Yes                                   |                                 |                                     |
| Performance Standards / Specifications                                              |                                                                                         |                                                                                  |                                       |                                 |                                     |
| None applicable under Section 514                                                   | Yes                                                                                     | Yes                                                                              | Yes                                   |                                 |                                     |

Device Description - Research - Comments - Research - Research - Mar - Mar - Mar -

The EMS Disposable Manual Resuscitator is a bag-valve-mask device with the capability of delivering supplemental oxygen. It incorporates a feature CO2 detection.

#### Intended Use of the may be the state of the county of the county of the first of the first

Intended for use where pulmonary support resuscitation is indicated and Indicated Use -exhaled CO2 detection is desirable.

Environment of Use --Hospital, Transport, Mobile and Home settings

{1}------------------------------------------------

# Non-Confidential Summary of Safety and Effectiveness

page 2 of 3 May 28, 1999

# Comparison to Predicate Devices:

{2}------------------------------------------------

## Non-Confidential Summary of Safety and Effectiveness

page 3 of 3 May 28, 1999

# Differences Between Other Legally Marketed Predicate Devices | | | | | | | | | | | | | | | | | | | | | | | | | | | | |

There are no differences between the proposed modification, that of adding CO2 detection capabilities to the exhalation port of the manual resuscitator, and the predicates.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which consists of three stylized human figures intertwined, representing health and well-being. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL - 1 1999

Ms. Bonnie Holly Engineered Medical Systems 2055 Executive Drive Indianapolis, IN 46241

Re: K991953 Disposable Manual Resuscitator with CO2 Detection Requlatory Class: II (two) Product Code: 73 BTM Dated: May 28, 1999 June 1, 1999 Received:

Dear Ms. Holly:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Ms. Bonnie Holly

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callehan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### ATIONS FOR USE

Pursuant to the Notice of February 6, 1996 regarding listing of Indications for Use on a separate sheet, the following is per that request.

| 510(k) Number: | (To be assigned)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:   | EMS Disposable Resuscitator with CO2 Detector                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Intended Use : | Disposable manual resuscitator with CO2 detection for use with patients requiring manual ventilatory support. The CO2 detector is located on the expiratory port. It may be supplied separately and attached to any resuscitator expiratory port. The CO2 detector assists in verification of the endotracheal tube placing during endotracheal or nasotracheal intubation. The CO2 detector detects approximate ranges of CO2 by color comparison. It is intended for use by qualified healthcare professionals in any environment where pulmonary support resuscitation is indicated such as in the hospital, transport, mobile or home settings. |

Concurrence of CDRH, Office of Device Evaluation (ODE)

Atl. A. Cirlouski.

(Division Sign-Off) Division of Cardiovascu and Neurological Devic 510(k) Number ..

**Prescription Use**
(Per CFR 801.109)

or

Over-the-counter use

---

**Source:** [https://fda-staging.innolitics.com/device/K991953](https://fda-staging.innolitics.com/device/K991953)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).