Ceramic abutments are indicated for use with endosseous dental implants, to extend the implant's coronal aspect through the mucosal tissues and into the oral cavity, for prosthetic attachment. Ceramic abutments are primarily indicated for use in the higher aesthetically required anterior regions of the oral cavity but are not restricted for that application.
Device Story
3i Ceramic Abutment is a dental prosthetic component used to connect endosseous dental implants to oral prosthetics. It consists of a zirconium ceramic outer sleeve bonded to a machined titanium alloy core using an apatite glass-ceramic seal. The device is placed by a dental professional in a clinical setting. It functions by extending the implant's coronal aspect through mucosal tissues into the oral cavity, providing an aesthetic base for prosthetic attachment. The device is intended for use in the anterior region of the oral cavity. Clinical benefit includes improved aesthetic outcomes for dental restorations. The healthcare provider uses the abutment to support the final prosthesis, directly impacting the functional and aesthetic success of the dental implant procedure.
Clinical Evidence
Bench testing only. Testing included cyclical fatigue at 30° angle (5,000,000 cycles at 300 N), static load at 30° angle (695 N), and torsional strength (166 Ncm). Results showed the 3i ceramic abutment performed as well as or better than equivalent devices.
Technological Characteristics
Three-component assembly: zirconium ceramic outer sleeve, machined titanium alloy core, and apatite glass-ceramic bond. Sterilized via gamma irradiation to SAL 10^-6. Dimensions/form factor designed for endosseous dental implant interface. No software or electronic components.
Indications for Use
Indicated for patients requiring prosthetic attachment to endosseous dental implants to extend the coronal aspect through mucosal tissues into the oral cavity. Primarily for anterior regions requiring high aesthetics. Contraindicated in patients with insufficient jawbone width/height, lack of osseointegration, poor oral hygiene, heavy smoking/tobacco use, blood disorders, infection, vascular impairment, uncontrolled diabetes, drug/alcohol abuse, chronic high-dose steroid therapy, or metabolic bone disease.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Nobel Biocare CerAdapt
Submission Summary (Full Text)
{0}------------------------------------------------
K991947
#### 20
### 510(k) SUMMARY 3i Ceramic Abutments
To 510(k) Summary Requestor:
Information contained in this summary is from an original Pre-Market Notification [510(k)] submission, provided the United States Food and Drug Administration. No pertinent or known releasable information regarding device safety or efficacy has been intentionally or otherwise knowingly deleted from that submission, for this summary.
William G. Conety
Regulatory Affairs
CLASSIFICATION NAME: Endosseous Dental Implant
COMMON/USUAL NAMES: Abutments; transmucosal abutments/elements
PROPRIETARY NAME: No established marketing name at time of submission.
CLASSIFICATION: Class III
PERFORMANCE STANDARDS: No established performance standards for abutments
FORM and CONSTRUCTION: The 3i ceramic abutment is comprised of three basic components: An outer cylinder or sleeve constructed from zirconium ceramic bonded to an inner machined titanium alloy core with a biocompatible, water-insoluble, apatite glass-ceranic seall.
Titanium alloy cores or inserts are machined. The ceranic sleeve is molded. The two components are bonded using proprietary process. After bonding, parts are cleaned, inspected, packaged and sterilized. Sterilization is accomplished by gamma irradiation to a SAL of 10 ° The 31 ceramic abutment is equivalent in physical strength characteristics as other devices.
{1}------------------------------------------------
Cyclical Fatigue at 30° Angle (Average result)
| Other device | 3i |
|-------------------------------------------|------------------------|
| 5000000 cycles @ 250 N | 5000000 cycles @ 300 N |
| Static load at 30° Angle (Average result) | |
| Other device | 3i |
| 517 N | 695 N |
| 03 Torsional Strength (Average result) | |
| Other device | 3i |
| 79.5 Ncm (¹) | 166 Ncm (²) |
| ¹ Abutment hex failed | ² Implant hex failed |
In all physical testing undertaken, the 3i ceramic abutment performed as well or better than equivalent other devices.
SUBSTANTIAL EQUIVALENCE: The 3i ceramic abutment is substantially equivalent to Nobel Biocare "CerAdapt" abutment in that they are similar in outward appearance, basic design configuration and indications for use do not differ between the two.
The Nobel Biocare device is a one-piece aluminum oxide ceramic device. The 3i ceramic abutment is a three component device: Inner titanium sleeve and outer ceramic body, bonded by a compatible ceramic.
INDICATIONS: Ceramic abutments are indicated for use with endosseous dental implants, to extend implant's coronal aspect through mucosal tissues and into the oral cavity, for prosthetic attachment. Ceramic abutments are primarily indicated for use in higher aesthetically required anterior regions of the oral cavity but are not restricted for that application.
CONTRAINDICATIONS: No known (reported) or anticipated contraindications in use of ceramic abutments when used with implants that have achieved integration in alveolar bone. 3i abutments are not indicated for use in locations other than alveolar ridges.
Implants/abutments should not be used in cases where remaining jawbone is too diminished to provide adequate width/height to surround implant. Lack of osseointegration or subsequent implant failure may occur in cases where there is insufficient available bone or poor bone quality, poor oral hygiene, heavy smoking or tobacco use or medical conditions such as blood disorders, infection(s), vascular impairment at surgical site, uncontrolled diabetes, heavy
{2}------------------------------------------------
smoking or tobacco abuse, drug or alcohol abuse, chronic high dose steroid therapy, medical conditions such as blood clotting disorders, current or ongoing anticoagulant therapy, metabolic bone disease or other metabolic or systemic disorders which may adversely affect bone or wound healing or cases in which the available bone is too diminished to provide adequate width or height to adequately hold implants and restorative appliances.
WARNINGS: For safe and effective use of 3i implants and abutments, it is strongly suggested that specialized training be undertaken since the surgical techniques required to place dental implants are highly specialized, complex procedures. Improper patient selection and technique can cause implant and/or abutment failure with possible loss of supporting bone. 3i implants are not indicated for use in sites other than alveolar ridges.
PRECAUTIONS: Thorough screening of prospective implant candidates must be performed. Visual inspection as well as panoramic and periapical radiographs are essential to determine anatomical landmarks, occlusal conditions, periodontal status, and adequacy of bone. Lateral cephalometric radiographs, CT scans and tomogram may also be beneficial.
ADVERSE EFFECTS: Ceramic abutments: No known (reported) or anticipated adverse effects.
Loss of implant anchorage (failure to integrate) and loss of the prosthesis are possible occurrences after surgery. Lack of quantity or quality of remaining bone, infections, poor patient oral hygiene or cooperation, and generalized diseases (diabetes, etc.) are some potential causes for loss of anchorage.
SURGICAL COMPLICATIONS: Ceramic abutments: No known (reported) or anticipated surgical complications. The implant procedure has risks, including localized swelling, dehiscence, tenderness of short duration, edema, hematoma, or bleeding. Numbness of the lower lip and chin region following lower jaw surgery, and of the tissue beside the nose following upper jaw surgery, is a possible side effect of the surgery. Though it would most probably be of a temporary nature, in very rare cases, the numbness has been permanent. Gingival/mucosal (gum tissue) ulceration, tissue reaction, or infection may occur, but generally responds to local care.
Summary _ End Summary _ _ _ _ _ _ _ _
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is a stylized image of an eagle with three lines representing its wings.
Public Hcalth Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 9 1999
Mr. William G. Conety Implant Innovations, Incorporated 4555 Riverside Drive Palm Beach Gardens, Florida 33410
Re : K991947 Trade Name: 3I Ceramic Abutment System Regulatory Class: III Product Code: DZE September 7, 1999 Dated: Received: September 8, 1999
Dear Mr. Conety:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, listi prohibitions against misbranding practice, fabering, and
If your device is classified (see above) into either class II (Special Controls) or class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of
{4}------------------------------------------------
Page 2 - Mr. Conety
the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4690. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdaj.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 510(k) Number: K
Page 1 of 1
# Device Name: Ceramic Abutment System
### INDICATIONS FOR USE:
Ceramic abutments are indicated for use with endosseous dental implants, to extend the implant's coronal aspect through the mucosal tissues and into the oral cavity, for prosthetic attachment. Ceramic abutments are primarily indicated for use in the higher aesthetically required anterior regions of the oral cavity but are not restricted for that application.
# DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Russo
Division Sign-Off) Division of Dental, Infection Control, and General Hospital D 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.