K991943 · Philips Medical Systems (Cleveland), Inc. · LNH · Aug 23, 1999 · Radiology
Device Facts
Record ID
K991943
Device Name
MRGP OPTICAL TRACKING UNIT
Applicant
Philips Medical Systems (Cleveland), Inc.
Product Code
LNH · Radiology
Decision Date
Aug 23, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Outlook system is indicated for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis. The MRGP Optical Tracking Unit is designed to support and guide diagnostic and therapeutic interventional procedures, such as needle biopsies and drainages.
Device Story
MRGP Optical Tracking Unit functions as an accessory to MRI systems to assist in interventional procedures. Input consists of infrared signals from a Position Sensor Assembly (PSA) reflected off markers on specialized tools. System transforms these spatial inputs into guidance for imaging planes; software reformats 3D image sets and provides graphical overlays for planning and intraoperative use. Used in clinical settings by physicians to shorten time required to identify correct imaging trajectories, particularly for complex double oblique planes. Benefits include improved accuracy and efficiency during needle biopsies and drainages.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and comparison of technological characteristics and performance specifications.
Technological Characteristics
System includes infrared digitizer hardware, specialized tracking tools, and software for image plane guidance. Tool bodies constructed of Polyethermide and Polyphenylsulfone. Digitizer operates at 30 Hz max rate with 1m x 1m active volume. Connectivity is integrated with Outlook MRI system. Sterilization: steam for tool bodies, ETO for reflective spheres. Software provides 3D slice reformatting and graphical overlays.
Indications for Use
Indicated for patients requiring diagnostic or therapeutic interventional procedures, such as needle biopsies and drainages, guided by MRI imaging.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
AUG 23 1999
99 1943
## SUMMARY OF SAFETY AND EFFECTIVENESS
(As required by 21 CFR 807.92)
#### General Information 1.
| Classification: | Class II<br>Magnetic Resonance Imaging (MRI) System |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Magnetic Resonance Imaging (MRI) Option |
| Proprietary Name: | MRGP Optical Tracking Unit |
| Establishment Registration: | Manufacturer:<br>Picker Nordstar, Inc.<br>Ayritie 4, Vantaa<br>FIN-01510 Vantaa Finland<br>FDA Facility Registration: #9680194<br>United States Representative:<br>Picker International, Inc.<br>World Headquarters<br>595 Miner Road<br>Highland Heights, Ohio 44143<br>Contact: Elaine K. Keeler, Ph.D.<br>Phone: (440) 473-3000<br>FDA Owner Number: #1580240<br>FDA Registration Number: #1525965 |
| Performance Standards: | No applicable performance standards have been<br>issued under section 514 of the Food, Drug and<br>Cosmetic Act. |
#### 2. Intended Uses
The Outlook system is indicated for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
The MRGP Optical Tracking Unit is designed to support and guide diagnostic and therapeutic interventional procedures, such as needle biopsies and drainages.
| PICKER INTERNATIONAL, INC. | (MRGP-OP) | 05/27/99 | I.- |
|----------------------------|-----------|----------|-----|
|----------------------------|-----------|----------|-----|
{1}------------------------------------------------
#### 3. Device Description
,
The main features in the Optical Tracking Unit include: 1) infrared digitizer hardware; 2) specialized tools and tracking devices for procedures; 3) software that guides the imaging plane based on information from the digitizer; 4) software utilities that allow for slice reformatting of 3D image sets and 5) MRGP graphical overlays for planning and intraoperative use.
This guidance mechanism shortens the time needed to find the correct imaging plane, and is useful in cases that require a difficult trajectory (i.e. double oblique imaging planes).
#### Safety and Effectiveness 4.
The following substantial equivalence chart has been compiled to demonstrate the equivalence of the MRGP Optical Tracking Unit described in this submission with the Picker MRGP Basic Package (K983342) and the Picker ViewPoint Passive Tool Option (K990868).
| Parameter | MRGP Optical Tracking<br>Unit | Predicate Devices |
|-------------------------|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| System Compatibility | Same. | 0.23T Outlook,<br>0.23T Outlook Proview<br>(See K983342) |
| Software | Special MRGP Software for<br>the Outlook. | Standard Outlook Software.<br>(See K983342) |
| Imaging Plane Selection | Optically Guided. | Manual. (See K983342) |
| Sequence type | Same. | 2D/3D Gradient echo, FSE, Single-<br>shot FSE. (See K983342) |
| Sequence capabilities | Same. | Dynamic imaging with auto start and<br>keyhole imaging capabilities.<br>Typical reconstruction of 200 ms<br>per image. (See K983342) |
| Sequence resolution | Same. | FOV- 4 to 40 cm<br><br>Slice thickness-<br>2D: 1-100mm (0.1mm steps)<br>3D: 0.4-100mm (0.1mm steps)<br>Matrix- up to 512<br>(See K983342) |
## Substantial Equivalence Chart
PICKER INTERNATIONAL, INC.
(MRGP-OP)
{2}------------------------------------------------
| Parameter | MRGP Optical Tracking<br>Unit | Predicate Devices |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of Digitizer | Same. | Infrared signals emitted from the<br>Position Sensor Assembly (PSA) are<br>reflected off reflective markers<br>mounted on the tool. The reflected<br>signal is detected by the PSA with<br>two optical detectors.<br>(See K990868) |
| Max. Digitizer Rate | Same. | 30 Hz for one or two tools; for three<br>tools, the rate is 30 Hz for one of<br>the tools and 15 Hz for the other<br>two. (See K990868) |
| Active Digitizer Volume | Same. | Silo shape, with 1 meter diameter<br>and 1 meter length. (See K990868) |
| Type of Sterilization | Same. | Tool bodies are steam sterilized and<br>the reflective spheres are sterilized<br>with ETO. (See K990868) |
| Material Composition of<br>tool bodies | Polyethermide and<br>Polyphenylsulfone | Aluminum, titanium or stainless steel<br>(See K990868) |
| Indications for Use | The MRGP Optical Tracking<br>Unit is designed to support and<br>guide diagnostic and<br>therapeutic interventional<br>procedures, such as needle<br>biopsies and drainages. | The MRGP Basic Package is<br>designed to support and guide<br>diagnostic interventional procedures,<br>such as needle biopsies and<br>drainages. (See K983342) |
PICKER INTERNATIONAL, INC.
:
(MRGP-OP)
05/27/99
.
I - 3
.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is centered in the image and is the only element present. The background is white.
Image /page/3/Picture/1 description: The image contains a logo of the U.S. Department of Health and Human Services (HHS). The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it, topped by a circle. The text "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged around the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 3 1999
Elaine Keeler, Ph.D. Manager, Clinical Science Picker International, Inc. 595 Miner Road Highland Heights, Ohio 44143
K991943 Re:
MRGP Optical Tracking Unit (Interventional MRI Accessory) Dated: June 7, 1999 Received: June 9, 1999 Regulatory Class: II 21 CFR 892.1000/Procode: 90 LNH
Dear Dr. Keeler:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page I of 1
# 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
MRGP Optical Tracking Unit Device Name:
### Indications for Use:
The Outlook system is indicated for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (TI), and spin-spin relaxation time (T2)) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
The MRGP Optical Tracking Unit is designed to support and guide diagnostic and therapeutic interventional procedures, such as needle biopsies and drainages.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|--------------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices | |
| 510(k) Number | K991943 |
| Prescription Use (Per 21 CFR 801.109) | OR | Over-The-Counter Use (Optional Format 1-2-96) |
|---------------------------------------|----|-----------------------------------------------|
|---------------------------------------|----|-----------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.