ADDITIONAL IMS ASSAYS FOR THE BAYER ADVIA IMS SYSTEM
Applicant
Bayer Corp.
Product Code
CZW · Immunology
Decision Date
Nov 9, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5240
Device Class
Class 2
Indications for Use
The Bayer ADVIA IMS C3 assay is an in vitro diagnostic device intended to measure Complement C3 (C3) in human serum. Such measurements are used as an aid in the diagnosis and treatment of immunologic disorders. The Bayer ADVIA IMS C4 assay is an in vitro diagnostic device intended to measure Complement C4 (C4) in human serum. Such measurements are used as an aid in the diagnosis and treatment of immunologic disorders. The Bayer ADVIA IMS IgA assay is an in vitro diagnostic device intended to measure Immunoglobulin A (IgA) in human serum. Such measurements are used as an aid in the diagnosis and treatment of abnormal protein metabolism and the body's inability to resist infectious agents. The Bayer ADVIA IMS IgG assay is an in vitro diagnostic device intended to measure Immunoglobulin G (IgG) in human serum. Such measurements are used as an aid in the diagnosis and treatment of autoimmune diseases, chronic or recurrent infections and abnormal protein metabolism. The Bayer ADVIA IMS Immunoglobulin M (IgM) assay is an in vitro diagnostic device intended to measure IgM in human serum. Such measurements are used as an aid in the diagnosis and treatment of chronic or recurrent infections and abnormal protein metabolism. The Bayer ADVIA IMS Transferrin assay is an in vitro diagnostic device intended to measure transferrin (TRF) in human serum. Such measurements are used as an aid in the diagnosis and treatment of malnutrition, chronic infection, acute hepatitis, polycythemia, pernicious anemia, and red blood cell disorders, such as iron deficiency anemia. The Bayer ADVIA IMS Vancomycin assay is an in vitro diagnostic device intended to measure vancomycin, an antibiotic drug, in human serum. Measurements of vancomycin are used as an aid in the diagnosis and treatment of vancomycin overdose and in monitoring therapeutic levels of vancomycin to ensure appropriate therapy. The Bayer ADVIA IMS Valproic Acid Assay is an in vitro diagnostic device intended to measure valproic acid, an anti-epileptic drug, in human serum. Measurements of valproic acid are used as an aid in the diagnosis and treatment of valproic acid overdose and in monitoring therapeutic levels of valproic acid to ensure appropriate therapy.
Device Story
Bayer ADVIA IMS system performs automated in vitro quantitative analysis of serum analytes (C3, C4, IgA, IgG, IgM, Transferrin, Vancomycin, Valproic Acid). System utilizes specific reagents and calibrators to process patient serum samples. Operation occurs in clinical laboratory settings by trained technicians. Device transforms sample inputs into quantitative concentration outputs; results assist clinicians in diagnosing immunologic/metabolic disorders, monitoring therapeutic drug levels, and managing patient treatment. System provides automated measurement, reducing manual processing requirements.
Clinical Evidence
Bench testing only. Performance validated via correlation studies against predicate devices (n=55 to 106 per assay). Correlation coefficients (r) ranged from 0.972 to 0.994. Precision (total CV) reported for various concentration levels, generally ranging from 1.4% to 3.9%. Interference testing conducted for bilirubin, hemoglobin, and triglycerides; effects on analyte measurement were documented.
Technological Characteristics
In vitro diagnostic assays for use on Bayer ADVIA IMS automated systems. Reagents and calibrators specific to each analyte. Quantitative measurement principle (immunoturbidimetric or similar immunoassay). System is a clinical laboratory analyzer. Connectivity and software integration are inherent to the ADVIA IMS platform.
Indications for Use
Indicated for quantitative measurement of C3, C4, IgA, IgG, IgM, Transferrin, Vancomycin, and Valproic Acid in human serum to aid in diagnosis and treatment of immunologic disorders, protein metabolism abnormalities, infectious diseases, autoimmune conditions, nutritional status, anemia, and therapeutic drug monitoring.
Regulatory Classification
Identification
A complement components immunological test system is a device that consists of the reagents used to measure by immunochemical techniques complement components C1q , C1r , C1s , C2 , C3 , C4 , C5 , C6 , C7 , C8 , and C9 , in serum, other body fluids, and tissues. Complement is a group of serum proteins which destroy infectious agents. Measurements of these proteins aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components.
Predicate Devices
Dade Behring BNA Complement C3 method
Dade Behring BNA Complement C4 method
Dade Behring BNA Immunoglobulin A method
Dade Behring BNA Immunoglobulin G method
Dade Behring BNA Immunoglobulin M method
Dade Behring BNA Transferrin method
Bayer Immuno 1 Vancomycin method
Abbott TDx Valproic Acid method
Submission Summary (Full Text)
{0}------------------------------------------------
### C3 Method for the Bayer ADVIA IMS Systems
Listed below is a comparison of the performance between the Bayer ADVIA IMS Complement C3 method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Complement C3 method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS C3 method sheet and the BNAC3 method sheet.
#### INTENDED USE
This in vitro method is intended to quantitatively measure C3 in human serum on the Bayer ADVIA IMS systems. Measurements of C3 are used to aid in the diagnosis and treatment of immunologic disorders.
| METHOD | ADVIA IMS | BNA |
|----------|-------------------------------------------------|--------------|
| Part No. | Reagents B41-3788-41<br>Calibrators B46-4088-60 | OSAP<br>OSAU |
2.1% @ 79.7 mg/dL 5.9% @ 72.4 mg/dL Precision (Total) 1.8% @ 154 mg/dL 2.1% @ 237 mg/dL
Correlation
y=0.92x + 9.7 where y=ADVIA IMS x=BNA n=101 r=0.972 Syx=12.1 mg/dL
Gabriel J. Munoz, Jr.
6/4/99
{1}------------------------------------------------
| Interfering Substances | | | |
|-------------------------|----------------------------------|---------------------|----------------------|
| Interfering Substance | Interfering Sub<br>Concentration | C3<br>Concentration | Effect<br>(% Change) |
| Bilirubin, unconjugated | 20 mg/dL | 106 mg/dL | +1 |
| Bilirubin, conjugated | 20 mg/dL | 113 mg/dL | +4 |
| Hemoglobin | 500 mg/dL | 114 mg/dL | -2 |
| Triglycerides | 1000 mg/dL | 107 mg/dL | -8 |
# Analytical Range
:
---
:
:
ಿರುವ ಸಾ
| Dilution Range | Concentration Range,<br>mg/dL |
|-------------------|-------------------------------|
| Normal | 27 - 360 |
| Out of Range Low | 6.8 - 90 |
| Out of Range High | 135 - 1800 |
{2}------------------------------------------------
### C4 Method for the Bayer ADVIA IMS Systems
Listed below is a comparison of the performance between the Bayer ADVIA IMS Complement C4 method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Complement C4 method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS C4 method sheet and the BNAC4 method sheet.
### INTENDED USE
This in vitro method is intended to quantitatively measure C4 in human serum on the Bayer ADVIA IMS systems. Measurements of C4 are used to aid in the diagnosis and treatment of immunologic disorders.
| METHOD | ADVIA IMS | BNA |
|-------------------|------------------------------------------------------------------------------------|-------------------|
| Part No. | | |
| Reagents | B41-3789-41 | OSAO |
| Calibrators | B46-4088-60 | OSAU |
| Precision (Total) | 1.9% @ 19.3 mg/dL<br>2.0% @ 34.8 mg/dL<br>1.9% @ 49.3 mg/dL | 2.9% @ 15.7 mg/dL |
| Correlation | y=0.92x - 0.3<br>where<br>y=ADVIA IMS<br>x=BNA<br>n=94<br>r=0.989<br>Syx=1.9 mg/dL | |
Gabriel J. Munoz, Jr.
6/4/99
{3}------------------------------------------------
| Interfering Substance | Interfering Sub<br>Concentration | C4<br>Concentration | Effect<br>(% Change) |
|-------------------------|----------------------------------|---------------------|----------------------|
| Bilirubin, unconjugated | 20 mg/dL | 19 mg/dL | 0 |
| Bilirubin, conjugated | 20 mg/dL | 21 mg/dL | 0 |
| Hemoglobin | 500 mg/dL | 20 mg/dL | 0 |
# Analytical Range
| Dilution Range | Concentration Range,<br>mg/dL |
|-------------------|-------------------------------|
| Normal | 7.2 - 96 |
| Out of Range Low | 4.0 - 53 |
| Out of Range High | 36 - 480 |
{4}------------------------------------------------
### IgA Method for the Bayer ADVIA IMS Systems
Listed below is a comparison of the performance between the Bayer ADVIA IMS Immunoglobulin A method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Immunoglobulin A method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS IgA method sheet and the BNA IgA method sheet.
### INTENDED USE
This in vitro method is intended to quantitatively measure IgA in human serum on the Bayer ADVIA IMS systems. Measurements of IgA are used to aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
| METHOD | | ADVIA IMS | BNA |
|-------------------|-------------|-----------------------------------------------------------------------------------|-----------------|
| Part No. | Reagents | B41-3791-41 | OSAR |
| | Calibrators | B46-4088-60 | OSAU |
| Precision (Total) | | 2.3% @ 122mg/dL<br>1.6% @ 233mg/dL<br>1.4% @ 344mg/dL | 3.5% @ 296mg/dL |
| Correlation | | y=1.00x - 1<br>where<br>y=ADVIA IMS<br>x=BNA<br>n=97<br>r=0.994<br>Syx=75.6 mg/dL | |
Gabriel J. Munoz, Jr.
6/4/99
{5}------------------------------------------------
| Interfering Substance | Interfering Sub<br>Concentration | IgA<br>Concentration | Effect<br>(% Change) |
|-------------------------|----------------------------------|----------------------|----------------------|
| Bilirubin, unconjugated | 20 mg/dL | 180 mg/dL | +2 |
| Bilirubin, conjugated | 20 mg/dL | 200 mg/dL | -3 |
| Hemoglobin | 500 mg/dL | 191 mg/dL | +1 |
| Triglycerides | 1000 mg/dL | 184 mg/dL | -8 |
:
### Analytical Range
| Analytical Range | |
|-------------------|------------------------------|
| Dilution Range | Concentration Range<br>mg/dL |
| Normal | 45 - 600 |
| Out of Range Low | 11.3 - 150 |
| Out of Range High | 360 - 4800 |
:
.
·
{6}------------------------------------------------
### IgG Method for the Bayer ADVIA IMS Systems
Listed below is a comparison of the performance between the Bayer ADVIA IMS Immunoglobulin G method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Immunoglobulin G method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS IgG method sheet and the BNA IgG method sheet.
### INTENDED USE
This in vitro method is intended to quantitatively measure IgG in human serum on the Bayer ADVIA IMS systems. Measurements of IgG are used to aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
| METHOD | | ADVIA IMS | BNA |
|----------|-------------|-------------|------|
| Part No. | Reagents | B41-3792-41 | OSAS |
| | Calibrators | B46-4088-60 | OSAU |
Precision (Total) 1.8% @ 772mg/dL 2.7% @ 1317mg/dL 1.6% @ 1416mg/dL 3.4% @ 2141mg/dL
Correlation
y=0.97 x - 3 where y=ADVIA IMS x=BNA n=97 r=0.994 Syx=118 mg/dL
Gabriel J. Murau, Jr.
6/4/99
{7}------------------------------------------------
| Interfering Substance | Interfering Sub<br>Concentration | IgG<br>Concentration | Effect<br>(% Change) |
|-------------------------|----------------------------------|----------------------|----------------------|
| Bilirubin, unconjugated | 20 mg/dL | 876 mg/dL | -3.0 |
| Bilirubin, conjugated | 20 mg/dL | 953 mg/dL | -3.0 |
| Hemoglobin | 500 mg/dL | 897 mg/dL | +7.0 |
| Triglycerides | 1000 mg/dL | 887 mg/dL | -3.0 |
:
# Analytical Range
| Dilution Range | Concentration Range,<br>mg/dL |
|-------------------|-------------------------------|
| Normal | 225 - 3,000 |
| Out of Range Low | 56 - 750 |
| Out of Range High | 1128 - 15,000 |
{8}------------------------------------------------
### IgM Method for the Bayer ADVIA IMS Systems
Listed below is a comparison of the performance between the Bayer ADVIA IMS Immunoglobulin M method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Immunoglobulin M method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS IgM method sheet and the BNA IgM method sheet.
### INTENDED USE
This in vitro method is intended to quantitatively measure IgM in human serum on the Bayer ADVIA IMS systems. Measurements of IgM are used to aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
| METHOD | | ADVIA IMS | BNA |
|----------|-------------|-------------|------|
| Part No. | Reagents | B41-3793-41 | OSAT |
| | Calibrators | B46-4088-60 | OSAU |
Precision (Total)
3.6% @ 69.0mg/dL 1.9% @ 117mg/dL 2.4% @ 128mg/dL 1.5% @ 177mg/dL
Correlation
y=1.05x - 13.4 where y=ADVIA IMS x=BNA n=89 r=0.990 Syx=123.5 mg/dL
Gabriel J. Murray, Jr.
6/4/99
{9}------------------------------------------------
| Interfering Substance | Interfering Sub<br>Concentration | IgM<br>Concentration | Effect<br>(% Change) |
|-------------------------|----------------------------------|----------------------|----------------------|
| Bilirubin, unconjugated | 20 mg/dL | 77 mg/dL | +3.9 |
| Bilirubin, conjugated | 10 mg/dL | 85 mg/dL | +1.2 |
| Hemoglobin | 500 mg/dL | 86 mg/dL | 0.0 |
.
# Analytical Range
·
| Dilution Range | Concentration Range, |
|---------------------|----------------------|
| | mg/dL |
| Normal | 30 - 400 |
| Out of Range Low | 10 - 133 |
| Out of Range High 1 | 241 - 3,200 |
| Out of Range High 2 | 1,203 - 16,000 |
:
:
{10}------------------------------------------------
### TRF Method for the Bayer ADVIA IMS Systems
Listed below is a comparison of the performance between the Bayer ADVIA IMS Transferrin method and a similar device that was granted clearance of substantial equivalence (Dade Behring BNA Transferrin method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS TRF method sheet and the BNATRF method sheet.
### INTENDED USE
This in vitro method is intended to quantitatively measure TRF in human serum on the Bayer ADVIA IMS systems. Measurements of TRF are used to aid in the diagnosis and treatment of malnutrition, chronic infection, acute hepatitis, polycythemia, pernicious anemia and red blood disorders, such as iron deficiency anemia.
| METHOD | | ADVIA IMS | BNA |
|----------|-------------|-------------|------|
| Part No. | Reagents | B41-3795-41 | OSAX |
| | Calibrators | B46-4088-60 | OSAU |
Precision (Total) 2.2% @ 127 mg/dL 2.7% @ 303 mg/dL 1.9% @ 268 mg/dL 3.1% @ 400 mg/dL
Correlation
y=0.96x - 3 where y=ADVIA IMS x=BNA n=106 r=0.979 Syx=16.8 mg/dL
Gabriel J. Munoz, Jr.
6/4/99
{11}------------------------------------------------
| Interfering Substance | Interfering Sub<br>Concentration | TRF<br>Concentration | Effect<br>(% Change) |
|-------------------------|----------------------------------|----------------------|----------------------|
| Bilirubin, unconjugated | 20 mg/dL | 212 mg/dL | +2 |
| Bilirubin, conjugated | 20 mg/dL | 231 mg/dL | 0 |
| Hemoglobin | 500 mg/dL | 261 mg/dL | 0 |
| Triglycerides | 1000 mg/dL | 215 mg/dL | -4 |
# Analytical Range
| Dilution Range | Concentration Range,<br>mg/dL |
|-------------------|-------------------------------|
| Normal | 54 - 720 |
| Out of Range Low | 13 - 180 |
| Out of Range High | 270 - 3,600 |
.
:
{12}------------------------------------------------
### Vancomycin Method for the Bayer ADVIA IMS Systems
Listed below is a comparison of the performance between the Bayer ADVIA IMS Vancomycin method and a similar device that was granted clearance of substantial equivalence (Bayer Immuno 1 Vancomycin method) The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS Vancomycin method sheet and the Immuno 1 Vancomycin method sheet.
### INTENDED USE
This in vitro method is intended to quantitatively measure vancomycin in human serum on the Bayer ADVIA IMS systems. Measurements of vancomycin are used to aid in attaining optimum therapy in patients treated with the drug.
| METHOD | | ADVIA IMS | Immuno 1 |
|-------------------|-------------|------------------------------------------------------------------------------------------|-------------------|
| Part No. | Reagents | B41-3767-41 | T01-3705-01 |
| | Calibrators | B46-4117-01 | T03-3714-01 |
| Analytical Range | | 0.4-50 µg/mL | |
| Precision (Total) | | 2.8% @ 8.6 µg/mL | 8.9% @ 6.7 µg/mL |
| | | 3.4% @ 21.5 µg/mL | 7.5% @ 23.3 µg/mL |
| | | 3.9% @ 35.8 µg/mL | 8.1% @ 32.4 µg/mL |
| Correlation | | y=1.02x + 0.68<br>where<br>y=ADVIA IMS<br>x=Immuno 1<br>n=55<br>r=0.987<br>Svx=1.6 µg/dL | |
#### Interfering Substances
| Interfering Substance | Interfering Substance<br>Concentration | | Vancomycin<br>Concentration | | Effect<br>% Change |
|--------------------------|----------------------------------------|------------|-----------------------------|---------|--------------------|
| | | | (µmol/L) | (µg/mL) | |
| Bilirubin (unconjugated) | 0.43 mmol/L | 25 mg/dL | 10.6 | 15.3 | +0.3 |
| Bilirubin (conjugated) | 0.43 mmol/L | 25 mg/dL | 11.0 | 15.9 | +0.6 |
| Hemoglobin | 10 g/L | 1000 mg/dL | 11.1 | 16.1 | -7.7 |
| Lipemia (Triglycerides) | 11.4 mmol/L | 1000 mg/dL | 11.0 | 16.0 | -0.2 |
Gabriel J. Murray Jr.
6/4/99
{13}------------------------------------------------
### Valproic Acid Method for the Bayer ADVIA IMS Systems
Listed below is a comparison of the performance between the Bayer ADVIA IMS Valproic Acid method and a similar device that was granted clearance of substantial equivalence (Abbott TDx Valproic Acid method). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA IMS Valproic Acid method sheet and the TDx Valproic Acid method sheet.
#### INTENDED USE
This in vitro method is intended to quantitatively measure valproic acid in human serum on the Bayer ADVIA IMS systems. Measurements of valproic acid are used to aid in attaining optimum therapy in patients treated with the drug.
| METHOD | | ADVIA IMS | TDx |
|-------------------|-------------|--------------------|-------------------|
| Part No. | Reagents | B41-3766-41 | 9514-20 |
| | Calibrators | B46-4118-01 | 9514-01 |
| Analytical Range | | 0.3-150 µg/mL | |
| Precision (Total) | | 3.8% @ 20.3 µg/mL | 3.4% @ 37.5 µg/mL |
| | | 2.4% @ 60.2 µg/mL | 3.4% @ 75 µg/mL |
| | | 2.8% @ 108.5 µg/mL | 3.7% @ 125 µg/mL |
| Correlation | | y=1.11x + 0.27 | |
| | | where | |
| | | y=ADVIA IMS | |
| | | x=TDx | |
| | | n=55 | |
| | | r=0.993 | |
| | | Sux=3.79 µg/mL | |
#### Interfering Substances
| Interfering Substance | Interfering Substance<br>Concentration | Valproic acid<br>Concentration | | Effect<br>% Change |
|--------------------------|----------------------------------------|--------------------------------|---------|--------------------|
| | | (µmol/L) | (µg/mL) | |
| Bilirubin (unconjugated) | 0.43 mol/L<br>25 mg/dL | 573 | 82.5 | +6 |
| Bilirubin (conjugated) | 0.43 mol/L<br>25 mg/dL | 542 | 78.1 | +1 |
| Hemoglobin | 10 g/L<br>1000 mg/dL | 557 | 80.2 | +4 |
| Lipemia (Triglycerides) | 5.7 mmol/L<br>500 mg/dL | 561 | 80.8 | +3 |
Gabriel J. Munoz Jr.
6/4/99
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Nov - 9 1999
Mr. Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corp. Business Group Diagnostics 511 Benedict Avenue Tarrytown, New York 10591-5097
Re: K991907
Trade Name: 8 Additional Assays for the Bayer ADVIA® Integrated Modular System (IMS) Regulatory Class: II Product Code: CZW, CFQ, DBI, DDG, KTO, LEG, LEH Dated: September 3, 1999 Received: September 7, 1999
Dear Mr. Muraca:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might
{15}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{16}------------------------------------------------
Page 1 of 3
# 510(k) Number (if known):
### Device Name: Bayer ADVIA® Integrated Modular System (IMS)
Indications For Use:
The Bayer ADVIA IMS C3 assay is an in vitro diagnostic device intended to measure Complement C3 (C3) in human serum. Such measurements are used as an aid in the diagnosis and treatment of immunologic disorders.
The Bayer ADVIA IMS C4 assay is an in vitro diagnostic device intended to measure Complement C4 (C4) in human serum. Such measurements are used as an aid in the diagnosis and treatment of immunologic disorders.
The Bayer ADVIA IMS IgA assay is an in vitro diagnostic device intended to measure Immunoglobulin A (IgA) in human serum. Such measurements are used as an aid in the diagnosis and treatment of abnormal protein metabolism and the body's inability to resist infectious agents.
Dran Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K991907
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
{17}------------------------------------------------
510(k) Number (if known):
### Device Name: Bayer ADVIA® Integrated Modular System (IMS)
Indications For Use:
The Bayer ADVIA IMS IgG assay is an in vitro diagnostic device intended to measure Immunoglobulin G (IgG) in human serum. Such measurements are used as an aid in the diagnosis and treatment of autoimmune diseases, chronic or recurrent infections and abnormal protein metabolism.
The Bayer ADVIA IMS Immunoglobulin M (IgM) assay is an in vitro diagnostic device intended to measure IgM in human serum. Such measurements are used as an aid in the diagnosis and treatment of chronic or recurrent infections and abnormal protein metabolism.
The Bayer ADVIA IMS Transferrin assay is an in vitro diagnostic device intended to measure transferrin (TRF) in human serum. Such measurements are used as an aid in the diagnosis and treatment of malnutrition, chronic infection, acute hepatitis, polycythemia, pernicious anemia, and red blood cell disorders, such as iron deficiency anemia.
Fein Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K991907
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use - -
(Optional Format 1-2-96)
{18}------------------------------------------------
Page 3 of 3
510(k) Number (if known):
### Device Name: Bayer ADVIA® Integrated Modular System (IMS)
Indications For Use:
The Bayer ADVIA IMS Vancomycin assay is an in vitro diagnostic device intended to measure vancomycin, an antibiotic drug, in human serum. Measurements of vancomycin are used as an aid in the diagnosis and treatment of vancomycin overdose and in monitoring therapeutic levels of vancomycin to ensure appropriate therapy.
The Bayer ADVIA IMS Valproic Acid Assay is an in vitro diagnostic device intended to measure valproic acid, an anti-epileptic drug, in human serum. Measurements of valproic acid are used as an aid in the diagnosis and treatment of valproic acid overdose and in monitoring therapeutic levels of valproic acid to ensure appropriate therapy.
Hean Coogh
(Division Sign-Off)
Division of Clinical Laboratory Devices
× 510(k) Number .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.