K991853 · Idem (Int'L Dist. of Electronics For Medicine) · GWJ · Oct 25, 1999 · Neurology
Device Facts
Record ID
K991853
Device Name
INTERACOUSTICS EP 15 ABR STAND ALONE UNIT
Applicant
Idem (Int'L Dist. of Electronics For Medicine)
Product Code
GWJ · Neurology
Decision Date
Oct 25, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1900
Device Class
Class 2
Indications for Use
The Interacoustics EP15 ABR Stand Alone Unit is a diagnostic tool for computer-based testing. Data is collected in a Windows 95,98 database format that allows for easy inspection of results. The EP15 Stand Alone Unit produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
Device Story
The Interacoustics EP15 ABR is a diagnostic system for evoked response auditory measurements; it functions as a computer-based stimulator and data collector. The device generates sound stimuli (clicks, tone bursts, various envelope shapes) delivered via transducers (e.g., insert phones) to the patient. It records electroencephalogram (EEG) signals through a 2-channel amplifier system. The system operates on a Windows 95/98 platform, allowing clinicians to inspect, store, and manage waveform data in a database. It features programmable stimulus parameters (frequency, intensity, rate) and signal processing filters (FIR filters). The device is used in clinical settings by healthcare professionals to assess auditory function. The output, presented on an industrial TFT display, assists clinicians in diagnosing auditory conditions by analyzing evoked potential waveforms. The system includes safety features like optical isolation and a built-in isolation transformer.
Clinical Evidence
No clinical data provided; bench testing only. The device's safety and effectiveness were established through compliance with performance and safety standards including VDE750, EN 60601-1 (General Safety), EN 60601-1-1 (Safety of Systems), EN 60601-1-2 (EMC), EN 60601-2-26 (Electroencephalographs), and EN 60645-3 (Auditory test signals).
Technological Characteristics
2-channel diagnostic system; 12.1" TFT display; Windows 95/98 OS. Amplifiers: 80 dB gain, 100-8000 Hz frequency response, >120dB CMR ratio. Stimulators: Click, Tone Burst, various envelope shapes (Blackman, Gaussian, etc.). Connectivity: Standard PC interface. Safety: Optical isolation, built-in isolation transformer. Standards: VDE750, EN 60601-1, EN 60601-1-1, EN 60601-1-2, EN 60601-2-26, EN 60645-3. Software: Rule-based signal processing with FIR filtering.
Indications for Use
Indicated for use as a diagnostic tool for evoked response auditory measurements or electroencephalogram activation in patients requiring auditory testing.
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
Predicate Devices
ICS Medical Corp. Chartr EP System
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 2 5 1999
Page 1 of 3
## 510(k) SUMMARY
K 991853
INTERNATIONAL DISTRIBUTORS OF PREPARED BY: ELECTRONICS FOR MEDICINE, INC. (IDEM) 4814 East Second Street Benicia, CA 94510
CONTACT PERSON:
Donna Ward, President
TELEPHONE:
707-746-6334
May 28, 1999
DATE ON WHICH THE SUMMARY WAS PREPARED:
NAME IF DEVICE:
Interacoustics Model EP15 ABR Stand Alone Unit
COMMON NAME:
Evoked Response Auditory Stimulator
PREDICATE DEVICE:
ICS Medical Corp. Chartr EP System
DESCRIPTION OF DEVICE:
The Interacoustics EP15 ABR Stand Alone Unit is a diagnostic tool for computerbased testing. Data is collected in a Windows 95,98 database format that allows for easy inspection of results. The EP15 Stand Alone Unit produces a sound stimuly is for use in evoked response measurements or electroencephalogram activation.
{1}------------------------------------------------
## Comparison of the Interacoustics Model EP15 and the ICS Medical Chartr EP System:
Indication for use – Identical for both units.
Similarities and differences:
| EP15 ABR | Chartr EP |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display Description: | |
| 12.1" Wide Angle Industrial TFT | 15" SVGA Monitor (others available) |
| Software Features: | |
| Windows 95/98 Operating System | Same |
| AMPLIFIERS: | |
| Channels: 2 | 2 (optionally up to 8) |
| Gain: 80 dB | X50 - X5000000 |
| PREAMPLIFIERS: | |
| Frequency response: 100 to 8000 Hz | Same |
| CMR Ratio: >120dB at 50/60 Hz. | >100dB at 50/60 Hz |
| Noise: 5.5nV/Hz, 0.3µV RMS (0-3 kHz); | <1.5 µV RMS (0.1 Hz – 5 kHz) |
| 5.5nV/Hz, 0.5µV RMS (100-8000 Hz) | |
| Points Per Trace: 450 (30 kHz sample) | 600 |
| ANALOG FEATURES: | |
| Low pass: 1000, 1500, 2000, 3000, 5000<br>Hz, none (12000Hz). 33 taps FIR filter. | 15 Hz to 25 kHz, 12 dB/octave |
| High pass: None (5 Hz), 50, 100, 150, 300<br>500 Hz. 6dB/octave. | DC, 0.002 – 1 kHz, 6 dB/octave |
| STIMULATORS: | |
| Stimuli: Click, Tone Burst, Blackman,<br>Gaussian, Hanning, Hamming, Bartlett,<br>Rectangle, Manual (rise/plateau/fall) | Click, Tone Burst |
| Parameters for tonal stimuli; frequency,<br>intensity, rise/fall time, plateau duration,<br>envelope shape: Programmable | Programmable<br>0.1 - 100/sec |
| Rate: 7.1, 13.1, 21.1, 23.1, 29.1, 31.1, 35.1,<br>41.1, 45.1, 49.1 Hz, ext. trigger input | |
| Intensity: 20 dB to 130 dB peSPL; 10 dB to<br>100 dB HL In 1 dB step. | -12 dB to +128 dB SPL |
| Masking: White noise; masking ear is<br>opposite side of stimulus | White noise, programmable intensity |
| Transducers: EAR 3A Insert Phones-ABR | TDH-49 earphones, insert earphones, bone<br>conduction transducer |
| SAFETY CHARACTERISTICS: | |
| Optical isolation: Yes | Yes |
| Built-in isolation transformer: Yes | Yes |
| (CONTINUED – COMPARISON) | |
| EP15 ABR | Chartr EP |
| Controlled parameters: Stimuli Rate; Number<br>of Stimuli; Stimuli Polarity; Click; Tone Burst<br>(frequency, number of sin waves, window);<br>Stimulus Intensity; Number of Curve Pr.<br>Intensity; Intensity (ascend, descend); Soft<br>Attenuator; Stimulus Ear (right, left,<br>simultaneously); Masking Level; Preliminary<br>Filter Setting (low, high pass filter); Recording<br>Onset; Automatic Next Intensity (wave repro<br>level setting); Ext. Trigger Output Duration;<br>Rejection System Rejection Level; Gain<br>(manual, automatic); Display Options (Invert<br>curves on screen; Origin line; Latency Norm.<br>Report Templates; After Filtering; Print out;<br>Manual Stimulus to Familiarization; Talk<br>Forward; Talk Back Monitor. | Stimulated Ear, Masked Ear, Stimulus Intensity,<br>Masking Stimulus, Stimulus Transducer,<br>Stimulus Type, Stimulus Polarity, Stimulus<br>Characteristics, Number of Sweeps Acquires,<br>Stimulus Presentation Rate, Sweep Time,<br>Number of Channels, Amplifier Gain, Filter<br>Characteristics, Inclusion of Notch Filter,<br>Inclusion of Artifact Rejection |
| Data Collection: Impedance Test; Waveform<br>Buffer (A/B, contra, ipsi-contra, A-B=Noise);<br>Curve (Hide, Fixate, Merge, Delete); Show<br>Online EEG, Store Waveforms in unlimited<br>Storage database | Tests Impedance of Patient Electrode<br>Connections, Display Waveform Buffers<br>During Examination, Displays on-going EEG<br>Activity, Stores Waveforms, Stores the<br>Waveform Presentation |
| Dimensions: 14" x 10" x 15" | 22.5" x 17.5" x 7.8" |
| Weight: 26.5 lbs. | 58 lbs. |
| Power Supply: Input volts: 90 to 250 VAC<br>Universal Input Switch Mode; Safety: VDE750,<br>EN60601-1, IEC601, IEC1010, UL544, CSA<br>22.2 | 200 Watts, 50/60 Hz, 120 or 240 volts |
| Keyboard:<br>101-key IBM standard | 101 - key IBM standard |
| Ancillary Functions: Help system.<br>Exports data of one patient to diskette. | Help system.<br>Exports data of one patient to diskette.<br>Backup procedure automatically implemented. |
{2}------------------------------------------------
## SAFETY AND EFFECTIVENESS:
ﺎﺭ ﺍﻟﻤﺮﺍﺟﻊ
The Interacoustics Model EP15 ABR Stand Alone Unit is in compliance with the following performance and safety standards:
VDE750; EN 60601-01 (General Safety) Class I, Type BF; EM 60601-1-1 (Safety of Systems); EN 60601-1-2 (EMC); EN 60601-2-26 (Electroencephalographs); EN 60645-3 (Auditory test signals)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, combined with three human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
APR - 9 2012
Ms. Donna Ward President IDEM INTERNATIONAL DISTRIBUTORS OF ELECTRONICS FOR MEDICINE, INC. 4814 East Second Street Benicia, California 94510
Re: K991853
Trade/Device Name: Interacoustics Model EP15 ABR Stand Alone Unit Regulation Number: 21 CFR 882.1900 Regulation Name: Evoked response auditory stimulator Regulatory Class: II Product Code: GWJ Dated (Date on orig SE ltr): August 5, 1999 Received (Date on orig SE ltr): August 20, 1999
Dear Ms. Ward:
This letter corrects our substantially equivalent letter of October 25, 1999.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Kesia Alexander
Image /page/4/Picture/7 description: The image shows a signature on the left side and some text on the right side. The signature appears to be a stylized cursive signature, possibly initials. The text on the right is arranged in lines, but the content is not clear enough to read.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K991853
Interacoustics Model EP15 ABR Stand Alone Unit
Device Name:
Indications For Usc:
The Interacoustics EP15 ABR Stand Alone Unit is a diagnostic tool for computerrne interacoucilled in a Windows 98 database format that allows for easy inspection of results. The EP15 Stand Alone Unit produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| <div align="center"> <img height="30" src="data:image/png;base64,iVBORw0KGgoAAAANSUhEUgAAAAUAAAAFCAYAAACNbyblAAAAHElEQVQI12P4//8/w+r8/08gZSAgYGRgYGBgYAAJzQG35+5X0AAAAABJRU5ErkJggg==" width="100"/></div> | |
| (Division Sign-Off) | |
| Division of General Restorative Devices | K991853 |
| 510(k) Number | |
Prescription Use 
OR
Over-The-Counter Use
મ
Prescription Use_ (Por 21 CFR 801.109)
... : (Optional Formal 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.