← Product Code [JOS](/productcode/JOS) · K991830

# CAPSULAR SHRINKAGE ELECTRODE (K991830)

_Linvatec Corp. · JOS · Aug 20, 1999 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K991830

## Device Facts

- **Applicant:** Linvatec Corp.
- **Product Code:** [JOS](/productcode/JOS.md)
- **Decision Date:** Aug 20, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures.

## Device Story

HeatWave Electrode is a sterile, single-use, monopolar electrosurgical device. It connects to an external electrosurgical generator via an electrosurgical pencil and dispersive pad. Used by surgeons during arthroscopic procedures to perform soft tissue electro-coagulation. The device delivers electrical energy to the target tissue to achieve coagulation. It is provided individually or in a kit containing the electrode, pencil, and dispersive pad.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Monopolar electrode. Materials: 303 Stainless Steel (tip/shaft), Kynar (shrink tube), Polystyrene (hub). Dimensions: 2.3mm, 3.0mm, 4.0mm diameters; 4-8.5 inch working length. Sterilization: ETO. Energy source: External electrosurgical generator.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- TAC-S Monopolar Cautery Probe ([K984185](/device/K984185.md))
- ArthoWand CAPS (Unknown)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K991830

、元 20 103

August 5, 1999

# SUMMARY OF SAFETY AND EFFECTIVENESS

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, Linvatec Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for the HeatWave™ Electrode, 510(k) Number K991830.

#### A. Submitter

Linvatec Corporation 11311 Concept Boulevard Largo, Florida 33773-4908

### Company Contact B.

Laura Seneff Manager, Regulatory Affairs

### C. Device Name

| Trade Name:                            | HeatWave™ Electrode                                                         |
|----------------------------------------|-----------------------------------------------------------------------------|
| Common Name:                           | Electrode                                                                   |
| Classification Names:                  | Electrosurgical cutting and coagulation device<br>and accessories, 878.4400 |
| Proposed Class/Device:<br>Product Code | Class II, 79 JOS, Electrode,<br>Electrosurgical                             |

### D. Predicate/Legally Marketed Devices

TAC™ Probes ORATEC Interventions, Inc.

ArthoWand CAPS ArthroCare Corporation

{1}------------------------------------------------

Summary of Safety and Effectiveness HeatWave Electrode 510(k) # K991830 August 5, 1999 Page 2 of 3

### E. Device Description

The HeatWave™ Electrode is a sterile, single-use, monopolar electrode designed to be used in conjunction with an electrosurgical generator via an electrosurgical pencil and a dispersive pad. The HeatWave Electrode will be sold individually and/or in a kit containing an electrode, electrosurgical pencil and an electrosurgical dispersive pad.

### F. Intended Use

The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures.

#### G. Substantial Equivalence

The HeatWave Electrode is substantially equivalent in design, function and intended use to the TAC™ Probes (ORATEC Interventions, Inc.)and the ArthoWand CAPS (ArthroCare Corporation).

Testing has been done to prove safety and effectiveness of the devices.

The similarities/dissimilarities to the predicates are shown in the attached table.

{2}------------------------------------------------

# Summary of Safety and Effectiveness HeatWave Electrode 510(k) # K991830 August 5, 1999 Page 3 of 3

# CHART OF SIMILARITIES AND DISSIMILARITIES

| Company                                                          | Device Name                           | Intended Use                                                                                                  | Material                                                                            | Single-Use<br>Reusable<br>Method of<br>Sterilization | Design/Dimensions                                                                |
|------------------------------------------------------------------|---------------------------------------|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------|----------------------------------------------------------------------------------|
| NEW PRODUCT<br>Linvatec                                          | HeatWaveTM<br>Electrode               | Electro-coagulation of soft<br>tissue in shoulder, ankle,<br>wrist, elbow and knee<br>arthroscopic procedures | Tip/Shaft: 303 Stainless Steel<br>Shrink Tube: Kynar<br>Hub: Polystyrene            | Single-Use<br>ETO                                    | Diameter: 2.3mm<br>3.0mm<br>4.0mm<br>Working length: 4-8½in.<br>Monopolar        |
| PREDICATE<br>ORATEC<br>Interventions, Inc.<br>510(k)#<br>K984185 | TAC-STM<br>Monopolar<br>Cautery Probe | Electro-coagulation of soft<br>tissue in shoulder, ankle,<br>wrist, elbow and knee<br>arthroscopic procedures | Tip/Shaft: 303 Stainless Steel<br>Shrink Tube: Teflon<br>Hub/Connector: ABS Plastic | Single Use<br>Sterilization<br>Method Unknown        | Diameter: 2.0mm<br>Working Length: 6 in.<br>Monopolar                            |
| PREDICATE<br>ArthroCare<br>Corporation<br>510(K) #<br>Unknown    | ARTHROWAND<br>CAPS                    | Electro-coagulation of soft<br>tissue during arthroscopic<br>procedures                                       | Shaft: 300 Series Stainless Steel<br>Tip: Ceramic<br>Hub/Connector: ABS Plastic     | Single-Use<br>Sterilization<br>Method Unknown        | Degree/Diameter:<br>25°/3.0mm<br>35°/3.0mm<br>Working length: 5.0 in.<br>Bipolar |

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with text around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 20 1999

Ms. Laura Seneff Manager, Regulatory Affairs Linvatec Corporation 11311 Concept Boulevard Largo, Florida 33773

K991830 Re: Trade Name: HeatWave™ Electrode Regulatory Class: II Product Code: JOS Dated: August 5, 1999 Received: August 10, 1999

Dear Ms. Seneff:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Ms. Laura Seneff

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely vours.

Ser

Plumell J. Rayer

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

August 5, 1999

Page 1 of 1

510(k) Number (if known): K991830

Device Name: HeatWave™ Electrode

Indications for Use:

The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-the-Counter Use

(Per 21 CFR 801.109)

Thunell V. Hagen for Jas

(Division Sign-Off Division of General Res 510lk) Number

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K991830](https://fda-staging.innolitics.com/device/K991830)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).