K991784 · Advanced Neuromodulation Systems · GZB · Jun 24, 1999 · Neurology
Device Facts
Record ID
K991784
Device Name
RENEW LEAD EXTENSION
Applicant
Advanced Neuromodulation Systems
Product Code
GZB · Neurology
Decision Date
Jun 24, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ANS Renew Lead Extensions are indicated for the treatment of chronic pain of the trunk and limbs, either as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. Lead extensions are intended to be used with Advanced Neuromodulation Systems, leads, receivers, transmitters, and antennae.
Device Story
Renew Lead Extensions are implantable accessories designed to increase the length of spinal cord stimulation leads. The device consists of a silicone connector body, Platinum/Iridium contacts, and a braided wire cable within a polyurethane sheath. It is used to connect a lead to a receiver or to externalize the connection for a trial stimulation period (up to 30 days). The device is implanted by a physician; it functions as a passive electrical conduit to facilitate signal transmission from the stimulator system to the spinal cord. By extending the reach of the lead, it allows for flexible placement of the receiver or external trial components, aiding in the management of chronic pain.
Clinical Evidence
Bench testing only. Verification testing included insulation resistance, conductor resistance, system impedance, dielectric strength, accelerated contact resistance, connector pull test, lead body flex life, strain relief flex life, stiffness, and setscrew block retention. All acceptance criteria were met.
Technological Characteristics
Materials: Platinum/Iridium contacts, polyurethane sheath, silicone connector body. Design: Braided wire cable, 10-30 cm length. Connectivity: Passive electrical extension for spinal cord stimulation leads. Sterilization: 100% ETO (SAL 1x10^-6).
Indications for Use
Indicated for patients with chronic pain of the trunk and limbs requiring spinal cord stimulation, either as a primary therapy or as an adjunct in a multidisciplinary approach.
Regulatory Classification
Identification
An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.
{0}------------------------------------------------
JUN 2 4 1999
Image /page/0/Picture/1 description: The image shows the logo "ANS" in bold, black letters. To the left of the letters is a square grid pattern, also in black. The grid is made up of small squares arranged in a 6x6 pattern. The letters are large and fill most of the horizontal space in the image.
.
K991784
Page 1 of 3
# 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Date Prepared: | May 21, 1999 |
|----------------|-----------------------------------------------------------------------------------------|
| Company: | Advanced Neuromodulation Systems, Inc.<br>One Allentown Parkway<br>Allen, TX 75002-4211 |
| Contact: | Katryna Warren |
| Phone Number: | 972-390-9800 Ext. 332 or<br>972-309-8000 |
| Fax Number: | 972-390-2881 or (972) 309-8150 |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for ANS, which is a professional organization for people in the nuclear industry. The logo consists of a grid of small squares on the left, followed by the letters "ANS" in a bold, sans-serif font. The grid of squares is likely meant to represent a nuclear reactor core, while the letters "ANS" stand for the American Nuclear Society.
K991784
Page 2 of 3
## Lead Extension 510(k) Summary of Safety and Effectiveness
#### Device Information:
| Trade Names: | Renew Lead Extension Model 3382<br>Renew Lead Extension Model 3383<br>Renew Dual Lead Extension Model 3341<br>Renew Dual Lead Extension Model 3342<br>Renew Dual Lead Extension Model 3343 |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Spinal Cord Stimulator |
| Classification Name:<br>Relief | Implanted Spinal Cord Stimulator for Pain |
#### Predicate Device:
Advanced Neuromodulation Systems, Inc., currently markets One Step Connect™ Leads under 510(k) # K960728 and Octrode Neurostim Receiver under 510(k) # K930536.
#### Device Description:
The Renew Lead Extension consists of a silicone connector body, with Platinum/Iridium contacts, and braided wire cable inside a polyurethane sheath. Lead extensions are designed to be implanted to add additional length to the lead for connection to a receiver or externalized for use during a trial stimulation period not to exceed 30 days.
#### Intended Use:
ANS Renew Lead Extensions are indicated for the treatment of chronic pain of the trunk and limbs, either as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. Lead extensions are intended to be used with Advanced Neuromodulation Systems, leads, receivers, transmitters, and antennae.
#### Comparison To Predicate Device:
The following table illustrates the comparison between the modified device, and the original legally marketed device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the letters 'ANS' in a bold, sans-serif font. To the left of the letters is a square grid pattern composed of smaller filled squares. The letters and the grid pattern are all in black, contrasting with the white background.
Page 3 of 3
| | ANS Predicate Device<br>510(k) K960728, K930536 | ANS Modified Device<br>K# Under Review |
|---------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Intended Use: | Stimulation of spinal cord for<br>treatment of chronic pain of<br>the trunk and limbs | Stimulation of spinal cord for<br>treatment of chronic pain of<br>the trunk and limbs |
| Materials: | | |
| • Electrode: | Platinum/Iridium | N/A |
| • Contact Terminal: | Stainless Steel 304 | Platinum/Iridium |
| • Insulator: | Polyurethane | Polyurethane |
| • Connector Coating | Silicone<br>(K930536) | Silicone |
| Design Features: | Braided Wire Cable | Braided Wire Cable |
| | 4,8,or 16 electrodes | N/A – The extension<br>connects with the terminal<br>end of the lead |
| | Percutaneous Introduction | Percutaneous<br>Introduction/<br>External Use |
| Dimensions: | | |
| • Length: | 58 - 80 cm | 10 - 30 cm |
| Sterilization | 100% ETO | 100% ETO |
| | Sterility Assurance Level<br>(SAL) 1x 10-6 | Sterility Assurance Level<br>(SAL) 1x 10-6 |
| Packaging: | Tray w/ Tyvek Lid | Tray w/ Tyvek Lid |
| Labeling: | Labeled as sterile,<br>prescription device | Labeled as sterile,<br>prescription device |
#### Non-clinical Testing:
The biocompatible materials used to construct the lead extension are the same type as those used for predicate devices with the same classification regulation, for the same intended use. The addition of this extension raises no significant safety or effectiveness questions.
Verification testing was performed as required by the risk analysis for this modification. The results demonstrated that the acceptance criteria were met for the following tests: Insulation Resistance, Conductor Resistance System Impedance, Dielectric Strength, Accelerated Contact Resistance, Connector Pull Test, Lead Body Flex Life, Strain Relief Flex Life, Stiffness, and Setscrew Block Retention.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of three human profiles facing to the right, arranged in a way that they resemble a bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 4 1999
Ms. Katryna Warren Advanced Neuromodulation Systems One Allentown Parkway Allen, Texas 75002-4211
K991784 Re: Trade Name: Renew TM Lead Extension Regulatory Class: II Product Code: GZB Dated: May 21, 1999 Received: May 25, 1999
Dear Ms. Warren:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Ms. Katryna Warren
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Page 1of 1
510(k) Number (if known): K 99/787
Device Name: Lead Extension
Indications For Use:
ANS Lead Extensions are indicated for the treatment of chronic pain of the trunk and limbs, either as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. Lead extensions are intended to be used with Advanced Neuromodulation Systems', leads, receivers, transmitters, and antennae.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use
les
eral Restorative Devices
0000000
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.