ROYAL SHIELD NON-STERILE BUBBLE GUM SCENTED POWDER-FREE LATEX EXAMINATION GLOVE WITH PROTEIN CONTENT LABELING CLAIM (5O

K991743 · Shield Gloves Manufacturer (M) Sdn Bhd · LYY · Jun 23, 1999 · General Hospital

Device Facts

Record IDK991743
Device NameROYAL SHIELD NON-STERILE BUBBLE GUM SCENTED POWDER-FREE LATEX EXAMINATION GLOVE WITH PROTEIN CONTENT LABELING CLAIM (5O
ApplicantShield Gloves Manufacturer (M) Sdn Bhd
Product CodeLYY · General Hospital
Decision DateJun 23, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A medical glove to be worn on the hand of the health care and similar personnel to prevent contamination between health care personnel and patient.

Device Story

Royal Shield is a non-sterile, powder-free, bubble-gum scented latex examination glove. Designed for use by healthcare personnel in clinical settings to serve as a protective barrier against contamination. The device is manufactured from natural rubber latex and processed to meet specific residual protein and powder content limits. It is intended for over-the-counter use. The glove provides a physical barrier that reduces the risk of pathogen transmission between the clinician and the patient during examinations or procedures.

Clinical Evidence

Bench testing only. Testing included physical property evaluation (tensile strength, elongation), water leak testing (AQL 4.0%), residual powder content (ASTM D6124-97), residual protein levels (ASTM D5712-95), and biocompatibility (primary dermal irritation and skin sensitization).

Technological Characteristics

Material: Natural rubber latex. Dimensions: 242mm length; 0.08mm min thickness. Physical properties: Tensile strength 21.0 MPa (before aging), 16.0 MPa (after aging); Elongation 800% (before aging), 750% (after aging). Residual powder: 1.5-1.9 mg/glove. Residual protein: <50 µg/g. Standards: ASTM D3578-95, ASTM D6124-97, ASTM D5712-95. Non-sterile.

Indications for Use

Indicated for use by healthcare and similar personnel as a protective barrier to prevent cross-contamination between the wearer and the patient.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ ## APPENDIX H ## 510(k) SUMMARY ## SUMMARY OF THE SAFETY AND EFFECTIVENESS FOR ROYAL SHIELD POWDER FREE LATEX EXAMINATION GLOVES WITH BUBBLE GUM SCENT AND PROTEIN CONTENT LABELING ### Contact person : Ong Lay Mau This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. #### Device Information: Trade Name - ROYAL SHIELD POWDER FREE GREEN LATEX EXAM GLOVES Common Name - Exam gloves Classification Name - Patient examination glove (per 21 CFR 880.6250) Classification Information - Class I latex patient examination glove 80L YY, powder free and meeting all the requirements of ASTM-D3578-95 Standard Specification for Latex Examination Gloves for Medical Application. #### Device Description: Class I latex patient examination gloves 80LYY, powder free and meeting all the requirements of ASTM-D3578-95 Standard Specification for Latex Examination Gloves for Medical Application. #### Intended Use of Device: A medical glove to be worn on the hand of the health care and similar personnel to prevent contamination between health care personnel and patient. Technological Characteristics of Device: #### 1. Dimension | DIMENSION | ASTM D3578-95 | ROYAL SHIELD | |-------------------------------------|-----------------------------------------------------------------------|--------------------------------------------------------| | X-Small<br>Small<br>Medium<br>Large | 70 mm +/- 10 mm<br>80 mm +/- 10mm<br>95 mm +/- 10mm<br>111mm +/- 10mm | 70 - 75 mm<br>80 - 85 mm<br>90 - 97 mm<br>105 - 111 mm | | Length | 230 mm minimum<br>for all sizes | 242mm | | Thickness -<br>Finger<br>Palm | 0.08mm min<br>0.08mm min | 0.08 mm min<br>0.08 mm min | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a sequence of handwritten characters. The characters are 'K991743'. The characters are written in black ink on a white background. The handwriting is somewhat stylized. | | TENSILE STRENGTH | | ULTIMATE ELONGATION | | |-----------------------------|------------------|----------|---------------------|----------| | | ASTM-D3578-95 | SHIELD's | ASTM-D3578-95 | SHIELD's | | Before Aging | 14.0 Mpa min | 21.0 Mpa | 700 % | 800% | | After Aging<br>22 hrs @100C | 14.0 Mpa min | 16.0 Mpa | 500% | 750% | # 2. Physical Properties (ASTM-D3578-95 Standard Specification for Latex Exam Gloves) ## 3. Water Tight Test Using the FDA specified 1,000 ml water leak test, 80 pieces of each size of the gloves were tested and our results are as given below: | BATCH # | SIZE | SAMPLE SIZE | LEAK STATUS | NUMBER<br>LEAKED | |------------|---------|-------------|-------------|------------------| | 9903031019 | X-Small | 200 | Yes | 2 | | 9903222024 | Small | 200 | Yes | 2 | | 9904011027 | Medium | 200 | Yes | 1 | | 9903162023 | Large | 200 | Yes | 4 | The above figures are within the FDA/ ASTM requirements for latex exam gloves of 4.0% AQL. ## 4. Biocompatibility The test results below show that the gloves meet FDA biocompatibility requirements: BIOCOMPATIBILITY TESTS Primary Dermal Irritation Test Skin Sensitization Study RESULTS Not a primary irritant Not a sensitiser {2}------------------------------------------------ 991743 | TESTS | FDA INTERNAL<br>REQUIREMENT | SHIELD's | |--------------------------------------------------|-----------------------------|-----------------------------------------------| | Residual Powder<br>Content per<br>ASTM D 6124-97 | 2 mg/glove max | Range: 1.5-1.9mg/glove<br>Mean : 1.7 mg/glove | | Presence of Cornstarch | Negative | Negative | ## 5. Total Residual Powder Content & Presence of Cornstarch ## 6. Residual Protein Level | TESTS | FDA ALLOWABLE LEVEL | CLAIMED LEVEL | |----------------|---------------------|---------------------------------------| | ASTM D 5712-95 | < 50 µg/g | < 50 µg/g<br>Range: 27-33<br>Mean: 30 | #### Conclusion :- The data presented indicate that the Royal Shield Powder Free latex examination glove with Bubblegum scent - meets/exceeds ASTM- D3578-95 Standard Specifications For Latex Examination Glove, ﻨﺴ - 2. meets FDA pinhole requirements, - meets SHIELD's labeling claim of its being a powder free glove. న. - 4. meets the protein labeling claim level at <50 µg/g {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure and an eagle, with the human figure facing the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 3 1999 Shield Gloves Manufacturer (M) Sdn. Bhd. c/o Mr. E.J. Smith Smith Associates P.O. Box 4341 Crofton, Maryland 21114 K991743 Re : Trade Name: Royal Shield™ Non-Sterile Bubble Gum Scented Powder-Free Latex Examination Glove with Protein Content Labeling Claim (50 micrograms or less) Requlatory Class: I Product Code: LYY Dated: May 10, 1999 May 21, 1999 Received: Dear Mr. Smith: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. {4}------------------------------------------------ Page 2 - Mr. Smith In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". > Sincer ely Timo Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE STATEMENT Applicant: Shield Gloves Manufacturer (M) Sdn Bhd. 510K Number: Device Name: Royal Shield Powder Free Latex Examination Gloves With Bubble gum seent thotein content labeling claim (50 microgram or Indications For Use : This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patient. Concurrence of CDRH Office of Device Evaluation (ODE) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Prescription Use ... ... Per 21 CFR 801.109 OR Over-The-Counter........X Chien S. Lin (Division Sign-Off) Division of Dental, Infection Control, and General Hospital 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...