← Product Code [CAW](/productcode/CAW) · K991722

# DEVILBISS 5 LITER OXYGEN CONCENTRATOR (K991722)

_Sunrise Medical Hhg, Inc. · CAW · Nov 3, 1999 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K991722

## Device Facts

- **Applicant:** Sunrise Medical Hhg, Inc.
- **Product Code:** [CAW](/productcode/CAW.md)
- **Decision Date:** Nov 3, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5440
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The revised DeVilbiss 5 Liter Oxygen Concentrator is designed to provide supplemental low flow oxygen therapy for patients suffering from COPD, cardiovascular disease, and lung disorders. The oxygen concentrator is used in home type environments: homes, patient care facilities, etc.

## Device Story

Device is a 1-5 lpm pressure swing adsorption oxygen concentrator; uses molecular sieve technology to produce 93% +/-3% oxygen. Input is room air drawn via piston compressor; air passes through filters; pneumatic valve directs air into one of two sieve beds. Nitrogen adsorbed in pressurized bed; oxygen flows to patient. Simultaneously, other bed depressurizes to exhaust nitrogen. Revised model uses fixed-time cycle controlled by microprocessor; predicate used pressure-based cycle. Device used in homes/care facilities; operated by patients or caregivers. Output is continuous oxygen flow. High-pressure relief valve provides mechanical safety. Benefits include reliable supplemental oxygen delivery for respiratory conditions.

## Clinical Evidence

Bench testing only. Performance evaluated by measuring oxygen purity output at 1-5 lpm. Both subject and predicate devices met the 90% minimum purity requirement, with theoretical maximum of 95.5%. No clinical data provided.

## Technological Characteristics

Pressure swing adsorption oxygen concentrator; 1-5 lpm flow. Uses UOP Oxysiv-5 molecular sieve material. Single pneumatic valve system; fixed-time cycle control via microprocessor. Mechanical pressure relief valve for high-pressure safety. Dimensions/form factor not specified. Standalone operation. Electronics modified from predicate to support timed cycling; reduced alarm set (High Pressure, Long Cycle, No Flow removed).

## Regulatory Identification

A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).

## Predicate Devices

- DeVilbiss 5 Liter Oxygen Concentrator ([K953815](/device/K953815.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 6.0 510(k) Summary

Submitter's Name: Greg Clites

Address: Sunrise Medical HHG Inc Respiratory Products Division 100 DeVilbiss Drive P.O. Box 635 Somerset, PA 15501-0635

Phone #: 814-443-7441

Fax #: 814-443-7571

Contact Person: Greg Clites

Date Summary Prepared: May 14, 1999

Name of Device: DeVilbiss 5 Liter Oxygen Concentrator

Common Name: Oxygen Concentrator

Classification Name: Portable Oxygen Generator

Legally Marketed Device to which Equivalence is Claimed: DeVilbiss 5 Liter Oxygen Concentrator*

Manufacturer: Sunrise Medical HHG Inc Respiratory Products Division 100 DeVilbiss Drive P.O. Box 635 Somerset, PA 15501-0635

Predicate Device 510(k) Number: K953815

* The original FDA 510(k) submittal name for the predicate device was the "5 LPM Oxygen Concentrator".

### Device Description:

The revised DeVilbiss 5 Liter Oxygen Concentrator is a 1-5 lpm, pressure swing adsorption, oxygen concentrator based on molecular sieve technology. Room air is drawn into the concentrator via a piston style compressor. The air then passes through a series of filters that remove dust, bacteria, and other particulates. A pneumatic valve directs air into one of the two sieve beds. Nitrogen is adsorbed in the bed as the pressure increases while oxygen flows through, thereby producing an enriched oxygen product for the patient. Simultaneously in the other bed, nitrogen is desorbed as the pressure decreases and is exhausted into the atmosphere. A momentary intermediate pneumatic sequence ties the beds together with the exhaust blocked for an enhanced nitrogen purge. The cycle continues, providing a continuous flow of oxygen at a purity of 93% +/-3% to the patient. The concentrator is available with or without an oxygen monitoring device.

#### Intended Use of the Device:

The revised DeVilbiss 5 Liter Oxygen Concentrator is designed to provide supplemental low flow oxygen therapy for patients suffering from COPD, cardiovascular disease, and lung disorders. The oxygen concentrator is used in home type environments: homes, patient care facilities, etc.

{1}------------------------------------------------

#### Technological Characteristics:

The key technological characteristic differences between the revised DeVilbiss 5 Liter Oxygen Concentrator and the predicate DeVilbiss 5 Liter Oxygen Concentrator are summarized below:

The operating principle of the single pneumatic valve revised DeVilbiss 5 Liter Oxygen Concentrator is the same as the dual pneumatic valve predicate DeVilbiss 5 Liter Oxygen Concentrator, with the revised concentrator operates with a timed cycle and the predicate concentrator cycles based on the monitoring of system pressure. Operating principle:

Room air is drawn into the pressure swing adsorption molecular sieve technology oxygen concentrator by a compressor. Exiting the compressor, the air enters a four-way directional valve and is directed into one of two cylinders containing molecular sieve material. Nitrogen is adsorbed in the bed being pressurized, while oxygen flows through, thereby producing an enriched oxygen product for the patient. The other bed is simultaneously being depressurized, cleansing the bed of nitrogen. The process is alternated between the two beds assuring the patient a continuous supply of oxygen. The additional predicate concentrator valve serves to enhance the nitrogen purge with the UOP 5A-MG molecular sieve material to meet the required performance specification. The revised concentrator accomplishes this enhanced nitrogen purge of UOP Oxysiv-5 molecular sieve material via a momentary intermediate valve position, enabling the revised concentrator to meet the required performance specification. Both UOP 5A-MG and Oxysiv-5 are non-hazardous and non-toxic.

The predicate DeVilbiss 5 Liter Oxygen Concentrator cycles based on a fixed pressure setting detected in the oxygen accumulator tank. When a pc board mounted pressure transducer detects the set pressure in the accumulator tank, the microprocessor shifts the pneumatic valving. The revised concentrator cycles based on a fixed cycle time. The microprocessor shifts the pneumatic valving according to this fixed time.

While the method of shifting the valve to cycle the system is different, the operating principle of the pressure swing adsorption concentrator is the same for both the revised DeVilbiss 5 Liter Oxygen Concentrator and the predicate device. The oxygen purity requirement to the patient is satisfied for both products, hence there is no effect on safety and effectiveness.

The revised and predicate concentrator electronics have been designed with the same microprocessor. Since the functionality of the electronics has been modified for the revised concentrator with implementation of the timed cycle and an accompanying decrease in alarms, the processor is different. With the exception of the No Power and the Low Pressure alarms, the alarm features (High Pressure, Long Cycle, and No Flow) have been removed from the revised concentrator. New software documentation will be generated accordingly. A high pressure indication is most directly tied to the safety of the patient. This will mechanically be maintained via the use of a pressure relief valve on the revised concentrator. The Long and Short Cycle alarms are not necessary on the revised concentrator since it is controlled by a fixed cycle time The High Pressure, Long Cycle, and Short Cycle alarms shut the predicate concentrator down in the event that one of these conditions was detected by the software, in an effort to protect the concentrator from deterioration, but are not directly tied to the oxygen performance. Hence, the modified software and accompanying reduction in the patient alert alarms from the revised DeVilbiss 5 Liter Oxygen Concentrator do not present new issues of concentrator safety or effectiveness.

#### Performance Data:

The key performance descriptor for the oxygen concentrator devices is the oxygen purity performance at 1-5 lpm of oxygen output. The theoretical maximum oxygen purity which is capable from a molecular sieve concentrator is 95.5%, and the commonly accepted minimum level in the concentrator market is 90%. Based on tests performed, both the revised DeVilbiss 5 Liter Oxygen Concentrator and the predicate concentrator meet this requirement.

Conclusion: The oxygen purity levels for the revised and predicate concentrators are equivalent, hence there is no effect on safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the words "Public Health Service" in bold, black font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The text appears to be extracted from a document or sign.

Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often used as a symbol of medicine. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.

NOV -3 1999 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Greq Clites Sunrise Medical HHG Inc. Respiratory Products Division 100 DeVilbiss Drive P.O. Box 635 Somerset, PA 15501-0635

Re: K991722 DeVilbiss 5 Liter Oxygen Concentrator Regulatory Class: II (two) Product Code: 73 CAW Dated: August 31, 1999 Received: September 1, 1999

Dear Mr. Clites:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Greg Clites

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

feccions. Hwacteristanse for,
Ceila M. Witten, Ph.D., M.D.
Acting Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

510(k) Number (if known): __ K991722

Device Name: DeVilbiss 5 Liter Oxygen Concentrator

Indications For Use:

The DeVilbiss 5 Liter Oxygen Concentrator is intended for use as an oxygen concentrator to provide supplemental low flow oxygen therapy in the home, nursing homes, patient care facilities, etc.

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

JoAu Westerman

(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number K971722

Prescription Use x (Per 21 CFR 801.109) Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K991722](https://fda-staging.innolitics.com/device/K991722)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
