← Product Code [MNI](/productcode/MNI) · K991721

# EBI OMEGA 21 SYSTEM (K991721)

_Ebi Electro Biology · MNI · Aug 13, 1999 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K991721

## Device Facts

- **Applicant:** Ebi Electro Biology
- **Product Code:** [MNI](/productcode/MNI.md)
- **Decision Date:** Aug 13, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The EBI Omega 21™ System is a spinal fixation device for pedicle screw fixation and a nonpedicle hook and sacral/iliac screw fixation system of the noncervical spine. When used as a pedicle screw fixation system, in the non-cervical spine of skeletally mature patients, the System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: Degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocations, scoliosis, kyphosis, spinal tumor, and failed previous fusion (psuedarthrosis). In addition, when used as a pedicle screw fixation system in skeletally mature patients, it is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass. When used as a posterior hook and sacral/iliac screw fixation system, the levels of attachment are the lumbar and thoracic spine, and screw fixation limited to the sacrum and ilium. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Psuedarthrosis; scoliosis; spondylolisthesis, fracture; previous failed fusion; or tumor resection. When used as an anterior fixation system, the levels of attachment are the anterolateral vertebral bodies of the lumbar and thoracic spine. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Psuedarthrosis; stenosis; scoliosis; spondylolisthesis, fracture; previous failed fusion; or tumor resection.

## Device Story

Spinal fixation system comprising rods, screws, and couplers; used to immobilize and stabilize spinal segments as adjunct to fusion. Operates via mechanical fixation to thoracic, lumbar, and sacral spine. Surgeon-implanted in clinical setting. Provides structural support to treat instabilities/deformities; facilitates bone fusion. Submission expands screw diameter range (4.75mm-8.5mm).

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Components manufactured from Ti-6Al-4V ELI per ASTM F136. Mechanical spinal fixation system consisting of rods, screws, and couplers. No software or electronic components.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- EBI Omega 21TM System ([K973683](/device/K973683.md))
- EBI Omega 21TM System ([K990303](/device/K990303.md))
- EBI SpineLink™ System
- Synthes Posterior Universal Spinal System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) Summary of Safety & Effectiveness

This 510(k) Summary of Safety and Effectiveness for the EBI Omega 21TM System is provided as required per Section 513(D)(3) of the Federal Food, Drug and Cosmetic Act.

Jon Caparotta, Manager Regulatory Affairs Sponsor: 1. Electro-Biology, Inc. 100 Interpace Parkway Parsippany, NJ 07054

Contact Person: Jonas Wilf Telephone: (973) 299-9300

Date prepared: May 19, 1999

- 2. Proprietary Name: Common Name: Classification Names:
EBI Omega 21 TM System Spinal Fixation Device Spondylolisthesis Spinal Fixation Device System Spinal Intervertebral Body Fixation Orthosis Spinal Interlaminal Fixation Orthosis

- 3. Predicate or legally marketed devices that are substantially equivalent:
	- · EBI Omega 21TM System
	- · EBI SpineLink™ System
	- · Synthes Posterior Universal Spinal System
- Description of the device: The EBI Omega 21™ System is a spinal fixation device 4. that uses rods, screws, and couplers. This submission is for expanding the range of screw sizes to 4.75mm-8.5mm diameter.
- 5. Intended Use: The EBI Omega 21™ System has not changed the Indications for Use or fundamental scientific technology of the previously cleared EBI Omega 21™ System (K973683 and K990303). The EBI Omega 21™ System is a spinal fixation device for pedicle screw fixation and a nonpedicle hook and sacral/iliac screw fixation system of the noncervical spine.

When used as a pedicle screw fixation system, in the non-cervical spine of skeletally mature patients, the System is intended to provide immobilization of spinal segments, as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: Degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocations, scoliosis, kyphosis, spinal tumor, and failed previous fusion (psuedarthrosis).

In addition, when used as a pedicle screw fixation system in skeletally mature patients, it is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.

{1}------------------------------------------------

When used as a posterior hook and sacral/iliac screw fixation system, the levels of attachment are the lumbar and thoracic spine, and screw fixation limited to the sacrum and ilium. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Psuedarthrosis; stenosis; scoliosis; spondylolisthesis, fracture; previous failed fusion; or tumor resection.

When used as an anterior fixation system, the levels of attachment are the anterolateral vertebral bodies of the lumbar and thoracic spine. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Psuedarthrosis; stenosis; scoliosis; spondylolisthesis, fracture; previous failed fusion; or tumor resection.

- 6. Materials: The components of the system are manufactured from Ti-6Al-4V ELI per ASTM F136.
- 7. Comparison of the technological characteristics of the device to predicate devices: There are no significant differences between the EBI Omega 21™ System and other currently marketed spinal systems. It is substantially equivalent* to the predicate devices in regards to intended use, materials and function

* Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate to whether the product can be lawfully market approval or reclassification and is not intended to be interpreted as an admission or any other type of evidence in patent litigation. [Establishment Registation and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.))

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 13 1999

Mr. Jon Caparotta Manager, Regulatory Affairs Electro Biology, Inc. 100 Interpace Parkway Parsippany, New Jersey 07054

Re: K991721

Trade Name: Omega 21TM System Regulatory Class: II Product Code: MNI, MNH, KWP and KWQ Dated: May 19, 1999 Received: May 20, 1999

Dear Mr. Capparotta:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General Regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2 - Mr. Jon Caparotta

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.

Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{4}------------------------------------------------

## Statement of Indications for Use:

The EBI Omega 21™ System is a spinal fixation device for pedicle screw fixation and a nonpedicle hook and sacral/iliac screw fixation system of the noncervical spine.

When used as a pedicle screw fixation system, in the non-cervical spine of skeletally mature patients, the System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: Degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocations, scoliosis, kyphosis, spinal tumor, and failed previous fusion (psuedarthrosis).

In addition, when used as a pedicle screw fixation system in skeletally mature patients, it is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.

When used as a posterior hook and sacral/iliac screw fixation system, the levels of attachment are the lumbar and thoracic spine, and screw fixation limited to the sacrum and ilium. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Psuedarthrosis; scoliosis; spondylolisthesis, fracture; previous failed fusion; or tumor resection.

When used as an anterior fixation system, the levels of attachment are the anterolateral vertebral bodies of the lumbar and thoracic spine. The System is intended for the treatment of degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Psuedarthrosis; stenosis; scoliosis; spondylolisthesis, fracture; previous failed fusion; or tumor resection.

Prescription Use
(Per 21 CFR 801.109)

Piseula

(Division S Division of General Restorative Devic 510(k) Number

6

---

**Source:** [https://fda-staging.innolitics.com/device/K991721](https://fda-staging.innolitics.com/device/K991721)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com