K991609 · Dermaphylyx, Inc. · KMF · Jul 20, 1999 · General Hospital
Device Facts
Record ID
K991609
Device Name
DERMAPHYLYX SUPER ABSORBENT FIBER DRESSING
Applicant
Dermaphylyx, Inc.
Product Code
KMF · General Hospital
Decision Date
Jul 20, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5090
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
Dermaphylyx Absorbent Fiber Wound Dressings are soft, non-woven, and absorptive. They are intended for use in the management of a variety of partial and full-thickness wounds. The dressings are additionally intended to help control minor bleeding by application of pressure to the wound. The following Indications are for Prescription Use or under the direction of a health care professional: Venous ulcers, Diabetic ulcers, Pressure ulcers, Arterial ulcers, Superficial burns, Abrasions and lacerations, Donor sites, Incisions. The following Indications are for Over-the-Counter Use: Abrasions, Minor Burns, Minor Cuts, Minor Lacerations.
Device Story
Dermaphylyx Absorbent Fiber Wound Dressing is a soft, non-woven, absorptive dressing available in felt or rope configurations. Designed for management of partial and full-thickness wounds; absorbs wound exudate to form a fiber mat, creating a moist healing environment and facilitating trauma-free removal. Used in professional clinical settings and by patients for OTC applications (abrasions, minor burns, minor cuts, minor lacerations). Applied topically to wounds; provides pressure to control minor bleeding. Benefits include exudate management and support of moist wound healing.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use compared to legally marketed predicate devices.
Technological Characteristics
Soft, non-woven, absorptive fiber dressing; felt or rope configuration. Principle of operation involves absorption of wound exudate to form a fiber mat. No energy source or software components.
Indications for Use
Indicated for management of partial and full-thickness wounds including venous, diabetic, pressure, and arterial ulcers, superficial burns, abrasions, lacerations, donor sites, and incisions. Also indicated for control of minor bleeding. Suitable for professional/prescription use and OTC use for abrasions, minor burns, minor cuts, and minor lacerations.
Regulatory Classification
Identification
A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.
{0}------------------------------------------------
JUL 20 1999
991609
## 510(k) Dermaphylyx Absorbent Fiber Wound Dressing K991609. Dermaphylyx, Inc.
### 510(k) Summary
| Proprietary Name: | Dermaphylyx Absorbent Fiber Wound Dressing |
|----------------------|-------------------------------------------------------------------------------------------------------------------|
| Common Name: | Dressing |
| Classification: | Unclassified |
| Submitter's Details: | Dermaphylyx, Inc.<br>78-E, Olympia Avenue,<br>Woburn, MA 01801-2057<br>Tel: (781) 933-4772<br>Fax: (781) 933-3933 |
#### Description:
Dermaphylyx Absorbent Fiber Wound Dressings are soft, non-woven, and absorptive. The Dressings consist of Absorbent fibers fabricated into a felt or rope configuration.
Dermaphylyx Absorbent Fiber Wound Dressings are engineered to absorb substantial quantities of wound exudate. During the absorption of exudate the dressing forms a fiber mat which provides an environment conducive to moist wound healing. The fiber mat facilitates traums free removal from the wound.
Dermaphylyx Absorbent Fiber Wound Dressings are intended for use in the management of partial and full- thickness wounds in both a professional and OTC environment. They may be used on the following wounds:
| Venous ulcers | Superficial burns |
|-----------------|---------------------------|
| Diabetic ulcers | Abrasions and lacerations |
| Pressure ulcers | Donor sites |
| Arterial ulcers | Incisions |
Dermaphylyx Absorbent Fiber Wound Dressings may also be used to control minor bleeding.
Over the Counter applications include abrasions, minor lacerations, as well as minor burns and the control of minor bleeding by application of pressure to the wound.
Dermaphylyx Absorbent Fiber Wound Dressings are substantially equivalent to Aquace!™ Hydrofiber™ Wound Dressings (Convated), and Kaltostat@ Wound Dressings (Calgon Vestal Laboratories). These devices are absorptive Wound Dressings, manufactured from Absorbent fibers. They are intended for use in the management of a wide variety of wounds.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with three human profiles embedded within the eagle's body.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 20 1999
Andrew M. Reed, Ph.D. Principal Dermaphylyx, Inc. 12106 West 75th Lane Arvada, Colorado 80005-5306
Re: K991609 Trade Name: Absorbent Fiber Wound Dressing Regulatory Class: Unclassified Product Code: KMF Dated: May 6, 1999 Received: May 10, 1999
Dear Dr. Reed:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 - Andrew M. Reed, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Dermaphylyx Absorbent Fiber Wound Dressing K991609, Dermaphylyx, Inc.
# 991609
Page 1 of 1
## PREMARKET NOTIFICATION INDICATIONS FOR USE STATEMENT
| 510(k) Number: | K991609 |
|----------------|---------|
| | Dermanh |
aphylyx, Inc.
Device Name:
Dermaphylyx Absorbent Fiber Wound Dressing
Indications for Use:
Dermaphylyx Absorbent Fiber Wound Dressings are soft, non-woven, and absorptive. They are intended for use in the management of a variety of partial and full-thickness wounds. The dressings are additionally intended to help control minor bleeding by application of pressure to the wound,
The following Indications are for Prescription Use or under the direction of a health care professional:
Venous ulcers Diabetic ulcers Pressure ulcers Arterial ulcers
Superficial burns Abrasions and lacerations Donor sites Incisions
The following Indications are for Over-the-Counter Use:
Abrasions Minor Burns Minor Cuts Minor Lacerations
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evelustion (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use X
२०
Tóledo
(Division Sign-Off) Division of General Re 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.