CELSIOR COLD FLUSH, STORAGE AND TRANSPORT SOLUTION FOR HEARTS
K991594 · Sangstat Medical Corp. · MSB · Aug 5, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K991594
Device Name
CELSIOR COLD FLUSH, STORAGE AND TRANSPORT SOLUTION FOR HEARTS
Applicant
Sangstat Medical Corp.
Product Code
MSB · Gastroenterology, Urology
Decision Date
Aug 5, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5880
Device Class
Class 2
Attributes
Therapeutic, Expedited Review
Indications for Use
Celsior™ is intended for flushing and cold storage of a heart at the time of its removal from a donor in preparation for storage, transportation, and eventual transplantation into a recipient.
Device Story
Celsior is a sterile, non-pyrogenic, extracellular solution used for hypothermic flushing, storage, and transport of donor hearts. The solution is supplied in 1-liter ethylene-vinyl acetate (EVA) bags, which are sterilized via ethylene oxide and packaged with an oxygen-absorbent sachet. The solution is used by transplant clinicians during the organ procurement process. It functions by maintaining the heart in a hypothermic state (2° to 8°C) to preserve tissue viability during the interval between donor removal and recipient transplantation. The solution's high buffering capacity and specific electrolyte composition are designed to support cardiac tissue during cold ischemia. Clinical use involves flushing the donor heart with the solution immediately upon removal, followed by storage in the solution during transport to the transplant site.
Clinical Evidence
Clinical evidence includes a multicenter, randomized, controlled, open-label trial (IDE G970052) comparing Celsior to standard solutions (including ViaSpan). Primary endpoint: 7-day patient survival (comparable). Secondary endpoints: 30-day patient and graft survival (comparable). Results showed significantly fewer cardiac-related serious adverse events in the Celsior group within 30 days post-transplant. Preclinical studies in rabbit and canine models demonstrated superior performance of Celsior-treated hearts compared to ViaSpan. Biocompatibility, sterility, and stability testing were also performed.
Technological Characteristics
Extracellular solution; pH 7.3; osmolarity 320-360 mOsmol; viscosity 1.15 cSt. Contains mannitol (impermeant) and histidine (buffer). Supplied in 1L ethylene-vinyl acetate (EVA) bags. Sterilized via ethylene oxide. Requires refrigerated storage (2° to 8°C).
Indications for Use
Indicated for flushing, storage, and transportation of donor hearts in preparation for transplantation into a recipient.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}------------------------------------------------
SangStat Medical Corporation
Celsior™ 510(k)
K991694 p.1/4
### 510(k) Summary of Safety and Effectiveness Information 9.0
Date Prepared: July 30, 1999
### I. Sponsor
- A. Sponsor Name
SangStat Medical Corporation 6300 Dumbarton Circle Fremont, CA 94555 Telephone: (510) 789-4300 (510) 789-4205 Fax:
## B. Official Correspondent
Mark D. Tolpin, M.D. Senior Vice President Worldwide Clinical Research & Regulatory Affairs SangStat Medical Corporation 6300 Dumbarton Circle Fremont, CA 94555 Telephone: (510) 789-4533 Fax: (510) 789-4205
### II. System Identification
- A. Proprietary Product Name
Celsior™ Cold Storage Solution
- B. Common or Usual Product Name
Cold Storage Solution
# C. Product Classification
Class II, Panel Number 78-KDL
{1}------------------------------------------------
K991594 p.2/4
## III. Predicate Device
## A. Name
ViaSpan® (Belzer UW-CSS), Du Pont Pharmaceuticals (Wilmington, Delaware) [510(k) Notification Number K900492 and K883782], Class Code with Panel Number 78-KDL.
# B. Indications for Use
The solution is intended for flushing and cold storage for organs including kidney, liver, and pancreas at the time of their removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.
## C. Device Description
ViaSpan®, (Belzer UW-CSS) Cold Storage Solution (Du Pont Pharmaceuticals, Wilmington, Delaware), is a clear to light yellow, sterile, non-pyrogenic solution used for flushing and storage of human organs. This solution has an approximate calculated osmolarity of 320 mOsm, a sodium concentration of 29 mEq/L, a potassium concentration of 125 mEq/L, and a pH of roughly 7.4 at room The composition of ViaSpan is thus consistent with that of an temperature. intracellular solution.
### Celsior™ Device Information IV.
# A. Indications for Use
Celsior™ is intended to be used for flushing, storage, and transportation of hearts in preparation for eventual transplantation into a recipient.
# B. Device Description
Celsior (SangStat Medical Corporation, Fremont, CA) is a clear to slightly yellow, sterile, non-pyrogenic, extracellular solution for hypothermic flushing and storage of hearts. The solution is slightly acidic (pH 7.3 + 0.10), slightly hypertonic (osmolarity 320-360 mOsmol) with low viscosity (1.15 cSt), and has a high buffering capacity (acidic approximately 11 mmol, alkaline approximately 7 mmol). The composition of Celsior is thus consistent with that of an extracellular solution.
Celsior is filled into 1 liter ethylene-vinyl acetate copolymer (EVA) bags. The EVA bags are sterilized (ethylene oxide) and aseptically filled up to 1,000 mL in a sterile area. After filling, each EVA bag is enveloped by an aluminum protected bag containing an oxygen absorbent sachet. This product must be stored under refrigerated conditions at 2° to 8°C (36° to 46°F) until it is used.
{2}------------------------------------------------
K991594 p.3/4
Celsior™ 510(k)
#### V. Substantial Equivalence
#### A. Indications for Use
ViaSpan and Celsior have the same intended use. Both solutions are intended for the flushing and cold storage of an organ at the time of its removal from a donor in preparation for storage, transportation, and eventual transplantation into a recipient.
#### B. Technological Characteristics
Celsior and ViaSpan are clear to light yellow, sterile, non-pyrogenic solutions. Celsion's chemical composition has a number of similarities to that of ViaSpan. However, the potassium content of Celsior is more similar to that of an extracellular solution, whereas the potassium content of ViaSpan is more consistent with that of an intracellular solution. Both Celsior and ViaSpan contain impermeants, buffers, reduced glutathione, electrolytes, and water for injection. The impermeant in Celsior is mannitol and the buffer is histidine, while the impermeant in ViaSpan is pentafraction and the buffers are potassium hydroxide and sodium hydroxide.
#### C. Performance Data
To demonstrate that Celsior is substantially equivalent to ViaSpan for its intended uses, SangStat conducted three clinical studies and several nonclinical studies using Celsior. The study conducted in the U.S. (IDE file number G970052) was a multicenter, randomized, controlled, open-label clinical trial comparing Celsior to standard preservation solutions, including ViaSpan. The primary efficacy endpoint of seven (7) day patient survival was comparable between groups as were the secondary endpoints of 30 day patient and graft survival. In the first 30 days posttransplant, significantly fewer subjects receiving Celsior-treated hearts had one or more cardiac-related serious adverse events based on one-sided 95% confidence interval analysis. Clinical studies demonstrated that Celsior and ViaSpan had similar adverse event profiles and that adverse events using Celsior were also similar to that for other hospital solutions currently used to prepare hearts for transplant.
Preclinical studies conducted in isolated rabbit hearts and mongrel canines demonstrate that Celsior-treated hearts performed better than ViaSpan-treated hearts. Preclinical studies also demonstrate the biocompatibility (e.g., cyotoxicity, intracutaneous, in vitro hemolysis, in vivo hemolysis, osmotic fragility), sterility (ethylene oxide, ethylene chlorohydrin, and ethylene glycol), and stability of Celsior.
{3}------------------------------------------------
K991594 p.4/4
## VI. Conclusion
ViaSpan and Celsior have the same intended use. Data contained in the 510(k) demonstrate that Celsior is substantially equivalent.
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 5 1999
Mark D. Tolpin, M.D Senior Vice President Worldwide Clinical Research and Requiatory Affairs SangStat Medical Corporation 1505 Adams Drive Menlo Park, CA 94025
Re: K991594. Celsior™ Cold Flush, Storage and Transport Solution for Hearts Dated: May 6, 1999 Received: Mav 7. 1999 Regulatory Class: II 21 CFR §876.5880/Procode: 78 MSB
Dear Dr. Tolpin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations,
This letter will allow you to begin marketing your device as described in your 510(t) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), piease contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| 510(k) Number (if known): | Subject of this Submission |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Celsior™ Cold Flush, Storage and Transport Solution<br>for Hearts |
| Indications for Use: | Celsior™ is intended for flushing and cold storage of a<br>heart at the time of its removal from a donor in<br>preparation for storage, transportation, and eventual<br>transplantation into a recipient. |
(Please Do Not Write Below This Line - Continue On Another Page If Needed Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use # (Per 21 CFR §801.109)
Or
Over-The-Counter Use (Optional Format 1-2-96)
David C. Digman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, Division of Reproductive, and Radiological Devices
and Radiological Devices
5100k) Number 991596 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.