K991588 · (Ems Sa) Electro Medical Systems · HSZ · Jul 21, 1999 · General, Plastic Surgery
Device Facts
Record ID
K991588
Device Name
EMS SWISS ORTHOCLAST
Applicant
(Ems Sa) Electro Medical Systems
Product Code
HSZ · General, Plastic Surgery
Decision Date
Jul 21, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4820
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The EMS Swiss OrthoClast® is used for the removal of cement during implant revision surgery.
Device Story
The Swiss OrthoClast® is a pneumatic-powered orthopedic surgical instrument used for bone cement removal during implant revision surgery. The system comprises a pressure regulator/control unit, pneumatic footswitch, handpiece, various chisels, a hollow drill/extractor for cement tips, and an Orthoscope. Operation involves a pressure pulse from the control unit driving an internal projectile within the handpiece to strike the chisel; this generates a shock wave causing high-speed, short-excursion tip movement to fragment bone cement. The Orthoscope provides direct visualization of body cavities (e.g., femoral cavity). The device is operated by a surgeon in a clinical setting, requiring only one hand to operate, which allows for precise guidance of the chisel tip. The intermittent pulse mechanism prevents heat build-up at the fragmentation site. The device benefits the patient by facilitating efficient cement removal during revision procedures.
Clinical Evidence
Bench testing only. Testing included risk analysis, maximum chisel tip excursion and velocity measurements, handpiece and chisel longevity testing, in vitro cement fragmentation efficiency, and Orthoscope temperature elevation monitoring. Results confirmed design objectives were met.
Technological Characteristics
Pneumatic-powered orthopedic surgical instrument. System includes pressure regulator, pneumatic footswitch, handpiece, and various steel chisels. Operates via ballistic principle: pneumatic pressure pulse drives internal projectile to strike chisel, creating shock waves. Includes fiberoptic-style Orthoscope for visualization. No software or electronic algorithms; purely mechanical/pneumatic operation.
Indications for Use
Indicated for patients undergoing implant revision surgery requiring the removal of bone cement.
Regulatory Classification
Identification
Surgical instrument motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. Accessories or attachments may include a bur, chisel (osteotome), dermabrasion brush, dermatome, drill bit, hammerhead, pin driver, and saw blade.
{0}------------------------------------------------
91588
# 510(k) Summary . for Swiss OrthoClast®
#### 1. SPONSOR
EMS SA Ch. de la Vuarpillière 31 CH-1260 Nyon Switzerland
| Contact Person: | Gianni Campana |
|-----------------|-------------------|
| Telephone: | 011 41 22 9944770 |
May 5, 1999 Date Prepared:
#### 2. DEVICE NAME
Swiss OrthoClast® Proprietary Name: Powered orthopedic surgical instrument Common/Usual Name: Instrument, surgical, orthopedic, pneumatic powered and Classification Name: accessories
#### 3. PREDICATE DEVICE
Ceboplane® (K933933)
#### 4. INTENDED USE
The Swiss OrthoClast® is used for the removal of bone cement during implant revision surgery.
#### 5. DEVICE DESCRIPTION
The EMS Swiss OrthoClast® system consists of a pressure regulator/control unit, pneumatic footswitch, handpiece, chisels of various sizes and shapes, a hollow drill and extractor for removal of the cement tip, and the Orthoscope. The Swiss OrthoClast® uses simple pneumatic and ballistic principles to fragment bone cement. A pressure pulse generated by the pressure regulator/control unit causes a moveable projectile in the handpiece to be driven forward to strike the proximal end of the chisel. The impact generates a shock wave along the chisel, which leads to a small, high-speed excursion of the chisel tip. The distal end of the chisel is
{1}------------------------------------------------
held in contact with the targeted cement. Repetitive shock waves result in fragmentation of the bone cement. A short tip excursion period followed by a longer rest period prevents the build-up of heat in the fragmentation site. Direct vision into body cavities, such as the femoral cavity, is facilitated using the Orthoscope.
### 6. PERFORMANCE TESTING
A comprehensive risk analysis was conducted to identify potential hazards associated with the use and misuse of the Swiss OrthoClast®. Verification tests conducted as a result of this risk analysis include the following
- Maximum excursion and velocity of the Swiss OrthoClast® chisel tips ●
- Handpiece and chisel longevity
- In vitro cement fragmentation efficiency .
- . Orthoscope temperature elevation
The results of these tests confirmed that the Swiss OrthoClast® met the design objectives as identified by the risk analysis.
## 6. BASIS FOR SUBSTANTIAL EQUIVALENCE
The Swiss OrthoClast® has the same intended use as the Ceboplane®, that is, removal of bone cement during implant revision surgery. The Ceboplane® is a pneumatic powered reciprocating handpiece that accepts a wide variety of osteotome chisels. A fiberoptic illuminator mounts directly on the predicate handpiece to enhance visualization for working in deep cavities. Like the Swiss OrthoClast®, the Ceboplane® uses an internal piston hammer to deliver a highenergy, short-excursion impact to the desired location under the guidance of the surgeon using either single or multiple pulse mode. Both the OrthoClast® and the Ceboplane® require only one hand to operate, unlike the traditional osteotome and mallet which require two hands, allowing the surgeon to focus on the impact sight and precisely guide the chisel tip to its optimum position.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
JUL 2 1 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Shelia Hemeon-Heyer, Esq., RAC Senior Staff Consultant MDCI Representing EMS SA 49 Plain Street North Attleboro, Massachusetts 02760
Re: K991588 Trade Name: Swiss OrthoClast® Regulatory Class: II Product Code: HSZ, LZV, and GET Dated: May 5, 1999 Received: May 7, 1999
Dear Ms. Hemeon-Heyer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above, and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General Regulation (21 CFR Part 820), and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{3}------------------------------------------------
# Page 2 - Ms. Shelia Hemeon-Heyer
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a scattering of small, dark, irregularly shaped objects against a bright white background. The objects are distributed unevenly across the frame, with a higher concentration in the upper left quadrant. The objects appear to be randomly oriented and vary slightly in size and shape.
510(k) Number (if known): _ K 991588
Device Name: EMS Swiss OrthoClast®
Indications For Use:
The EMS Swiss OrthoClast® is used for the removal of cement during implant revision surgery.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sian-Off) Division of General Restorative Devi Kgg1598 510(k) Number
Prescription Use X (Per 21 CFR 801.109) OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.