K991571 · Alpha-Omega Services, Inc. · KXK · Feb 22, 2000 · Radiology
Device Facts
Record ID
K991571
Device Name
10 CURIE IRDIUM 192, CSN00100-192
Applicant
Alpha-Omega Services, Inc.
Product Code
KXK · Radiology
Decision Date
Feb 22, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
A high specific activity radionuclide brachytherapy source to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy. Device is the active source for a Nucletron Microselection Device. Prescription Device. Federal Law (US) restricts this device to sale by or on the order of a physician.
Device Story
Device is a hermetically sealed Iridium-192 radionuclide brachytherapy source (Model CSN0010-192) for use with Nucletron Microselection HDR afterloader systems. Input is radioactive Iridium-192 seed; output is nuclear radiation for therapeutic treatment. Device consists of stainless steel 316LVM capsule (1.1mm diameter) welded to 7x7 stranded stainless steel cable (2000mm length). Used in clinical oncology settings; operated by trained, licensed personnel (physicians/technicians) within established irradiation equipment installations. Device delivers radiation to target tissue/cavity; clinical decision-making relies on physician-directed treatment planning. Benefits include precise delivery of radiation therapy for cancer treatment.
Clinical Evidence
Bench testing only. No clinical data provided. Comparison based on physical specifications, material biocompatibility, and performance testing (50,000 transfer cycles).
Indicated for patients requiring brachytherapy; source placed on body surface, in body cavity, or in tissue for radiation therapy. Used as active source for Nucletron Microselection HDR afterloader. Prescription only.
Regulatory Classification
Identification
A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
Special Controls
*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Section 2 Summary 510(k)
The following is a Summary of the Alpha-Omega Services, Inc., 10 Curie IR-92 Brachytherapy Source for Nucletron Microselection device.
| Applicant | Alpha-Omega Services, Inc.<br>9156 Rose Street<br>Bellflower, California 90706<br>(800) 346-7894<br>(562) 804-0610 Fax |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Manufacturing Facility | Alpha-Omega Services, Inc.<br>1282 Big Woods and Starks Rd.<br>Edgerly, LA 70668<br>(318) 589-5720 |
| Contact Person | Bruce Hedger, President and RSO |
| Classification Name | Radionuclide Brachytherapy Source. |
| Common/Usual Name | 10 Curie IR-92 Brachytherapy Source for Nucletron<br>Afterloader device. |
| Proprietary Name | 10 Curie Iridium 192 Source, Model CSN0010-192. |
| Establishment Registration Number | 2022694 California, Louisiana registration has been sent in. |
| Classification | 21 CFR 892.5730, Class II, Product Code: JAQ |
| Special Controls | Product will comply with the nuclear regulatory requirements<br>for the State of Louisiana NRC and US NRC. |
| Substantial Equivalence | K864210 Nucletron Microselection Source. |
## General Information
The contents of this premarket notification summary will demonstrate the substantial equivalence of the subject device, 10 Curie IR-92 Brachytherapy Source for Nucletron Afterloader device. The substantial equivalence will be based on the following features of the device:
- . Indications
- Physical Size .
- Radioisotope .
- . Radiation activity
The Alpha-Omega Services, Inc., 10 Curie Iridium 192 Source, Model CSN0010-192 device is a hermetically sealed radionuclide brachytherapy source indicated for, to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
The source is made of a stainless steel capsule with an outer diameter of approximately 1.1mm. A hole is drilled through the center of the capsule to a depth of the ir 192 seed. The closed end of the capsule is welded to a 1.1 mm diameter 7 X 7 stranded stainless steel cable. On the opposite end, a solid connector is welded in the same manner to the free end of the cable A Ir 192 seed is inserted into the open end of the capsule. The Ir 192 seed is secured by welding the open end of the capsule. The post-weld length of capsule is approximately 4.5 mm in length
{1}------------------------------------------------
and overall length of the finished source cable assembly is approximately 2000 mm. These physical specifications are the source for the Nucletron Microselection HDR device.
The following table compares indications statement for Alpha-Omega Services, Inc., 10 Curie IR-92 Brachytherapy Source for Nucletron Microselection HDR device and the predicate device.
| 10 Curie IR-92 Brachytherapy Source for | IR-192 (K864210), Nucletron Microselection |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nucletron Afterloader device | Predicate Device |
| A high specific activity radionuclide<br>brachytherapy source to be placed onto a body<br>surface or into a body cavity or tissue as a<br>source of nuclear radiation for therapy. Device<br>is the active source for a Nucletron<br>Microselection Device. | A high specific activity radionuclide<br>brachytherapy source to be placed onto a body<br>surface or into a body cavity or tissue as a<br>source of nuclear radiation for therapy. Device<br>is the active source for an Nucletron<br>Microselection |
Based on the intent of the indications statement for the subject device, the Alpha-Omega Services, Inc., 10 Curie IR-92 Brachytherapy Source for Nucletron Afterloader device, is substantially equivalent to the predicate device with respect to its indications.
The following table compares the physical size. materials of construction, and the radioisotope for the subject device and predicate device.
| Feature Description | 10 Curie Iridium 192 Source, Model<br>CSN0010-192 for Nucletron<br>Microselection device | IIR-192 (K864210), Nucletron<br>Microselection<br>Predicate Device |
|-----------------------------|------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Diameter | 1.1 mm | 1.1 mm |
| Length | 2000 mm ± 5% | 2000 mm ± 5% |
| Cable Type | 7X7 Stranded | 7X7 Stranded |
| Material | Stainless steel 316LVM | Stainless steel 316LVM |
| Seal Method | Laser / Tig Welding | Laser Welding |
| Radioisotope | IR - 192 Seed 3.5 mm | IR - 192 Seed 3.5 mm |
| Tested transfers cycles | 50,000 | 20,000 |
| Recommended cycles | 5,000 | 5,000 |
| Radioactive half life | 75 days | 75 days |
| Apparent Activity<br>Levels | 9 - 13 Ci at time of shipping | 9 - 13 Ci at time of shipping |
## LIMITATIONS AND/ OR OTHER CONSIDERATIONS OF USE
- The source cable shall be distributed in the United States to persons specifically 1. licensed by the United States NRC or an agreement state.
- 2. The source cable shall be distributed internationally to users with approval from local competent authorities.
- 3. The source cable shall be leak tested at six (6) month intervals using techniques capable of detecting 0.005 microcuries of removable contamination.
- Handling, storage, use, transfer and disposal will be determined by the licensing 4. authority. The sealed source must be installed and transferred by experienced, trained and licensed personnel only using adequate handling procedures.
- ક. The licensee shall not cut, splice or altar the source cable in any way.
- To be used as an active source cable for the Nucletron Oldelft Microselectron HDR 6.
{2}------------------------------------------------
Afterloader at existing irradiation equipment installations. The licensee shall strictly abide by all requirements of the customer's license and applicable requirements of the U.S.N.R.C., agreement state or international competent authorities in which the authorized work is performed
## External Radiation Levels
The following are radiation dose rates for a 10 curie source
| Distance (in cm) | Dose Rate (mr/hr) |
|------------------|-------------------|
| 1 | 48,000,000 |
| 25 | 76,800 |
| 50 | 19,200 |
| 100 | 4,800 |
These values are for health physics ONLY and not for clinical use.
Based on the dimensions of the subject device being the same as the predicate device, the body tissue contacting materials being made of a known biocompatible material, stainless steel, and both devices using IR-192 as the radionuclide, the subject device, Alpha-Omega Services, Inc., 10 Curie Indium 192 Source, Model CSN0010-192 for use with Nucletron Microselection HDR device is substantially equivalent to the predicate device with respect to features and indications.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 2 2000
Alpha-Omega Services, Inc. C/O J. Harvey Knauss Delphi Consulting Group 11874 South Evelyn Circle Houston, TX 77071
Re:
K991571 10 Curie Iridium 192 Source, Model CSN0010-192 Dated: January 18, 2000 Received: January 19, 2000 Regulatory class: II 21 CFR 892.5730/Procode: 90 KXK
Dear Mr. Knauss:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive. Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number K 991571
10 Curie Iridium 192 Source, Model CSN0010-192 Device Name:
- Indications for use: A high specific activity radionuclide brachytherapy source to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy. Device is the active source for a Nucletron Microselection Device.
Prescription Device. Federal Law (US) restricts this device to dale by or on the order of a physician.
## (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K991571
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
ર્વ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.