# MR SPECTROSCOPY PACKAGE (K991568)

_Philips Medical Systems (Cleveland), Inc. · LNI · Jul 19, 1999 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K991568

## Device Facts

- **Applicant:** Philips Medical Systems (Cleveland), Inc.
- **Product Code:** [LNI](/productcode/LNI.md)
- **Decision Date:** Jul 19, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Picker MR Spectroscopy Package is intended for use as a non-invasive diagnostic device that provides information based on relative concentrations of metabolites in body tissues. This NMR data in the form of spectra or spectral images reflect the NMR properties of proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and chemical shift. When interpreted by a trained medical practitioner, these spectral data provide information that can be useful in making a diagnosis.

## Device Story

Proton (1H) MR spectroscopy software option for existing MRI systems; performs single voxel proton (SVP) spectroscopy and chemical shift imaging (CSI). Inputs: NMR signals from body tissues; processes data to calculate relative metabolite concentrations (NAA, Cr, Cho, mI, Glx, Lac). Features automated solvent suppression (MOIST, MOIST2, FATFREE, DUALSAT) and localization (STEAM, PRESS). Output: spectra or spectral images displayed to clinicians for diagnostic interpretation. Used in clinical settings by trained medical practitioners; aids in non-invasive biochemical characterization of tissues. Benefits patients by providing diagnostic information without invasive procedures.

## Clinical Evidence

No clinical data; bench testing only. Substantial equivalence established through comparison of technological characteristics, hardware, and software processing capabilities against predicate devices.

## Technological Characteristics

Proton (1H) spectroscopy software package. Localization: STEAM and PRESS. Solvent suppression: Automated (MOIST, MOIST2, FATFREE, DUALSAT). Minimum voxel size: 5x5x5 mm3. Minimum TE: 20ms (STEAM), 35ms (PRESS). Connectivity: Integrates with MRI system database, film, and post-script printers. Software performs FFT, phase correction, peak fitting, and baseline correction.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- GE Probe ([K930265](/device/K930265.md))
- Siemens Clinical Proton Spectroscopy Package ([K951650](/device/K951650.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a handwritten string of characters. The string starts with the letters 'K99' followed by the numbers '1568'. The characters are written in a casual, slightly slanted style, giving it a personal touch. The image is simple, focusing solely on the handwritten text.

# SUMMARY OF SAFETY AND EFFECTIVENESS

(As required by 21 CFR 807.92)

#### General Information 1.

| Classification:             | Class II<br>Magnetic Resonance Imaging (MRI) Option                                                                                                                                                                                    |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name:          | Magnetic Resonance Spectroscopy Option                                                                                                                                                                                                 |
| Proprietary Name:           | MR Spectroscopy Package                                                                                                                                                                                                                |
| Establishment Registration: | Picker International, Inc.<br>World Headquarters<br>595 Miner Road<br>Highland Heights, Ohio 44143<br>Contact: Elaine K. Keeler, Ph.D.<br>Phone: (440) 473-3000<br><br>FDA Owner Number: #1580240<br>FDA Registration Number: #1525965 |
| Performance Standards:      | No applicable performance standards have been<br>issued under section 514 of the Food, Drug and<br>Cosmetic Act                                                                                                                        |

#### 2. Intended Uses and Indications for Use

The Picker MR Spectroscopy Package is intended for use as a non-invasive diagnostic device that provides information based on relative concentrations of metabolites in body tissues. This NMR data in the form of spectra or spectral images reflect the NMR properties of proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and chemical shift. When interpreted by a trained medical practitioner, these spectral data provide information that can be useful in making a diagnosis.

| This package is indicated for use as follows: |                  |
|-----------------------------------------------|------------------|
| Anatomical region:                            | Head, whole body |
| Nuclei Excited:                               | $^{1}H$          |

#### 3. Device Description

Picker's MR Spectroscopy Package described in this submission is a proton ('H) spectroscopy software option. The package includes both capabilities for single voxel proton (SVP) spectroscopy and chemical shift imaging (CSI). These techniques provide a non-invasive method for analyzing metabolite concentrations in the brain and throughout

PICKER INTERNATIONAL, INC. 4/27/99 I - I (SPEC)

{1}------------------------------------------------

the body. The MR Spectroscopy Package does not include any additional risks to the patient other than those for standard MR imaging.

### 4. Safety and Effectiveness

The Picker MR Spectroscopy Package is similar in technological characteristics and intended use to the Siemens Clinical Proton Spectroscopy Package and the GE Probe/SV Option. The following table has been created to demonstrate their substantial equivalence.

| Parameter                   | Picker MR<br>Spectroscopy Option                                                                                                                                                                                            | Predicate Devices: GE Probe<br>(K930265), Siemens (K951650)                                                                                                                                               |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nucleus                     | Same.                                                                                                                                                                                                                       | Proton. (Both predicates)                                                                                                                                                                                 |
| Type of Spectroscopy        | Same.                                                                                                                                                                                                                       | Single Voxel and Chemical Shift<br>Imaging. (See K951650)                                                                                                                                                 |
| Spectroscopic<br>Hardware   | Same.                                                                                                                                                                                                                       | Phantom for QA. (See K951650)                                                                                                                                                                             |
| Shimming Technique          | Same.                                                                                                                                                                                                                       | Automated. (See K930265)                                                                                                                                                                                  |
| Solvent Suppression         | Automated with the choice of<br>MOIST, MOIST2, FATFREE and<br>DUALSAT.                                                                                                                                                      | Automated CHESS. (See K930265)                                                                                                                                                                            |
| Localization<br>Techniques  | Same.                                                                                                                                                                                                                       | STEAM and PRESS. (See K930265)                                                                                                                                                                            |
| Voxel positioning           | Graphical positioning.                                                                                                                                                                                                      | Alphanumerical or graphical<br>positioning. (Both predicates)                                                                                                                                             |
| Minimum voxel size          | 5 x 5 x 5 mm3.                                                                                                                                                                                                              | 10 x 10 x 10 mm3. (See K951650)                                                                                                                                                                           |
| Minimum TE                  | STEAM – 20ms, PRESS – 35 ms.                                                                                                                                                                                                | STEAM – 10ms, PRESS – 35ms.<br>(See K930265)                                                                                                                                                              |
| Typical TR Range            | 1.5 – 3 sec.                                                                                                                                                                                                                | 1.5 – 6 sec. (See K930265)                                                                                                                                                                                |
| Acquisition Time            | Approximately 3 – 7 minutes.                                                                                                                                                                                                | Approximately 10 minutes. (See<br>K930265)                                                                                                                                                                |
| Post-processing<br>features | Phase correction, filtering, data<br>zeroing, retrospective voxel<br>boundary shifting, FFT, baseline<br>correction, peak fitting and<br>analysis, chemical shift assignment,<br>variable smoothing and<br>differentiation. | Zero filling, water reference<br>processing, FID shifting, baseline<br>correction, phase correction, FFT,<br>apodization, Curve fitting, peak<br>information display, spectrum labeling.<br>(See K951650) |
| Key Metabolites             | NAA, Cr, Cho, mI, Glx and Lac.                                                                                                                                                                                              | NAA, Cr/PCr, Cho, lipids and some<br>other metabolites only at short TEs.<br>(See K951650)                                                                                                                |

## SUBSTANTIAL EQUIVALENCE TABLE

PICKER INTERNATIONAL, INC. (SPEC) I - 2 4/27/99

{2}------------------------------------------------

|                      | Picker MR<br>Spectroscopy Option                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Devices: GE Probe (K930265), Siemens (K951650)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Parameter            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Data display options | Same.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Spectra, results of fit, images showing<br>metabolite distributions or ratios. (See<br>K951650)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Output type          | Archived to database, film, post-<br>script printer.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Archived in database, Windows bitmap<br>format, laser camera, standard paper<br>laser printer, digital printer. (See<br>K951650)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Indications for use  | Head, whole body.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Brain and any anatomy containing little<br>or no fat. (See K930265)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Intended Use         | The Picker MR Spectroscopy<br>Package is intended for use as a<br>non-invasive diagnostic device that<br>provides information based on<br>relative concentrations of<br>metabolites in body tissues. This<br>NMR data in the form of spectra<br>or spectral images reflect the NMR<br>properties of proton density, spin-<br>lattice relaxation time (T1), spin-<br>spin relaxation time (T2), and<br>chemical shift. When interpreted<br>by a trained medical practitioner,<br>these spectral data provide<br>information that can be useful in<br>making a diagnosis. | The GE Signa Advantage MR<br>diagnostic system is indicated for use<br>as a diagnostic medical device that<br>produces computer processed,<br>localized spectra that display the<br>internal biochemical characteristics of<br>the human body. Other parameters<br>derived from the spectra may also be<br>displayed. The spectra reflect the<br>frequency distribution of several<br>biologically relevant nuclei exhibiting<br>nuclear magnetic resonance (NMR).<br>The NMR properties that determine<br>the spectral appearance are nuclear<br>density, spin-lattice relaxation time<br>(T1), spin-spin relaxation time (T2),<br>chemical shift in resonant frequency,<br>and fluid flow. When interpreted by a<br>trained specialist, these spectra provide<br>information that may be useful in the<br>determination of a diagnosis. (See<br>K930265) |

PICKER INTERNATIONAL, INC.

(SPEC)

4/27/99

.

I - 3

. : 上一

:

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/3/Picture/16 description: The image shows a partial view of a logo or emblem. The visible portion includes a stylized graphic of three curved lines stacked on top of each other, resembling a simplified representation of waves or motion. To the left of the graphic, there is a partial arc of text, suggesting the presence of a circular border or frame around the logo. The text is oriented along the curve, indicating that it is part of the emblem's design.

JUL 1 9 1999

Elaine Keeler, Ph.D. Picker International, Inc. 595 Miner Road Highland Heights, Ohio 44143 RE:

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

K991568 Magnetic Resonance Spectroscopy Package Dated: April 30, 1999 Received: May 5, 1999 Regulatory Class: II 21 CFR 892.1000/Procode: 90 LNI

Dear Dr. Keeler:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: MR Spectroscopy Package

Indications for Use:

The Picker MR Spectroscopy Package is intended for use as a non-invasive diagnostic device that provides information based on relative concentrations of metabolites in body tissues. This NMR data in the form of spectra or spectral images reflect the NMR properties of proton density, spinlattice relaxation time (T1), spin-spin relaxation time (T2), and chemical shift. When interpreted by a trained medical practitioner, these spectral data provide information that can be useful in making a diagnosis.

This package is indicated for use as follows:

Head, whole body Anatomical region: Nuclei Excited: ,H

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
  

  

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices

| 510(k) Number | K991568 |
|---------------|---------|
|---------------|---------|

| Prescription Use<br>(Per 21 CFR 801.109) | <div style="display:inline-block;width:30px;"></div> OR <div style="display:inline-block;width:30px;"></div> Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|

---

**Source:** [https://fda-staging.innolitics.com/device/K991568](https://fda-staging.innolitics.com/device/K991568)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).