← Product Code [IYE](/productcode/IYE) · K991523

# TLC PEDIATRIC FRAME, MODEL TLC-SYS (K991523)

_Radionics, Inc. · IYE · Jul 30, 1999 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K991523

## Device Facts

- **Applicant:** Radionics, Inc.
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** Jul 30, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The TLC Pediatric Frame is intended for use as a stereotactic platform for fractionated (multiple, lower dose) LINAC-based radiation treatments and related CT and angiographic imaging in the cranium for the treatment of pediatric patients.

## Device Story

The TLC Pediatric Frame is a non-invasive, re-locatable head holder system used in clinical radiation oncology settings. It functions as a stereotactic platform to ensure precise, repeatable patient positioning for fractionated LINAC-based radiation treatments and diagnostic imaging (CT/angiography). The device consists of a base frame, plastic occipital head cups, earplug tips, and a custom-molded interface for the patient's frontal, nasion, and lateral eye regions. A 'Depth Helmet' attaches to the frame to maintain specific depth measurements, allowing clinicians to accurately re-locate the patient for subsequent treatment fractions. By ensuring consistent cranial positioning, the device facilitates accurate radiation delivery, potentially improving treatment outcomes for pediatric patients.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Non-invasive re-locatable head holder system. Components include base frame, plastic occipital head cups, earplug tips, and custom-molded patient interface. Includes Depth Helmet for measurement maintenance. Mechanical positioning device; no energy source or software.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Radionics Gill-Thomas-Cosman (GTC) Re-locatable Head Holder System ([K934523](/device/K934523.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary of Safety & Effectiveness for the Radionics TLC Pediatric Frame

K991523

### 1. Sponsor

Radionics Software Applications, Inc. 22 Terry Avenue Burlington, MA 01803

| Contact Person: | Nancy C. MacDonald      |
|-----------------|-------------------------|
|                 | Sr. Regulatory Engineer |
|                 | (781) 272-1233          |

#### 2. Device Name

| Proprietary Name:    | TLC Pediatric Frame                                                 |
|----------------------|---------------------------------------------------------------------|
| Common/Usual Name:   | Stereotactic Radiation Treatment Planning<br>System and Accessories |
| Classification Name: | X-ray Radiation Therapy System                                      |

## 3. Predicate Device

Radionics Software Applications, Inc. believes that within the meaning of the Medical Devices Amendments of 1976, the TLC Pediatric Frame addressed in this premarket notification is substantially equivalent to the following medical device:

> The Radionics Gill-Thomas-Cosman (GTC) Re-locatable Head Holder System, cleared under 510(k) #K934523.

### 4 . Device Description

The TLC Pediatric Frame is a Re-locatable Head Holder System. The device consists of a base frame that combines the attachment of various sized plastic occipital head cups, left and right earplug tips, and a mold of the patient's frontal, nasion and lateral of eyes. The Radionics' Depth Helmet attaches to the frame and depth measurements are maintained to allow the operator to re-locate the patient for each radiation therapy treatment.

{1}------------------------------------------------

# 5. Intended Use

The TLC Pediatric Frame is intended for use as a stereotactic platform for fractionated (multiple, lower dose) LINAC-based radiation treatments and related CT and angiographic imaging in the cranium for the treatment of pediatric patients.

# General Safety and Effectiveness 6 .

The TLC Pediatric Frame labeling contains Instructions for Use. The Operator's Manual includes: indications for use, cautions, warnings and user quality assurance procedures. This information promotes safe and effective use of this device.

# Comparison of Technological Characteristics 7 .

Radionics Software Applications, Inc. believes that the information and testing provided in this submission clearly describes the technological characteristics of the TLC Pediatric Frame and demonstrates that the TLC Frame is substantially equivalent to the commercially available GTC Re-locatable Head Holder System.

# 8. Performance Testing

Non-clinical tests were conducted to demonstrate that the TLC Pediatric Frame meets all product requirements. This testing also demonstrates that the performance is substantially equivalent to the predicate device cited above.

# 9. Safety Summary

The TLC Pediatric Frame has quality assurance equipment and procedures to verify, prior to patient treatment, that the TLC Frame is accurately positioned with respect to the patient's cranium.

# 10. Basis for Substantial Equivalence

The TLC Frame, addressed in this premarket notification, has the same technological characteristics as the commercially available Radionics GTC Relocatable Head Holder System. Like the GTC, the TLC Frame ins non-invasive, and allows repeat positioning for lower dose multiple fractionated stereotactic LINAC-based radiation treatments and related CT and angiographic imaging,

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of an eagle.

JUL 30 1999

Nancy macDonald Senior Regulatory Engineer Radionics Software Applications, Inc. 22 Terry Avenue Burlington, Massachusetts 01803-2516 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

K991523 TLC Pediatric Frame Dated: April 30, 1999 Received: May 3, 1999 Regulatory Class: II 21 CFR 892.5050/Procode: 90 IYE

Dear Ms. MacDonald:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Re:

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page of

510(k) Number (if known):

TLC Pediatric Frame Device Name:

# Indications for Use:

The TLC Frame is indicated for use as a stereotactic platform for fractionated (multiple, lower dose) LINAC-based radiation treatments and related CT and angiographic imaging in the cranium for the treatment of pediatric patients.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Thind A. beymm
(Division Sign-Off)

( Division of Reproductive, Abdominal, ENT,
and Radiological Devices and Radiological Devi 510(k) Number

Prescription Use: 
(Per 21 CFR 801.109)

OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda-staging.innolitics.com/device/K991523](https://fda-staging.innolitics.com/device/K991523)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).