K991509 · Sultan Chemists, Inc. · LRJ · Jul 14, 1999 · General Hospital
Device Facts
Record ID
K991509
Device Name
PRO-PORTION CLEANER/DISINFECTANT
Applicant
Sultan Chemists, Inc.
Product Code
LRJ · General Hospital
Decision Date
Jul 14, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6890
Device Class
Class 1
Indications for Use
Pro-Portion is a general-purpose cleaner/disinfectant. Its use is restricted to a health care setting where low and intermediate level disinfection of hard, non-porous surfaces is required. It is intended to be used in a dental office as an instrument ultrasonic cleaning solution. In order to disinfect, a contact time of 10 minutes at 69°F / 20°C is required. This product may be used to pre-clean or decontaminate of critical or noncritical medical devices prior to sterilization. Must be followed by appropriate terminal sterilization or high-level disinfection process.
Device Story
Pro-Portion is a phenolic-based, ready-to-use liquid cleaner/disinfectant; intended for hard, non-porous surfaces in dental offices and healthcare settings. Functions as an ultrasonic cleaning solution for instruments; requires 10-minute contact time at 20°C for disinfection. Used by dental staff to pre-clean or decontaminate critical/semi-critical devices prior to terminal sterilization or high-level disinfection. Not for use on devices entering sterile body areas or contacting mucous membranes. Benefits include efficient cleaning and intermediate-level disinfection of dental units, lights, and various materials including plastics, stainless steel, and glass.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Phenolic-based liquid cleaner/disinfectant; 10.5% phenolics in concentrate; pH 9.1-11.1 (concentrate), 6.8-8.2 (use-dilution); specific gravity 0.9-1.1; flash point: none; phosphate-free. Compatible with polycarbonate, PVC, polypropylene, polystyrene, vinyl, stainless steel, Plexiglas, and glass.
Indications for Use
Indicated for use in healthcare settings, specifically dental offices, for low and intermediate level disinfection of hard, non-porous surfaces and as an ultrasonic cleaning solution for instruments. Contraindicated for use as a terminal sterilant or high-level disinfectant on devices introduced into the bloodstream, normally sterile areas, or contacting intact mucous membranes.
Regulatory Classification
Identification
A general purpose disinfectant is a germicide intended to process noncritical medical devices and equipment surfaces. A general purpose disinfectant can be used to preclean or decontaminate critical or semicritical medical devices prior to terminal sterilization or high level disinfection. Noncritical medical devices make only topical contact with intact skin.
Submission Summary (Full Text)
{0}------------------------------------------------
:
# 510 K Summary K 991509
### Physical Characteristics
| Product designation: | A phenolic cleaner/disinfectant |
|---------------------------|---------------------------------|
| Appearance (concentrate): | Clear light amber liquid |
| Odor: | Slight alcohol |
| Percent of Phenolics: | Concentrate 10.5% |
| pH (concentrate): | 9.1 - 11.1 |
| pH (use-dilution): | 6.8 - 8.2 |
| Specific Gravity (25°C): | .9-1.1 |
| Flash Point: | None |
| Detergency: | Excellent |
| Phosphate Content: | None |
Image /page/0/Picture/4 description: The image shows a logo for "TUV CERT ISO 9001 REGISTERED". The logo features the "TUV" acronym in a stylized font above the word "CERT", which is enclosed in a rounded rectangle. Below "CERT" is the text "ISO 9001 REGISTERED". The entire logo is surrounded by two concentric oval shapes.
Image /page/0/Picture/5 description: The image shows a logo with the letters 'OSAP' stacked on top of each other. The letters are white and are set against a black square background. The logo is simple and modern in design.
Image /page/0/Picture/6 description: This image shows the logo and contact information for Sultan Chemists, Inc. The logo includes the company name in bold, stylized font, with a black triangle behind it. The address is listed as 85 West Forest Ave., Englewood, NJ 07631. The phone number is (201) 871-1232, the toll-free number is (800) 637-8582, and the fax number is (201) 871-0321.
Image /page/0/Picture/7 description: The image shows the letters DMA in a bold, sans-serif font. The letters are large and take up most of the frame. The letters are black and the background is white.
: 上一篇:
Image /page/0/Picture/8 description: The image shows a logo with the letters "ADEA" in a stylized, bold font. The letters are interconnected, with the "A" and "D" sharing a vertical line, and the "E" and "A" also sharing a line. The overall design is modern and geometric, with a sense of unity and connection between the letters.
{1}------------------------------------------------
## 510 K Summary K 991509
Sultan Chemists Submitter: 85 W. Forest Avenue Englewood, NJ 07631
| Contact Person: | Les Heimann |
|------------------------|---------------------------------------|
| Date Prepared: | June 16, 1999 |
| Propriety Name: | Pro-Portion ™ cleaner/ disinfectant |
| Common Usual Name: | General Purpose Cleaner/ Disinfectant |
| Disinfectant Category: | Intermediate level |
This product is not to be used as a terminal sterilant/high-level disinfectant on any surface or instrument that (1) is introduced directly into the human body, either into or in contact with the bloodstream or normally sterile areas of the body, or (2) contacts intact mucous membranes but which does not ordinarily penetrate the blood barrier or otherwise enter normally sterile areas of the body. This product may be used to preclean or decontaminate critical or semi-critical medical devices prior to sterilization/high-level disinfection. To be effective a ten-minute contact time at room temperature (69°F / 20°C) is required.
#### Use, Functions, Concepts, and Performance Characteristics
Pro-Portion is a broad spectrum, multi-purpose ready to use, effective cleaner and disinfectant for use on the surfaces of inanimate objects. Pro-portion will efficiently clean and disinfect, when used as directed, such items as laminated counter tops, dental units, dental instruments, lights, and other inanimate surfaces, including those made of plastics (such as: polycarbonate, polyvinylchloride, polypropylene and polystyrene), nonporous vinyl and upholstery, stainless steel, painted surfaces, Plexiglas, glass, and other hard non-porous surfaces.
Image /page/1/Picture/6 description: The image shows a logo for "TUV CERT ISO 9001 REGISTERED". The logo consists of the "TUV" acronym stacked on top of the word "CERT", all enclosed within an oval shape. Below the oval, the text "ISO 9001 REGISTERED" is printed in a smaller font size. The logo appears to be a certification mark, indicating that the organization has been registered under the ISO 9001 standard.
Image /page/1/Picture/7 description: The image shows a logo with the letters "OSAP" stacked on top of each other. The letters are white and are set against a black square background. The logo appears to be a simple, graphic design.
#### 510 K Summary
ultan
Sultan Chemists, Inc.
85 West Forest Ave., Englewood, NJ 07631
Phone: (201) 871-1232 • Toll Free: (800) 637-8582
Fax: (201) 871-0321 • http://www.sultanintl.com
Image /page/1/Picture/10 description: The image shows two sets of bold, sans-serif letters. On the left, the letters "DMA" are displayed, with the "A" appearing slightly smaller and nestled between the "D" and "M". To the right, the letters "ADCA" are shown, with a similar style and thickness as the first set of letters. The letters are all in black, contrasting with the white background.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three curved lines representing its wings. The eagle is positioned above the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which are arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 14 1999
Mr. Les Heimann Vice President of Quality Sultan Chemists, Incorporated 85 West Forest Avenue 07631 Englewood, NJ
Re : K991509 Pro-Portion™ Cleaner/Disinfectant Trade Name: Requlatory Class: Unclassified Product Code: LRJ Dated: April 24, 1999 Received: April 30,1999
Dear Mr. Heimann:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Mr. Heimann Page 2 -
this response to your premarket notification Please note: submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director
Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page of
| 510(k) Number (if known): | K991509 |
|---------------------------|---------|
|---------------------------|---------|
Pro-Portion™ Device Name:
Indications For Use:
Pro-Portion is a general-purpose cleaner/disinfectant. Its use is restricted to a health care setting where low and intermediate level disinfection of hard, non-porous surfaces is required. It is intended to be used in a dental office as an instrument ultrasonic cleaning solution. In order to disinfect, a contact time of 10 minutes at 69°F / 20°C is required.
This product may be used to pre-clean or decontaminate of critical or noncritical medical devices prior to sterilization. Must be followed by appropriate terminal sterilization or high-level disinfection process.
(PLEASE DO NOT VRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Qlin S. Lin
(Division Sign-Off)
Division of Dental Infection Control,
and General Hospital Devices
510(k) Number
K941509
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.