← Product Code [DCK](/productcode/DCK) · K991385

# DADE BEHRING N HIGH SENSITIVITY CRP ASSAY (K991385)

_Dade Behring, Inc. · DCK · Oct 25, 1999 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K991385

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [DCK](/productcode/DCK.md)
- **Decision Date:** Oct 25, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5270
- **Device Class:** Class 2
- **Review Panel:** Immunology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K991385 · Oct 25, 1999 | DADE BEHRING N HIGH SENSITIVITY CRP ASSAY | Dade Behring, Inc. | Published clinical literature | The sponsor utilized a large, published literature base to support the expansion of the device's intended use to include the evaluation of 'associated diseases' in addition to inflammatory disorders, tissue injury, and infection. | Literature review; Intended use expansion; Clinical literature |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Published literature base; Literature review of clinical studies | Patients undergoing CRP testing in clinical practice | Not applicable for this study | Clinical utility of high sensitivity CRP measurement |

## Indications for Use

The Dade Behring N High Sensitivity CRP assay is an in vitro diagnostic test for the quantitative determination of C-reactive protein in human serum as well as heparin- and EDTA-plasma by means of particle-enhanced immunonephelometry using Dade Behring Nephelometer Systems. In acute phase response, increased levels of a number of plasma proteins, including Creactive protein, are observed. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders, and associated diseases.

## Device Story

The Dade Behring N High Sensitivity CRP assay is an in vitro diagnostic test used for the quantitative measurement of C-reactive protein (CRP) in human serum or plasma samples. The device utilizes particle-enhanced immunonephelometry on Dade Behring Nephelometer Systems. The assay measures CRP levels to assist clinicians in the detection and evaluation of inflammatory disorders, tissue injury, infection, and associated diseases. The device is intended for professional use in clinical laboratory settings. Results are provided to healthcare providers to support clinical decision-making regarding patient inflammatory status.

## Clinical Evidence

No clinical data provided; the labeling change is supported by a large, published literature base using the Dade Behring methodology and other high sensitivity CRP methods.

## Technological Characteristics

Particle-enhanced immunonephelometric assay for quantitative CRP measurement. Designed for use with Dade Behring Nephelometer Systems. Analyte: C-reactive protein in human serum, heparinized plasma, or EDTA plasma.

## Regulatory Identification

A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 2 5 1999

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

## Dade Behring N High Sensitivity CRP

## Summary of Safety and Effectiveness

N High Sensitivity CRP is a nephelometric assay for the quantitative measurement of C-reactive protein in human serum, or heparinized or EDTA plasma. FDA has previously cleared this assay via the 510(k) process for the intended use for "the detection and evaluation of inflammatory disorders, tissue injury and infection".

The proposed revision is limited to modifications in the intended use. The intended use will be "for the detection and evaluation of tissue injury, inflammatory disorders, and associated diseases."

There are no other changes to the assay reagents or test system.

This labeling change is supported by a large, published literature base that uses both the Dade Behring methodology as well as other high sensitivity CRP methods.

Rebecca S. Aycock

Reberra S Manager, Regulatory Affairs

9/28/99

Date

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a triple-stranded helix intertwined with a staff. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA". The text is arranged in a circular fashion around the caduceus.

OCT 2 5 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Rebecca S. Ayash Manager, Regulatory Affairs Dade Behring Inc. Glasgow Building 500, Mailbox 514 P.O. Box 6101 Newark, Delaware 19714

Re: K991385

> Trade Name: Dade Behring N High Sensitivity CRP Assay Regulatory Class: II Product Code: DCK Dated: September 28, 1999 Received: September 29, 1999

Dear Ms. Ayash:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895, A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

## Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications Statement

Device Name: Dade Behring N High Sensitivity CRP assay

- Indications for Use: The Dade Behring N High Sensitivity CRP assay is an in vitro diagnostic test for the quantitative determination of C-reactive protein in human serum as well as heparin- and EDTA-plasma by means of particle-enhanced immunonephelometry using Dade Behring Nephelometer Systems. In acute phase response, increased levels of a number of plasma proteins, including Creactive protein, are observed. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders, and associated diseases.
ban looph

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K 991385

Rebecca S. Ayers

becca S. Avash Manager, Regulatory Affairs

9/20/99

---

Date

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

K991385
510(k) Number

Division Sign-Off Office of Device Evaluation

prescription use

---

**Source:** [https://fda-staging.innolitics.com/device/K991385](https://fda-staging.innolitics.com/device/K991385)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
