The intended use of Medi-Cult Sperm Preparation Medium is for sperm washing and for sperm separation by swim-up method, since seminal plasma must be separated from the spermatozoa in order to facilitate capacitation. The selection of motil spermatozoa is done in order to increase the likelihood of fertilization. The Sperm Preparation Medium is also used as holding medium for the oocytes before ICSI.
Device Story
Medi-Cult Sperm Preparation Medium is a liquid culture medium used in clinical IVF laboratories. Formulation includes Earl's Balanced Salt Solution (EBSS), sodium pyruvate, synthetic serum replacement (SSR), HEPES, sodium bicarbonate, human serum albumin (HSA), penicillin, streptomycin, and phenol red. Device facilitates sperm washing and separation of motile spermatozoa via swim-up method; also serves as holding medium for oocytes prior to ICSI. Used by embryologists and clinicians in fertility clinics. Output is prepared sperm or held oocytes, intended to improve fertilization outcomes. Safety verified through bioburden, sterility, pH, osmolarity, endotoxin, and sperm survival testing.
Clinical Evidence
Clinical data provided from multiple IVF clinics in Norway (1992-1996) and the UK (1996-1997). Data includes live birth rates per started cycle and per embryo replacement. For UK clinics using the medium, live birth rates per embryo replacement ranged from 22.3% to 34.5%, compared to a national average of 21.8%. Insemination data from the University of Trondheim (1988-1998) showed a 13.4% pregnancy rate per insemination (117 pregnancies from 877 inseminations).
Technological Characteristics
Liquid culture medium containing Earl's Balanced Salt Solution (EBSS), sodium pyruvate, synthetic serum replacement (SSR), HEPES, sodium bicarbonate, human serum albumin (HSA), penicillin, streptomycin, and phenol red. Quality control parameters include bioburden, filter integrity, sterility, pH, osmolarity, endotoxin levels, and sperm survival.
Indications for Use
Indicated for use in assisted reproductive technology (ART) procedures, specifically for sperm washing, separation of motile spermatozoa via swim-up method to facilitate capacitation and increase fertilization likelihood, and as a holding medium for oocytes prior to intracytoplasmic sperm injection (ICSI).
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Submission Summary (Full Text)
{0}------------------------------------------------
# MAY = 4 2000
## 510(k) Summary Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination
## (K991332)
#### The products:
- -"Medi-Cult Sperm Preparation Medium" Cat.No. 1070
- -"Medi-Cult Sperm Preparation Medium w/o Pen/Strep." Cat.No 1068
- . "Medi-Cult Sperm Preparation Medium w/o Phenol Red" Cat.No 1069
Indication for use:
"Medi-Cult Sperm Preparation Medium" is for sperm washing and for sperm separation by swim-up method. The "Medi-Cult Sperm Preparation Medium" is also used as holding medium for the occytes before ICSL
Product formulation: Earl's Balanced Salt Solution (EBSS) Sodium Pyruvate Synthetic Serum Replacement (SSR) HEPES Sodium Bicarbonate Human Serum Albumin (HSA) Penicillin Streptomycin Phenol Red
Product testing control contents:
- Bioburden, production-test .
- Integrity filter testing, production-test -
- -Sterility, QC-test
- pH, QC-test
- -Osmolarity, QC-test
- -Endotoxin, QC-test
- -Sperm Survival, QC-test
The culture media from Medi-Cult have been used by many European IVF-units since the end of 1980. (Produced and distributed by GEA-Biotech 1987 to 1989 and by Medi-Cult a/s from 1989). The Medi-Cult media were at that time introduced as a replacement for in-house prepared culture media.
A number of publications in peer- reviewed books or journals have presented data using Medi-Cult media. The clinical result of a given IVF-clinic will depend on the patient waters of the clinical procedures, the laboratory routines and on the various culture media used. A comparison of the clinical performance of clinics using only Medi-Cult media with other IVF-clinics using similar clinical performance of give an indicate on the performance of the Medi-Culture media. In countries were we have access to both national data and data from individual clinics, clinics using Medi-Cult media generally performs well.
The IVF-units in the Nordic countries use similar clinical and laboratory protocols and their patient populations are comparable. The clinical data from most Nordic and their patient of the performance of individual IVF-clinics are available.
The largest IVF clinics in Norway have been using only Medi-Cult media since the end of the 1980's. The data collected in the years from 1992 to 1996 are shown below as live birth rates per started cycle.
| Clinic | 1992 | 1993 | 1994 | 1995 | 1996 |
|-------------------------------------|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|-------|------|
| Dept of Ob & Gyn, Univ of Trondheim | 99.9<br>And for a loss | c Carola Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara<br>And lest in days | 19.8 | 25.35 | 18.4 |
| The National Hospital, Oslo | 13.3 | 19.0 | 15.8 | 11.42 | 17:4 |
| National average | 14.6 | 17.5 | 17.9 | 16.84 | 16.3 |
Table. Live birth rates per started cycle obtained in IVF units in Norway
{1}------------------------------------------------
In Sweden, the Carl von Linne Clinic uses only products from Medi-Cult for culture of gametes and embryos. For the year 1997 their birth rate per embryo replacement was 34% their clinical results were the best in Sweden and well above the national average of 25.8%.
The Human Fertilisation and Embryology Authority (HFEA, in the UK) collects clinical data from all centres licensed to offer treatment for infertility by assisted reproduction. HFEA publishes these data in a booklet and on the web. The clinical data are presented both unadjusted for the demographics of the clinics patient population.
In the Table below the clinical data from some UK-clinics that use only Medi-Cult products is shown. These clinics have all given their consent to be presented as clinics using only Medi-Cult products.
Table 2. Live birth rates per embryo replacement obtained in the UK in the period April 1* 1996 to March 31". 1997.
| Name of clinic | Number of<br>IVF- cycles | Live birth rate per embryo<br>replacement % |
|----------------------------------------------|--------------------------|---------------------------------------------|
| Northamptonshire Fertility service | 503 | 26.0 |
| The Bridge Centre | 852 | 27.2 |
| Chelsea and Westminster Hospital | 388 | 22.3 |
| CARE at Park Hospital | 650 | 34.5 |
| BMI Ross Hall Hospital | 68 | 27.5 |
| The Woking Nuffield Hospital | 163 | 32.6 |
| Leeds General Infirmary | 1434 | 24.0 |
| Nurture | 1245 | 28.8 |
| Holly House Fertility and IVF | 333 | 30.5 |
| Guy's and St Thomas Assisted Conception Unit | 651 | 27.1 |
| UK national average | 33520 | 21.8 |
Data from insemination are not reported along the same lines as that of IVF/ICSI National data is therefore difficult to obtain. Many studies using the Medi-Cult sperm preparation medium have been published.
Some clinical data are available from clinics using Medi-Cult sperm preparation medium for in conjunction with intrauterine insemination. The outcome after insemination with husbands sperm (AIH) at the IVF-Unit University of Trondheim, Norway in the period 1988-1998.
| Number of<br>inseminations | Clinical pregnancies | Pregnancy rate per<br>insemination |
|----------------------------|----------------------|------------------------------------|
| 877 | 117 | 13.4 |
There have been no registered complaints on the product and there is no evidence in the last 1.5 year that the product has been the cause of any serious adverse events in connection with its intent use.
Thus based on the Federal Register Notice (Final Rule, Vol. 63, No. 175, page 48429, September 10, 1998), "Obstetric and Gynecologic Devices; Reclassification and Classification of Medical Devices Used for Lly-Vitro Fertilization and Related Assisted Reproduction Procedures" effective on October 13, 1998 and the upportive clinical data we feel that the safety and effectiveness of the product for its intended use is shown in the present submission.
**Prepared and Submitted by:**
Ronald G. Leonardi, Ph. D. 12/14/99
Date
President, R & R Registrations P.O. Box 262069, San Diego CA 92196 1-858-586-0751
Date
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three wavy lines representing the agency's mission to promote health and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 4 2000
Medi-Cult A/S c/o Ronald G. Leonardi, Ph.D. President R & R Registrations P.O. Box 262069 San Diego, CA 92196
Re: K991332 Medi-Cult Sperm Preparation Medium Dated: April 20, 2000 Received: April 21, 2000 Requlatory Class: II 21 CFR §884.6180/Procode: 85 MQL
Dear Dr. Leonardi,
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the We have reviewed your Ookland (for the indications for use stated in the enclosure) to legally marketed predicate devices device is Substantially equivality (1976) the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, nave been reclassince in acoordance that the provisions of the Act. The general controls provisions of the Act include mante the device, subject to the general senses, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Such additional Controls. Externination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and requirements, as set form in the Quant Nogalar (FDA) will verify such assumptions. Failure to complions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish futher announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might the Podrial Regions 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vites if you done specific device for your do habbing e at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Mishranding by reference to premarket notification" (21CFR 807.97) Other general information on your responsibilities under the mission of Childred from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html"
Sincerely yours,
Daniel G. Schultz, M.D.
Captain USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiolonical Health
Enclosure(s)
{3}------------------------------------------------
### B4. Indication for Use Statement
510(k) Number (if know)\$^h\$ K991332
Device Name: Medi-Cult Sperm Preparation -- Medium
Indications for use:
The intended use of Medi-Cult Sperm Preparation Medium is for sperm washing and for sperm separation by swim-up method, since seminal plasma must be separated from the spermatozoa in order to facilitate capacitation. The selection of motil spermatozoa is done in order to increase the likelihood of fertilization.
The Sperm Preparation Medium is also used as holding medium for the oocytes before ICSI.
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|-----------------------------------------------------------------------|---------------------------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109) | <span style="text-decoration: overline;">✓</span> |
| OR | |
| Over-The-Counter Use | <span style="text-decoration: overline;"></span> |
| (Optional Format 1-2-96) | |
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices | |
| 510(k) Number | K991332 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.