K991296 · Tutogen Medical U.S., Inc. · FTM · Oct 6, 2000 · General, Plastic Surgery
Device Facts
Record ID
K991296
Device Name
TUTOPATCH
Applicant
Tutogen Medical U.S., Inc.
Product Code
FTM · General, Plastic Surgery
Decision Date
Oct 6, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Tutopatch bovine pericardium is intended for use to reinforce soft tissue where weakness exists in general and plastic surgery applications and is indicated for repair of pericardial structures and for use as a prosthesis for the surgical repair of soft tissue deficiencies which include: gastric banding, muscle flap reinforcement, repair of rectal prolapse using an abdominal approach (excluding rectocele), reconstruction of the pelvic floor using an abdominal approach (excluding transvaginal repair of pelvic organ prolapse), and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, paracolostomy, ventral, scrotal, and umbilical).
Device Story
Tutopatch is a processed bovine pericardium surgical mesh; derived from bovine pericardial tissue; processed via Tutoplast method involving cleaning, solvent dehydration, and terminal gamma-irradiation. Device consists of collagenous connective tissue with three-dimensional intertwined fibers; provides multidirectional mechanical strength and elasticity; supports replacement by endogenous tissue. Supplied in sterile peelable pouches in various sizes. Used by surgeons in general and plastic surgery procedures to reinforce soft tissue or repair defects. Implanted during surgery to provide structural support; reapproximates well around sutures; resists fluid leakage. Benefits patient by providing a biocompatible scaffold for tissue repair.
Clinical Evidence
Bench testing only. Device evaluated for safety, biocompatibility, toxicity, pyrogenicity, sterility, and mechanical strength (tensile and burst strength). All test results met expected performance criteria for a surgical mesh.
Technological Characteristics
Bovine pericardium collagenous tissue; Tutoplast processed (solvent-dehydrated, gamma-irradiated). Multidirectional fiber structure. Sterile, flexible, wall thickness similar to native pericardium. No software or electronic components.
Indications for Use
Indicated for reinforcement of soft tissue in general and plastic surgery, including repair of pericardial structures and soft tissue deficiencies such as gastric banding, muscle flap reinforcement, rectal prolapse (abdominal approach), pelvic floor reconstruction (abdominal approach), and various hernias (diaphragmatic, femoral, incisional, inguinal, lumbar, paracolostomy, ventral, scrotal, umbilical).
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Peri-Strips Staple Line Reinforcement - Sleeve Configuration (K983162)
Peri-Strips Staple Line Reinforcement - Strip Configuration (K983162)
Peri-Strips Dry Staple Line Reinforcement (K983162)
Submission Summary (Full Text)
{0}------------------------------------------------
ADVANCING SCIENCE, SAFETY & INNOVATION
RTI BIOLOGICS"
# Attachment 1:
# 510(k) PREMARKET NOTIFICATION SUMMARY OF SAFETY AND EFFECTIVENESS In Accordance with the Safe Medical Device Act of 1990
# Tutopatch™ bovine pericardium
### 1) Submitted by:
RTI Biologics, Inc. 11621 Research Circle Alachua, FL 32615 Phone: 386-418-8888 x4347 Fax: 386-418-1627
Contact person:
Travis Arola, MS, RAC, CTBS Regulatory Affairs Manager Telephone: (386) 418-8888 Facsimile: (386) 418-1627 E-mail: tarola@rtix.com
- 2) Original Date Prepared: August 25, 2000 Date Updated: October 24, 2012
- 3) Name of Medical Device: Tutopatch™ bovine pericardium Proprietary Name: Classification Name: Mesh, Surgical
### 4) Device Classification:
The Center for Devices and Radiological Health has classified bovine pericardium used for general and plastic surgery indications (Mesh, Surgical, Polymeric) as a Class II device pursuant to 21 CFR §878.3300.
| Device: | Mesh, Surgical |
|--------------------|-----------------------------|
| Medical Specialty: | General and Plastic Surgery |
| Product Code: | FTM |
| Class: | 2 |
| 510(k) Exempt? | No |
| Regulation Number: | 21 CFR §878.3300 |
{1}------------------------------------------------
**RTI|BIOLOGICS**® K991296 page 2/3
ADVANCING SCIENCE, SAFETY & INNOVATION
## 5) Device Description:
Tutopatch bovine pericardium is a Tutoplast® Processed, solvent-dehydrated gamma irradiated preserved bovine pericardium surgical mesh. Tutopatch bovine pericardium consists of collagenous connective tissue with three-dimensional intertwined fibers. Therefore, it has a multidirectional mechanical strength and can be fixed regardless of the direction of the graft. Collagenous connective tissue with multidirectional fibers retains the mechanical strength and elasticity of the native tissue, while providing the basic formulative structure to support replacement by new endogenous tissue. Tutopatch bovine pericardium is supplied in sterile double-packaged peelable pouches and comes in a variety of sizes.
## 6) Indication for Use:
Tutopatch bovine pericardium is intended for the reinforcement of tissue during general and plastic surgery repair. The labeling reveals that the applicant and all predicate devices have the same intended use as a general and plastic surgery surgical mesh.
# 7) Technological Characteristics and Substantial Equivalence
Tutopatch bovine pericardium is a Tutoplast® Processed, solvent dehydrated, gamma-irradiated preserved bovine pericardium surgical mesh. All tissues are recovered from cattle free of BSE. The processing of Tutopatch bovine pericardium consists of a strict, quality controlled procedure, which involves thorough cleaning, processing, dehydration and preservation. The process leaves no deleterious residue and minimizes antigenic potential.
The Tutoplast® Process begins with thorough screening of the raw tissue such as the bovine pericardium. Tissues meeting the screening criteria are then cleaned with saline solutions of various concentrations, resulting in the osmotic destruction of cells so that the source tissue is reduced to its fiber and mineral components.
After processing with inorganic agents such as hydrogen peroxide and 1N sodium hydroxide, the tissue is preserved by the extraction of water using organic solvents, e.g., acetone. The dense collagenous fiber structure of the tissue is retained during this method of preservation. Collagenous connective tissue with three dimensionally intertwined fibers retains the multidirectional and mechanical . strength of native tissue, while providing the basic formative structure to support replacement by new endogenous tissue.
{2}------------------------------------------------
## ADVANCING SCIENCE, SAFETY & INNOVATION
RTI BIOLOGICS" 191296
After preservation, the tissue is cut into standardized sizes and packaged in transparent, double peel pouches. The finished products are terminally sterilized by gamma-irradiation. The gamma-irradiated material exhibits excellent tensile strength, burst strength, reapproximates well around suture holes, resisting leakage of fluids. It has a wall thickness similar to that of native pericardium and is flexible, making it convenient to implant. RTI Biologics, Inc. believe that product subjected to the Tutoplast® Process posed no additional questions of safety and effectiveness.
## 8) Statement of Substantial Equivalence:
Tutopatch bovine pericardium is substantially equivalent in function and intended use to the following devices:
a) Peri-Guard® Pericardium 510(k) K983162
b) Supple Peri-Guard® Pericardium 510(k) K983162
c) Peri-Strips Staple Line Reinforcement - Sleeve Configuration 510(k) K983162
d) Peri-Strips® Staple Line Reinforcement - Strip Configuration 510(K) K98310 e) Peri-Strips DryTM Staple Line Reinforcement 510(k) K983162
### 9) Performance Data:
Tutopatch bovine pericardium has been evaluated by a number of tests for safety, biocompatibility, toxicity, pyrogenicity, sterility and mechanical strength. In all instances, Tutopatch bovine pericardium functioned as intended and all test results and characteristics observed were as expected. Summaries and reports of all data are contained in the Premarket Notification submission.
### 10) Conclusions:
The data collected demonstrate that the device has the necessary structural, biocompatibility, and biomechanical characteristics to function safely and effectively as a surgical mesh.
### THIS SPACE IS LEFT BLANK INTENTIONALLY
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird-like figure with three curved lines representing its body or wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Tutogen Medical U.S., Inc. % Mr. P.J. Pardo, MS, MT (ASCP) Director of Quality Assurance 13709 Progress Boulevard, Box 19 Alachua, Florida 32615
K991296 Re: Trade/Device Name: TUTOPATCH® Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM. OXH, PAJ, DXZ Dated: August 25, 2000 Received: September 6, 2000
Dear Mr. Pardo:
This letter corrects our substantially equivalent letter of October 6, 2000.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
July 1, 2013
{4}------------------------------------------------
Page 2 - Mr. P.J. Pardo, MS, MT (ASCP)
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, FOR Peter D. Rumm -S
Mark N. Melkerson Acting Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Attachment 2: Indications for Use
510(k) Number (if known): K991296
Tutopatch™ bovine pericardium Device Names:
#### Indications for Use:
Tutopatch bovine pericardium is intended for use to reinforce soft tissue where weakness exists in general and plastic surgery applications and is indicated for repair of pericardial structures and for use as a prosthesis for the surgical repair of soft tissue deficiencies which include: gastric banding, muscle flap reinforcement, repair of rectal prolapse using an abdominal approach (excluding rectocele), reconstruction of the pelvic floor using an abdominal approach (excluding transvaginal repair of pelvic organ prolapse), and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, paracolostomy, ventral, scrotal, and umbilical).
Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# David Krause -S
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K991296
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.