← Product Code [MQK](/productcode/MQK) · K991249

# FALCON IVF FOUR WELL PLATE (K991249)

_Becton Dickinson Labware · MQK · May 5, 1999 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K991249

## Device Facts

- **Applicant:** Becton Dickinson Labware
- **Product Code:** [MQK](/productcode/MQK.md)
- **Decision Date:** May 5, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.6160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The FALCON® IVF Four Well Plate is intended for use in preparing, storing, manipulating, or transferring human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other reproduction procedures.

## Device Story

FALCON® IVF Four Well Plate is a single-use, sterile, non-pyrogenic, non-embryotoxic plastic labware designed for assisted reproduction laboratories. Device features four wells (1.39 cm2 area, 1.8 mL volume each) with a specialized lid allowing access to two wells while keeping others covered. Surfaces are optically clear and treated to be hydrophilic for optimal observation and manipulation of ova and embryos. Used by embryologists or laboratory technicians in clinical IVF settings to hold gametes/embryos during preparation, storage, or transfer procedures. Benefits include aseptic manipulation, consistent humidification via lid venting, and clear sample identification via numbered wells and writing patch.

## Clinical Evidence

No clinical data. Bench testing only: device is sterile (SAL 10^-6), non-pyrogenic (LAL < 20 EU/device), and non-embryotoxic per mouse embryotoxicity assay (MEA) 2-cell method. Materials tested for USP Class IV, V, and VI cytotoxicity.

## Technological Characteristics

Single-use crystalline polystyrene plate; four-well configuration; hydrophilic surface treatment; sterile (SAL 10^-6); non-pyrogenic; non-embryotoxic. Dimensions: 1.39 cm2 well area, 1.8 mL well volume. Lid design provides aseptic access and consistent venting for humidification.

## Regulatory Identification

Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

5/5/99

### Attachment D

## 510(k) Summary

The assigned 510(k) number is _ K 99/249

#### Submitter Information (21 CFR 807.92(a)(1))

| Submitter:    | Becton Dickinson<br>1 Becton Drive<br>Franklin Lakes, NJ 07417-1880                                     |
|---------------|---------------------------------------------------------------------------------------------------------|
| Contact:      | Cindy Morrow<br>Sr. Regulatory Specialist<br>(408) 954-2694<br>(408) 954-2495 (FAX)<br>cmorrow@bdis.com |
| Summary date: | April 12, 1999                                                                                          |

#### Device Name/Classification (21 CFR 807.92(a)(2))

| Name:           | FALCON® IVF Four Well Plate                                      |
|-----------------|------------------------------------------------------------------|
| Classification: | Assisted reproduction labware, Class II, 884.6160, Code: 85 MQI- |

#### Substantially Equivalent/Predicate Device (21 CFR 807.92(a)(3))

This product is being submitted according to the Federal Register notice located on page 48428 of Vol.63, No. 175 on September 10, 1998 under the title Obstetric and Gynecologic Devices; Reclassification and Classification of Medical Devices Used for In Vitro Fertilization and Related Assisted Reproduction Procedures.

#### Device Description (21 CFR 807.92(a)(4))

The FALCON® IVF Four Well Plate is sterile (SAL of 10%), non-pyrogenic by Limulus Amebocyte Lysate (LAL of < 20 EU/device), and nonembryotoxic as tested by the mouse embryotoxicity assay (MEA) 2-cell method. The single-use plastic plate has four wells, each well area is 1.39 cm2 and each well volume is 1.8 mL. The plate has a unique lid which provides access to two wells at a time, while two remain covered. The wells are numbered and a large writing patch allows clear sample identification. Plates are packaged in individual peel-open trays for sterile presentation and are shipped in cases of 100 plates per case.

The plates have perfectly flat, optically clear surfaces for optimum manipulation and observation of the ova and embryos. The lids are designed for aseptic manipulation and

{1}------------------------------------------------

K 991249

consistent venting to maintain proper humidification. The plates are manufactured from virgin crystalline polystyrene tested for USP Class IV, V, and VI cytotoxicity. and the surfaces are treated to provide a more wetable or hydrophilic surface for tissue culture.

### Intended Use (21 CFR 807.92(a)(5))

The FALCON® IVF Four Well Plate is intended for use in preparing, storing, manipulating, or transferring human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other reproduction procedures.

David L. Ball

4/12/99

---

Date

David Ball Director of Quality Assurance/ Regulatory Affairs Becton Dickinson

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 5 1999

Mr. David Ball Director of Quality Assurance/ Regulatory Affairs Becton Dickinson & Company 1 Becton Drive Franklin Lakes, NJ 07417-1884

Re: K991249

FALCON® IVF Four Well Plate Dated: April 12, 1999 Received: April 13, 1999 Regulatory Class: II 21 CFR 884.6160/Procode: 85 MQK

Dear Mr. Ball:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regult in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register , Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Attachment C Indications for Use

510(k) Number: _K 991249 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Device Name: FALCON® IVF Four Well Plate

The FALCON® IVF Four Well Plate is sterile, nonpyrogenic, embryotoxicity tested, singleuse plasticware intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF) or other assisted reproduction techniques.

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use       | <div style="display:inline-block;">✓</div> |
|------------------------|--------------------------------------------|
| (Per 21 CFR § 801.109) |                                            |

Or

| Over-the-Counter Use | <div style="display:inline-block;">✓</div> |
|----------------------|--------------------------------------------|
|----------------------|--------------------------------------------|

| (Division Sign-Off)                                                |        |
|--------------------------------------------------------------------|--------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices |        |
| 510(k) Premarket Notification                                      |        |
| 510(k) Number                                                      | 899249 |

8

---

**Source:** [https://fda-staging.innolitics.com/device/K991249](https://fda-staging.innolitics.com/device/K991249)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).