SAUFLON AEROTAB DISINFECTING SYSTEM FOR THE DISINFECTION OF SOFT (HYDROPHILIC) CONTACT LENSES
Applicant
Sauflon Pharmaceuticals, Ltd.
Product Code
LPN · Ophthalmic
Decision Date
Sep 14, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
The SAUFLON AEROTAB DISINFECTING SYSTEM is indicated for chemical disinfection and storage of clear (untinted) soft (hydrophilic) contact lenses. The AEROSOLV UNPRESERVED BUFFERED SALINE SOLUTION is indicated for the rinsing of soft contact lenses after cleaning, dissolving of the Sauflon AEROTAB Disinfecting Tablet to make the disinfecting/ storage solution, and rinsing of the soft contact lenses prior to insertion onto the eye. The SAUFLON Barrel Lens Case is to be used with the AEROSOLV UNPRESERVED BUFFERED SALINE SOLUTION and Sauflon AEROTAB Disinfecting Tablets as the lens case for chemical disinfection and storage of clear (untinted) soft (hydrophilic) contact lenses. The SAUFLON barrel lens case may also be used as the lens case for storage of soft (hydrophilic) contact lenses during chemical disinfection only. DO NOT USE WITH HEAT OR HYDROGEN PEROXIDE SYSTEMS.
Device Story
System comprises Aerotab disinfecting tablet, Aerosolv unpreserved buffered saline, and barrel lens case. Tablet contains halazone (active ingredient) with effervescent base (adipic acid, anhydrous sodium carbonate). User dissolves tablet in saline within barrel case to create disinfecting/storage solution. Used by contact lens wearers for daily chemical disinfection and storage. Output is disinfected lens ready for insertion. Benefits include effective microbial control without preservatives. System is non-sterile; compatible with clear soft lenses.
Clinical Evidence
6-month clinical trial with 334 subjects (276 test, 58 control). Compared Aerotab system to commercial peroxide system. Results: No adverse reactions; 90% of eye exams showed no positive slit lamp findings (vs 67% for peroxide); 95% of test eyes achieved visual acuity within 2 Snellen lines. Lens cleanliness rated 93% clinically clean. No significant gender differences. Study confirms safety and efficacy.
Indicated for chemical disinfection and storage of clear (untinted) soft (hydrophilic) contact lenses in patients requiring lens care. Contraindicated for use with heat or hydrogen peroxide systems.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
PD 1343 Disinfecting Tablet with Unisol (PMA 8890059)
Submission Summary (Full Text)
{0}------------------------------------------------
SEP | 4 1955
K991125
### 510(K) SUMMARY FOR FREEDOM OF INFORMATION
SAUFLON AEROTAB DISINFECTING SYSTEM FOR SOFT (HYDROPHILIC) CONTACT LENSES
- Sauflon Pharmaceuticals, Ltd 1. Submitted by: 49-53 York St Twickenham, Middlesex TW1 3LP United Kingdom
- John M. Szabocsik, Ph.D. Contact Person: Szabocsik and Associates 203 N. Wabash, Ste 1200 Chicago, IL 60601 TEL: 312-553-0828
July 16, 1999 2. Date Prepared
- Sauflon AEROTAB Disinfecting System: 3. Common/Usual Sauflon AEROTAB Disinfecting Tablet with Name AEROSOLV Unpreserved Buffered Saline Solution and Sauflon Barrel Lens Case
- Sauflon AEROTAB Disinfecting System: 4. Trade/ Sauflon AEROTAB Disinfecting Tablet with Proprietary AEROSOLV Unpreserved Buffered Saline Solution and Sauflon Barrel Lens Case
- Class II (Performance Standards) 21 CFR 5. FDA 886.5928 Soft (hydrophilic) contact lens Classification disinfecting tablet with saline and lens case. Sauflon AEROTAB Disinfecting System: Product code 86LPN
- 6. Substantial This product is substantially equivalence equivalent to PD 1343 Disinfecting Tablet with Unisol®, approved PMA 8890059.
#### I. INDICATIONS FOR USE
The SAUFLON AEROTAB DISINFECTING SYSTEM is indicated for chemical disinfection and storage of clear (untinted) soft (hydrophilic) contact lenses.
The AEROSOLV UNPRESERVED BUFFERED SALINE SOLUTION is indicated for the rinsing of soft contact lenses after cleaning, dissolving of the Sauflon AEROTAB Disinfecting Tablet to make the disinfecting/ storage solution, and rinsing of the soft contact lenses prior to insertion onto the eye.
{1}------------------------------------------------
K991125, FOI pg 2
The SAUFLON Barrel Lens Case is to be used with the AEROSOLV UNPRESERVED BUFFERED SALINE SOLUTION and Sauflon AEROTAB Disinfecting Tablets as the lens case for chemical disinfection and storage of clear (untinted) soft (hydrophilic) contact lenses .
The SAUFLON barrel lens case may also be used as the lens case for storage of soft (hydrophilic) contact lenses during chemical disinfection only. DO NOT USE WITH HEAT OR HYDROGEN PEROXIDE SYSTEMS.
#### PRODUCT DESCRIPTION AND CHEMISTRY II.
Sauflon AEROTAB Disinfecting System
The Sauflon Aerotab Disinfecting Tablet contains halazone (4-carboxyphenyl N-chloro sulphonamide) as the active ingredient, with an effervescent base of adipic acid and anhydrous sodium carbonate. There is no preservative in the tablet. AEROSOLV Unpreserved Buffered Saline Solution is a sterile, aqueous, isotonic solution of sodium chloride buffered with boric acid and sodium borate.
The Sauflon AEROTAB disinfecting system was shown to be compatible with clear (untinted) soft (hydrophilic) contact lenses .
### III. TOXICOLOGY
Hydrophilic lenses representative of all four groups disinfected for 30 cycles with the Sauflon AEROTAB Disinfecting System showed no cytotoxicity and no ocular irritation.
All toxicology concerning the AEROSOLV Unpreserved Buffered Saline Solution is contained in P850074 and its supplements.
The components of the lens case have passed the requirements for Plastics for Ophthalmic Products (USP XXIII).
#### MICROBIOLOGY IV.
The Sauflon AEROTAB Disinfecting Tablet is a nonsterile tablet, and the system contains no preservative. The Sauflon AEROTAB Disinfecting System meets the requirements of the Stand-Alone Procedure for Disinfecting Products, and maintains efficacy throughout the labeled shelf-life.
{2}------------------------------------------------
## K991125, FOI pg 3
All microbiology concerning the AEROSOLV Unpreserved Buffered Saline Solution is contained in P850074 and its supplements.
#### V. CLINICAL STUDIES
A. AEROTAB Disinfecting Tablet
A 6 month clinical trial of 334 subjects (276 test, 58 control) compared the safety and efficacy of the Sauflon AEROTAB Disinfecting Tablet when used with commercially available saline, daily cleaners and wetting solutions to that of a commercial peroxide disinfecting system.
There were no adverse reactions with the Sauflon AEROTAB Disinfecting System, and there were no positive slit lamp findings at 90% of the eye examinations, compared to 67% with no findings with peroxide. No symptoms were reported at 81% of visits, compared to 85% in the peroxide controls, a difference considered not significantly different. The most common symptoms were lens awareness and discomfort in both test and control groups. Appropriate visual acuity (within 2 Snellen lines of original best corrected acuity) was obtained by 95% of the test eyes and all of the control eyes. Wear time was not affected by the care regimen. Evaluation of lens cleanliness showed 93% of test lenses and 94% of control lenses were clinically clean at scheduled visits. There were no significant gender differences in the findings.
All clinical information concerning the AEROSOLV Unpreserved Buffered Saline Solution is contained in P850074 and its supplements.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 4 1999
Sauflon Pharmaceuticals, Ltd. c/o John Szabocsik, Ph.D. Szabocsik and Associates 203 North Wabash Avenue Suite 1200 Chicago, Illinois 60601
Re: K991125
Trade Name: Sauflon Aerotab Disinfecting System (Sauflon Aerotab Disinfecting Tablet, Aerosolv Unpreserved Buffered Saline Solution, Sauflon Barrel Leng Case)
Regulatory Class: II Product Code: 86 LPN, 86 LRX Dated: July 19, 1999 Received: July 30, 1999
Dear Dr. Szabocsik:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - John Szabocsik, Ph.D.
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Afrez L. Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
់
510(k) NUMBER (IF KNOWN) K991125
SAUFLON AEROTAB DISINFECTING SYSTEM DEVICE NAME (SAUFLON AEROTAB DISINFECTING TABLET, AEROSOLV UNPRESERVED BUFFERED SALINE, SAUFLON BARREL LENS CASE)
### INDICATIONS FOR USE
The preservative-free SAUFLON AEROTAB DISINFECTING SYSTEM is indicated for chemical disinfection and storage of clear (untinted) soft (hydrophilic) contact lenses.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter-Use
(Optional Format 1-2-96)
Myra Smith
(Division Jivision of Ophthalmic Devices 510(k) Number
{6}------------------------------------------------
ﻨﻲ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟ
510(k) NUMBER (IF KNOWN) K991125
AEROSOLV UNPRESERVED BUFFERED SALINE DEVICE NAME
INDICATIONS FOR USE
The AEROSOLV UNPRESERVED BUFFERED SALINE is indicated for the rinsing of soft contact lenses after cleaning, dissolving of the AEROTAB tablet to make the disinfecting/ storage solution, and rinsing of the soft contact lenses prior to insertion onto the eye.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use (Per 21 CFR 801.109) Over-The-Counter-Use
(Optional Format 1-2-96)
Myra Smith
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K991125
してこ
{7}------------------------------------------------
- - - -
# 510(k) NUMBER (IF KNOWN) K991125
#### SAUFLON BARREL LENS CASE DEVICE NAME
### INDICATIONS FOR USE
The SAUFLON Barrel Lens Case is to be used with the AEROSOLV UNPRESERVED BUFFERED SALINE and Sauflon AEROTAB Disinfecting Tablets as the lens case for chemical disinfection and storage of clear, untinted soft (hydrophilic) contact lenses.
The SAUFLON Barrel Lens Case may also be used as the lens case for storage of soft (hydrophilic) contact lenses during chemical disinfection only. DO NOT USE WITH HEAT OR HYDROGEN PEROXIDE SYSTEMS.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use (Per 21 CFR 801.109) Over-The-Counter-Use X _
(Optional Format 1-2-96)
Myra Smith
Division of Ophthalmic Devices 510(k) Number_ K 9911 2 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.