K991122 · Influence, Inc. · GEH · Jul 1, 1999 · General, Plastic Surgery
Device Facts
Record ID
K991122
Device Name
CRYO-MONO CRYOGRNIC SYSTEM
Applicant
Influence, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Jul 1, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cryo-Mono Cryogenic system is intended for cryogenic destruction of tissues during surgical procedures. It is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurosurgery, thoracic surgery, ENT, gynecology, oncology, proctocology, and urology for the ablation of tissue, including liver metastases, skin lesions, warts, tonsils, turbinate reduction and prostate tissues. Cryo-Mono is specifically indicated for tonsil reduction in patients with blood dyscrasias or other bleeding problems and/or patients for whom general anesthesia presents an increased risk.
Device Story
Cryo-Mono Cryogenic System is a cryosurgical unit for tissue ablation. Device uses gas-based cooling delivered through cryoprobes to destroy target tissues; supports single-probe operation. Operated by surgeons in clinical settings. System includes console with 4 gas cylinders and interchangeable cryoprobes (needle, tonsillectomy, turbinectomy specific). Optional foot pedal control. Unlike predicate, lacks external thermosensors and 'Stick' mode. Healthcare providers use device to perform cryosurgery; benefits include tissue destruction in patients with bleeding risks or contraindications to general anesthesia.
Clinical Evidence
No clinical data provided. Biocompatibility based on predicate device. Device conformance certified to IEC 60601-1, IEC 60601-1-2, and ASTM F882-84. Literature regarding cryotonsillectomies and cryoturbinectomies summarized in pre-510(k) submission.
Technological Characteristics
Cryosurgical unit; gas-powered cooling. Supports single cryoprobe. Materials identical to predicate (Influence, Inc. In-Probe/In-Fast). Conforms to IEC 60601-1, IEC 60601-1-2, and ASTM F882-84. Features 4 gas cylinders, optional foot pedal. Lacks external thermosensors. Software-controlled operation.
Indications for Use
Indicated for tissue ablation in general surgery, dermatology, neurosurgery, thoracic surgery, ENT, gynecology, oncology, proctocology, and urology. Specifically indicated for tonsil reduction in patients with blood dyscrasias, bleeding disorders, or those at increased risk from general anesthesia.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
Predicate Devices
Cryo-Hit Cryogenic System (Galil Medical, Ltd.)
Submission Summary (Full Text)
{0}------------------------------------------------
JUL - 1 1999
# 510(k) Summary Influence, Inc.'s Cryo-Mono Cryogenic System 510(k) Number K991122
Submitter's Name:
Peter A. Bick. M.D., President and CEO, Influence, Inc. 71 Stevenson Street, Suite 1120 San Francisco, California 94105 Telephone: 415-546-7700 / Fax: 415-546-7744
Trade Name:
Cryo-Mono Cryogenic System
Classification Name:
Cryosurgical unit
Product Code:
GEH
Predicate Devices:
Galil Medical, Ltd.'s Cryo-Hit Cryogenic System.
Intended Use and Indication for Use:
The Cryo-Mono Cryogenic system is intended for cryogenic destruction of tissues during surgical procedures. It is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurosurgery, thoracic surgery, ENT, gynecology, oncology, proctocology, and urology for the ablation of tissue, including liver metastases, skin lesions, warts, tonsils, turbinate reduction and prostate tissues. Cryo-Mono is specifically indicated for tonsil reduction in patients with blood dyscrasias or other bleeding problems and/or patients for whom general anesthesia presents an increased risk.
## Technological Characteriatics:
The Cryo-Mono is a modified version of Cryo-Hit. They have the same technological characteristics with several exceptions: (1) Cryo-Mono can support one cryoprobe only, while its predicate, Cryo-Hit, is capable of supporting up to 8 probes; (2) the Cryo-Mono is smaller than Cryo-Hit. (3) the Cryo-Mono has 4 gas cylinders compared to the Cryo-Hit' s six; (4) Cryo-Mono software has been modified, and does not have a "Stick" option for
:
{1}------------------------------------------------
six; (4) Cryo-Mono software has been modified, and does not have a "Stick" option for holding the probe to the tissue during freezing; (5) Cryo-Mono can accommodate needle shaped, tonsillectomy and turbinectomy specific probes; (6) Cryo-Mono has an optional foot pedal; (7) The Cryo-Mono does not have external thermosensors, which the Cryo-Hit has; and (8) The Cryo-Mono's single-use turbinectomy and tonsillectomy cryoprobes do not have built-in thermosensors, unlike the Cryo-Hit and Cryo-Mono' s multi-use probes.
The Cryo-Mono's single-use probes are within the range of probes that can be used with the Cryo-Hit, i.e. one to eight probes.
### Performance Data:
The biocompatibility of the Cryo-Mono is based on FDA premarket clearance of the Cryo-Hit. The only component of the Cryo-Mono that comes into contact with the patient's body is its cryoprobe. Its components are made of the same materials, as the Cryo-Hit's probes and Influence, Inc.'s In-Probe and In-Fast. Influence has certified to the Cryo-Mono's conformance with IEC 60601-1, IEC 60601-1-2 and ASTM F882-84. Influence has summarized representative articles in the published medical literature regarding cryotonsillectomies and cryoturbinectomies in a pre-510(k) submission concerning the device.
## Substantial Equivalence:
The Cryo-Mono has the same intended use and virtually identical technological characteristics as the 510(k) cleared Cryo-Hit. The minor technological differences between these devices, namely, the Cryo-Mono smaller size, use of fewer probes, lack of a "stick" option, foot pedal, modified software, fewer gas cylinders and acceptance of tonsillectomy and turbinectomy specific probes do not raise any new questions of safety and effectiveness. Thus, the Cryo-Mono is substantially equivalent to a legally marketed cryosurgical device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like strokes above a wavy line, symbolizing care and protection. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", indicating the department's name and national affiliation. The logo is presented in black and white, giving it a clean and official appearance.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 1 1999
Influence. Inc. c/o Mr. Jonathan S. Kahan, Esq. Hogan and Hartson, L.L.P. Columbia Square 555 Thirteenth St., NW Washington, D.C. 20004
Re: K991122
> Trade Name: Cryo-Mono Cryogenic System Regulatory Class: II Product Code: GEH Dated: March 30, 1999 Received: April 2, 1999
Dear Mr. Kahan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Jonathan S. Kahan, Esq.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): 11991122
Device Name:
Cryo-Mono Cryogenic System.
Indications for Use:
The Cryo-Mono Cryogenic system is intended for cryogenic destruction of tissues during surgical procedures. It is indicated for use as a cryosurgical tool in the of general surgery, fields dermatology, neurosurgery, thoracic surgery, ENT, gynecology, oncology, proctocology, and urology for the ablation of tissue, including liver metastases, skin lesions, warts, tonsils, turbinate reduction and prostate tissues. Cryo-Mono is specifically indicated for tonsil reduction in patients with blood dyscrasias or other bleeding problems and/or patients for whom general anesthasia presents an increased risk.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
| (Division Sign-off) |
| Division of General and Restorative Devices |
| 510(k) Number | K991122 |
|---------------|---------|
|---------------|---------|
| Prescription Use (Per 21 CFR 801.109) | OR | Over-The-Counter-Use |
|---------------------------------------|----|----------------------|
|---------------------------------------|----|----------------------|
| (Division Sign-Off) |
|-----------------------------------------|
| Division of General Restorative Devices |
| 510(k) Number | K991122 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.