Bartels Legionella Urinary Antigen ELISA Test is intended as an adjunct to culture for the presumptive diagnosis of past or current Legionnaires' Disease by qualitative detection of Legionella pneumophila Serogroup 1 antigen in human urine.
Device Story
Device is an enzyme-linked immunosorbent assay (ELISA) for qualitative detection of Legionella pneumophila Serogroup 1 antigen in human urine. Input: 100 µL urine specimen added to microelisa stripwells pre-coated with rabbit anti-Legionella capture antibodies. Process: Addition of HRP-conjugated rabbit antibodies; incubation; wash cycles; addition of TMB/H2O2 substrate; addition of phosphoric acid stop solution. Output: Colorimetric signal measured at 450 nm via plate reader or compared against visual interpretation card. Used in clinical laboratories; operated by laboratory personnel. Results aid clinicians in presumptive diagnosis of Legionnaires' Disease as an adjunct to culture, facilitating timely patient management.
Clinical Evidence
Clinical study of 274 urine specimens (94 culture-positive for L. pneumophila, 150 non-Legionella, 30 healthy). Compared to culture gold standard, device showed 94.7% sensitivity, 91.1% specificity, and 92.3% accuracy. Comparison of automated plate reader vs. visual interpretation card showed 92.6% sensitivity and 93.9% specificity for visual method.
Indicated for individuals suspected of acute infection by Legionella pneumophila Serotype 1 for the presumptive diagnosis of past or current Legionnaires' Disease via qualitative detection of antigen in human urine.
Regulatory Classification
Identification
Haemophilus spp. serological reagents are devices that consist of antigens and antisera, including antisera conjugated with a fluorescent dye, that are used in serological tests to identify Haemophilus spp. directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Haemophilus and provides epidemiological information on diseases cause by these microorganisms. Diseases most often caused by Haemophilus spp. include pneumonia, pharyngitis, sinusitis, vaginitis, chancroid venereal disease, and a contagious form of conjunctivitis (inflammation of eyelid membranes).
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
Predicate Devices
Binax Legionella Urinary Antigen EIA
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 2 1 1999
K991074
000018
## 510(k) SUMMARY
### A. Name of Device and Classification
Bartels Legionella Urinary Antigen ELISA Test Trade Name:
Bartels Legionella Urinary Antigen ELISA Test is an enzyme-Classification: linked immunoassay that detects Legionella pneumophila Serogroup 1 antigen in human urine. This test has been classified as a Class II (performance standards) device, product code MJH (21 CFR 866.3300).
#### B. Legally Marketed Device
Bartels Legionella Urinary Antigen ELISA Test claims substantial equivalence to the Binax Legionella Urinary Antigen EIA.
#### C. Device Description
Bartels Legionella Urinary Antigen ELISA Test is an enzyme-linked immunoassay intended for the qualitative detection of Legionella pneumophila Serogroup 1 antigen in human urine. The kit consists of microelisa stripwells with lid, positive and negative control reagents, conjugate, wash concentrate, colorimetric substrate (2 components) and stop solution. Sufficient materials are provided to perform 96 analyses.
The microelisa wells have been pre-coated with purified rabbit antibodies to Legionella pneumophila Serotype 1 (capture antibody). An undiluted urine specimen (100 µL), or positive or negative controls (100 µL each) are each placed in a single microelisa well followed by the addition of 50 uL of Conjugate (horseradish peroxidase-conjugated rabbit antibodies to L. pneumophila). The loaded microelisa plate is then covered with the lid and incubated for 48 to 52 minutes at 34-37°C followed by 4 cycles of wash/aspiration using diluted Wash Solution (manual or automated wash procedure). Colorimetric substrate (tetramethyl benzidine/H2O2 100 µL/well) is then added and incubated for 10 to 12 minutes at 34-37℃ followed by the addition of Stop Solution (1M Phosphoric acid, 100 uL/well). The stopped plate is then read on a microelisa plate reader at 450 nm against an air blank. For a test to be considered valid, the Negative Control must have an optical density (OD) value of less than 0.100 and the Positive Control must be greater than the Positive Cutoff (pco). The pco is equal to 4X the O.D. value of the Negative Control value. Any specimen with an O.D. value ≥ the pco is considered positive. Any specimen with an O.D. value < the pco is considered negative. Alternatively, the results can be visually read. For this purpose, a visual interpretation card, and written instructions, are provided in the kit for interpretation of results. Any well that produces definite yellow color is considered to be positive.
{1}------------------------------------------------
The following components are provided in each Bartels LUA ELISA kit:
- Micro Wells, 96 break-away coated wells with purified rabbit anti-Legionella 1. pneumophila Serogroup 1, IgG fraction. Ready to use. Store at 2-8°C.
- Conjugate, 6.5 mL bottle of purified rabbit anti-Legionella pneumophila 2. Serogroup 1 IgG conjugated to horseradish peroxidase (HRP) in a buffer with protein stabilizer in a red tracking dye. Ready to use. Store at 2-8°C.
- 3. Wash Concentrate (20X), 100 mL bottle of buffered saline with detergent and thimerosal. 20X concentrate; dilute 1:20 with distilled or deionized water before use. Store at 2-30°C.
- Substrate A, 12mL bottle of tetramethyl benzidine (TMB). Ready to use. 4. Store at 2-8°C.
- ર . Substrate B, 12mL bottle of hydrogen peroxide (H2O2). Ready to use. Store at 2-8°C.
- Stop Solution, 12mL bottle of 1M phosphoric acid. Ready to use. Store at 2-6. 30°C.
- Positive Control. 1.5mL bottle of human urine containing Legionella 7. pneumophila serogroup 1 antigen and preservative. Ready to use. Store at 2-8°C.
- 8. Negative Control, 1.5mL bottle of normal human urine and preservative. Ready to use. Store at 2-8°C.
- இ Visual Interpretation Card - Visual interpretation card displaying 5 levels of vellow color intensity.
- 10. Package Insert, Detailed description of test procedure and interpretation of results.
#### D. Intended Use
Bartels Legionella Urinary Antigen ELISA Test is intended as an adjunct to culture for the presumptive diagnosis of past or current Legionnaires' Disease by qualitative detection of Legionella pneumophila Serogroup 1 antigen in human urine.
#### E. Comparison with the Predicate Device
Bartels Legionella Urinary Antigen ELISA Test, bv the antigen capture sandwich ELISA technique described in this application, is substantially equivalent to the currently marketed Binax Legionella Urinary Antigen EIA which has been cleared by FDA through 510(k) premarket notification. Both kits share common intended use statements, target populations, design/format. kit materials, performance characteristics, absence of risk to patients, specimen type and analytes. The following table (Table 1) summarizes the similarities
{2}------------------------------------------------
| Table 1: Comparison of Devices | | |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | Bartels Legionella Urinary Antigen<br>ELISA Test | Binax Legionella Urinary Antigen<br>EIA |
| Intended Use<br>Statement | "Bartels Legionella Urinary Antigen<br>ELISA Test is intended as an adjunct<br>to culture for the presumptive<br>diagnosis of past or current<br>Legionnaires' Disease by qualitative<br>detection of Legionella pneumophila<br>Serogroup 1 antigen in human urine." | "This kit is an enzyme immunoassay<br>(EIA) based system intended for in<br>vitro diagnostic use to qualitatively<br>detect the presence of Legionella<br>pneumophila Serogroup 1 antigen (L.<br>pneumophila Serogroup 1 antigen) in<br>human urine as an adjunct to culture<br>for the presumptive diagnosis of past<br>or current Legionnaires' Disease." |
| Target Population | Individuals suspected of acute<br>infection by Legionella pneumophila<br>Serotype 1. | Individuals suspected of acute<br>infection by Legionella pneumophila<br>Serotype 1. |
| Design/format | Antigen capture (sandwich) enzyme<br>immunoassay. | Antigen capture (sandwich) enzyme<br>immunoassay. |
| Materials | Microelisa stripwells coated with<br>rabbit polyclonal antibodies to L.<br>pneumophila Serotype 1, HRP<br>polyclonal antibodies to L.<br>pneumophila Serotype 1, positive and<br>negative controls, wash concentrate,<br>tetramethyl benzidine/H2O2 (TMB)<br>colorimetric substrate, stop solution,<br>visual interpretation card, package<br>insert. | Microelisa stripwells coated with<br>rabbit polyclonal antibodies to L.<br>pneumophila Serotype 1, HRP<br>polyclonal antibodies to L.<br>pneumophila Serotype 1, positive and<br>negative controls, wash concentrate,<br>tetramethyl benzidine/H2O2 (TMB)<br>colorimetric substrate, stop solution,<br>package insert. |
| Performance<br>Characteristics<br>(Described in<br>Section 10). | Using culture method as "gold<br>standard":<br>Sensitivity: 94.7%<br>Specificity: 91.1%<br>Accuracy: 92.3% | Using culture method as "gold<br>standard":<br>Sensitivity: 87.2%<br>Specificity: 86.7%<br>Accuracy: 86.9% |
| Risk to patient | No unique issues of safety or<br>effectiveness. | No unique issues of safety or<br>effectiveness. |
| Specimen Type | Voided urine. | Voided urine. |
| Analyte | Legionella pneumophila Serotype 1<br>soluble urinary antigen | Legionella pneumophila Serotype 1<br>soluble urinary antigen |
{3}------------------------------------------------
of Bartels Legionella Urinary Antigen ELISA Test and Binax Legionella Urinary Antigen EIA:
#### Performance Data F.
A clinical study was performed on 274 clinically well-defined urine specimens at a major infectious disease reference laboratory. The study consisted of: a) 94 urine specimens from patients whose respiratory specimens were culture-positive for Legionella pneumophila Serogroup 1; b) 150 urine specimens from who did not have a diagnosis of Legionella pneumophila; and c) 30 urine specimens from normal healthy volunteers. The Bartels Legionella Urinary Antigen ELISA Test had a sensitivity of 94.7%, specificity of 91.1% and accuracy of 92.3%. Substantial equivalence was established between the Bartels LUA and the predicate device.
In this study, a comparison was also made for the Bartels LUA ELISA Test between results obtained with the automated plate reader vs. the use of a visual interpretation card. The sensitivity of the visual interpretation was 92.6%, specificity was 93.9% and accuracy was 93.4%.
#### G. Conclusion
The Bartels Legionella Urinary Antigen ELISA Test is substantially equivalent to the Binax Legionella Urinary Antigen EIA. Furthermore, with the Bartels LUA ELISA Test, there is equivalence between reading the results with an automated plate reader and performing a visual interpretation.
{4}------------------------------------------------
Confidential
## Name of Submitter:
INTRACEL Corporation 1330 Piccard Dr. Rockville, MD 20850 Phone: 301-258-5200 301-296-0076 Fax:
Name of Contact Person:
Sheryl Ruppel Regulatory Affairs Manager INTRACEL Corporation 1330 Piccard Dr. Rockville, MD 20850 Phone: 301-258-5200 Ext. 1089 Fax:
March 30, 1999 Date Prepared:
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the department's name encircling a symbol. The symbol is a stylized representation of a human figure embracing a bird, which is meant to represent the department's mission of protecting the health of all Americans and providing essential human services.
DEC 2 1 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Sheryl Ruppel Regulatory Affairs Manager INTRACEL Corporation 1330 Piccard Drive Rockville, Maryland 20850
Re: K991074 Trade Name: Bartels Legionella Urinary Antigen ELISA Test Regulatory Class: II Product Code: MJH Dated: November 19, 1999 Received: November 19, 1999
Dear Ms. Ruppel:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours.
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Page_1 of_1
# 510(k) Number (if known): K991074
Bartels Legionella Urinary Device Name:
Antigen ELISA Test
## Indications For Use:
Bartels Legionella Urinary Antigen ELISA Test is intended as an adjunct to culture for the presumptive diagnosis of past or current Legionnaires' Disease by qualitative detection of Legionella pneumophila serogroup l antigen in human urine.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Dubois
(Division Sign Off)
Division of Clinical Laboratory Devices
510(k) Number K991074
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.