ZYNERGY ZOLUTION ELECTROPHYSIOLOGY CATHETER, MODEL Z50000
Applicant
Zcv ,Inc.
Product Code
DRF · Cardiovascular
Decision Date
Apr 6, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
Zynergy Zolution Electrophysiology Catheter with Zyp-Lok Extension Set is intended for temporary use in electrophysiology studies for intracardiac stimulation and/or ECG recording.
Device Story
The Zynergy Zolution Electrophysiology Catheter is a diagnostic tool used by physicians during electrophysiology studies. It functions as an electrode recording and pacing catheter, capturing intracardiac electrical signals and delivering stimulation to the heart. The device consists of a catheter (5Fr or 6Fr) with 2-20 platinum electrodes and a reusable Zyp-Lok accessory cable. The catheter is inserted into the cardiac chambers to map conduction pathways and assess arrhythmia vulnerability. The physician observes the recorded electrograms to determine conduction times and identify aberrant pathways. The device is supplied sterile for single-use in a clinical setting. It provides diagnostic information that assists the physician in clinical decision-making regarding cardiac rhythm management.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including tensile, engagement, electrical impedance, and torque transmission performance testing, and biocompatibility testing (cytotoxicity, hemolysis, pyrogenicity, sensitization, etc.) conducted in accordance with ISO 10993 and FDA guidelines.
Indicated for patients undergoing electrophysiology studies requiring intracardiac stimulation and/or ECG recording for the determination of conduction times, identification of aberrant conduction pathways, and assessment of cardiac chamber arrhythmia vulnerability.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
{0}------------------------------------------------
- 6 2000
Image /page/0/Picture/1 description: The image shows the letters ZCV in a bold, sans-serif font. The letters are large and take up most of the frame. A horizontal line underlines the letters, adding a sense of completion to the design.
Zynergy CardioVascular, Inc. 298 Fernwood Avenue 08837-6803 Edison. NJ USA Tel: (732) 225-3800 Fax: (732) 225-4454
Safety & Effectiveness Zynergy Zolution Electrophysiology Catheter with Zyp Lok Accessory Cable Summary: 74 DRF Catheter. Electrode Recording, CFR 870.1220 Classification Name: Electrode Recording Catheter Common / Usual Name: Priscilla Whitehead Cox, Regulatory Affairs Manager Contact: Monday, March 29, 1999 Prepared:
The Zynergy Zolution Electrophysiology Catheter is intended as a primary diagnostic tool for intracardiac electrogram recording and pacing, for the determination of conduction times from one location to another, to identify aberrant conduction pathways and to assess arryhthmia vulnerability of the cardiac chambers.
Initial sizes to be marketed are 5Fr and 6Fr which include lengths of 1-5mm, electrode spacing of 1mm -1cm and varying in the number of electrodes per catheter of 2 - 20. Various curve styles are available in order to meet anatomical constraints and/or physician preference. The Zynergy Zolution, Zyp-Lok Accessory Cable is a reusable, medical grade cable in lengths of 18", 36" and 72" and 4, 10 & 20 conductor configurations.
Zynergy Zolution Electrophysiology Catheter and Zyp-Lok Accessory Cable in single use pouches. Zynergy Zolution Electrophysiology Catheter and Zyp-Lok Accessory Cable are supplied sterile in single use pouches. The packaged product is ETO sterilised with a sterility assurance level of 10°. Ethylene oxide residuals and bacterial endotoxin levels are monitored for compliance with ISO/FDA guidelines.
Performance testing including tensile, engagement, electrical impedence and torque transmission was successfully completed.
Biocompatibility testing, including cytotoxic injection, intracutaneous injection, hemolysis, complement activation, unactivated partial thromboplastin time assay, Lee White Clotting Time, pyrogenicity, intravenous toxicity, sensilisation has been successfully completed per ISO10993 and FDA guidelines.
Zynergy Zolution Electrophysiology Catheter is similar in design, composition and function to the Elecath Defector Series 510(k)# K933450, manufactured by Electro-Catheter Corporation and the Arrow Intracardiac Electrode Catheter Products, 510(k)# K953651 manufactured by Arrow International.
| COMPARATIVE FEATURES | | | |
|------------------------------|-------------------------------------------------------------|------------------------------------------------------------------------|--------------------------------------------------------------------|
| Characteristics | ZCV, Inc<br>Zynergy Zolution<br>Electrophysiology Catheters | Electro-Catheter Co.<br>Detector Series<br>Electrophysiology Catheters | Arrow International<br>Intracardiac Electrode<br>Catheter Products |
| Distal Tip Design | Soft, non-braided,<br>atraumatic tip | Soft, non-braided,<br>atraumatic tip | Soft, non-braided,<br>atraumatic tip |
| Tip Curve Styles | Coumand, Josephson, Damato,<br>Straight | Coumand, Josephson, Damato | Coumand, Josephson, Damato, Straight |
| Tubing Materials | Polyamide TPE /<br>SS Wire Braid | Polyamide TPE /<br>SS Wire Braid | Polyurethane |
| Electrode Materials | Platinum | Platinum | Gold/Platinum |
| French Size | 3/5/6 | 5/6 | 2/5/6 |
| Electrode Spacing | 1-10mm | 1-10mm | 1-10mm |
| Connector Type | Positive Locking Multi-Pin | Positive Locking Multi-Pin | Positive Locking Multi-Pin |
| Tip Electrode length | 2-5mm | 2-5mm | Not specified |
| Electrode Number | 2-20 | 2-10 | 2/4/6/10 |
| Proximal Electrode<br>Length | 1-5mm | 1-5mm | Not specified |
| Useable Length | 110cm | 110cm | 110cm |
| Extension Cable | 18"/3'/6' | 3'/6' | 18"/3'/6' |
| Packaging | Blister/Poly/Tyvek | Blister/Poly/Tyvek | Blister/Poly/Tyvek |
| Sterilisation Method | ETO | ETO | ETO |
Tunarni Cardin\/acoular Inc
7 unarau 7 alution Elastranhicinlany Cathato
510(k) Submission
II-2
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR = 6 2000
Ms. Priscilla Whitehead Cox Zynergy CardioVascular, Inc. 298 Fernwood Ave. Edison, NJ 08837
Re: K991060 Zynergy Zolution Electrophysiology Catheter Model Z5000 and Cable Model Z9000 Regulatory Class: II (two) Product Code: DRF Dated: January 21, 2000 Received: January 24, 2000
Dear Ms. Cox:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice reguirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act
{2}------------------------------------------------
Page 2 - Ms. Pricilla Whitehead Cox
for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
pax maynehen
James E. Dillard III Director
Division of Cardiovascular, Respiratory and Neuroloqical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page lof l
510(k) Number (if known): K991060
## Device Name: Zynergy Zolution Electrophysiology Catheter with Zyp-Lok Extension Set
Indications For Use:
.
Zynergy Zolution Electrophysiology Catheter with Zyp-Lok Extension Set is intended for temporary use in electrophysiology studies for intracardiac stimulation and/or ECG recording.
## (PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Reggantiche Counter Use____________________________________________________________________________________________________________________________ Prescription Use_ V and Neurological Devices (Per 21 CFR 801.109) (Optional Format 1-2-96) 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.