K990968 · Biosound Esaote, Inc. · LNH · May 27, 1999 · Radiology
Device Facts
Record ID
K990968
Device Name
E-SCAN
Applicant
Biosound Esaote, Inc.
Product Code
LNH · Radiology
Decision Date
May 27, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The E-Scan is intended for diagnostic nuclear magnetic resonance imaging of the hip, leg, knee, ankle, foot, shoulder, elbow, forearm, wrist and hand. The device produces transverse, sagittal, coronal and oblique cross-sectional images, displaying the internal structure of the limbs and joints being imaged. The images that are produced correspond to the spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon the MR parameters (spin-lattice relaxation time (T1), spinspin relaxation time (T2), nuclei density, flow velocity and chemical shift). If interpreted by a medical expert, these images can provide diagnostically useful information. The system preliminarily named "Artoscan S" is going to be commercialized under the trade name "E-Scan". The changes performed, with respect to the cleared version, are due to the optimization process of the system, in particular in relation to an improvement of the accessibility inside the gantry and therefore toward a greater patient comfort. The intended use of the modified device has not changed as a result of the modifications nor such modifications alter its fundamental scientific technology.
Device Story
E-Scan is a magnetic resonance imaging (MRI) system for limb and joint diagnostics. It captures proton density and relaxation signals (T1, T2) to generate transverse, sagittal, coronal, and oblique cross-sectional images. Operated by trained professionals in clinical settings, the device provides visual data for medical experts to interpret internal anatomical structures. Modifications from the predicate Artoscan S focus on gantry accessibility and patient comfort, alongside industrialization improvements to magnet homogeneity and RF shielding. The system aids clinical decision-making by providing non-invasive diagnostic imaging of musculoskeletal regions.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Magnetic resonance diagnostic device. Features integrated RF shielding or optional shielding box. Magnet homogeneity < + 50 ppm (pk pk) on 140 mm DSV. Gradient rise time 0.8 ms (0 to 20 mT/m). Operates via nuclear magnetic resonance principles using T1, T2, proton density, flow velocity, and chemical shift parameters.
Indications for Use
Indicated for diagnostic nuclear magnetic resonance imaging of the hip, leg, knee, ankle, foot, shoulder, elbow, forearm, wrist, and hand in patients requiring cross-sectional imaging of limbs and joints.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K990968
MAY 2 7 1999
510(k) Summary E-Scan Biosound Esaote
# 510(k) Summary
The following safety and effectiveness summary has been prepared pursuant to requirement for 510(k) summaries specified in 21CFR¶807.92(a).
807.92(a)(1)
### Submitter Information
| Colleen Hittle, Official Correspondent | |
|----------------------------------------|----------------|
| 8000 Castleway Drive | |
| Indianapolis, IN 46250 | |
| Phone: | (317) 849-1916 |
| Facsimile: | (317) 577-9070 |
| Contact Person: | Colleen Hittle |
| Date: | March 10, 1999 |
807.92(a)(2)
| Trade Name: | E-Scan |
|-------------------------|--------------------------------------------|
| Common Name: | Magnetic resonance diagnostic device |
| Classification Name(s): | System, Nuclear Magnetic Resonance Imaging |
| Classification Number: | |
807.92(a)(3)
Predicate Device(s)
| Esaote | Artoscan S | K981358 |
|--------|------------|---------|
|--------|------------|---------|
Additional Substantial Equivalence Information is provided in the following substantial Equivalence Comparison Table.
{1}------------------------------------------------
510(k) Summary E-Scan Biosound Esaote
#### 807.92(a)(5)
#### Intended Use(s)
The E-Scan is intended for diagnostic nuclear magnetic resonance imaging of the hip, leg, knee, ankle, foot, shoulder, elbow, forearm, wrist and hand. The device produces transverse, sagittal, coronal and oblique cross-sectional images, displaying the internal structure of the limbs and joints being imaged. The images that are produced correspond to the spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon the MR parameters (spin-lattice relaxation time (T1), spinspin relaxation time (T2), nuclei density, flow velocity and chemical shift). If interpreted by a medical expert, these images can provide diagnostically useful information.
The system preliminarily named "Artoscan S" is going to be commercialized under the trade name "E-Scan". The changes performed, with respect to the cleared version, are due to the optimization process of the system, in particular in relation to an improvement of the accessibility inside the gantry and therefore toward a greater patient comfort. The intended use of the modified device has not changed as a result of the modifications nor such modifications alter its fundamental scientific technology.
{2}------------------------------------------------
510(k) Summary E-Scan Biosound Esaote
| Item | Artoscan S | E-SCAN | Comments |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| # slices per<br>echo | 1 - 128 | 1 – 96 | |
| Homogeneity | < + 30 ppm on 120 mm<br>DSV | < + 50 ppm (pk pk) on 140<br>mm DSV | Final result allows to<br>enlarge the Field of View |
| Fringe field<br>(0,5 mT line) | 0,5 mT line: 140 cm<br>from the center of the<br>magnet<br>X axis: 140 cm front; 90 cm rear;<br>Y axis: 120 cm<br>right/left;<br>Z axis: 90 cm<br>top/bottom | 0,5 mT line: 150 cm from<br>the center of the magnet<br>X axis: 150 cm front, 120<br>rear;<br>Y axis: 130 cm right/left<br>Z axis: 150 cm top/bottom | Variations due to the<br>industrialization of the<br>magnet |
| RF Shielding | Self contained RF<br>shielding (up to 30<br>dBμV/m)<br>External shielding (up to<br>100 dBμV/m) | Integrated RF shielding<br>(up to 30 dBμV/m)<br>Shielding Box (up to 60<br>dBμV/m)<br>External shielding (up to<br>100 dBμV/m) | The term "integrated"<br>represents better the<br>concept than "self-<br>contained"<br>A shielding box has been<br>added as intermediate<br>solution, in terms of both<br>performances and costs,<br>between the integrated RF<br>shielding and the<br>traditional RF cage (as this |
| | | | latter is always possible<br>but more expensive). |
| Gradients' rise<br>time | 0.5 ms from 0 to 20 mT<br>or from -20 to 0 mT | 0.8 ms from 0 to 20 mT/m<br>from 0 to 99% | More precise<br>characterization of the<br>data |
| Sequences | | The sequence preliminary<br>indicated as GEFS has<br>been renamed into GE-<br>STIR | |
| Item | Artoscan S | E-SCAN | Comments |
| Anatomical<br>regions | Limbs: in particular Hip,<br>Knee, Leg, Ankle, Foot,<br>Hand, Wrist, Forearm,<br>Elbow and Shoulder | Limbs: in particular Hip,<br>Knee, Leg, Ankle, Foot,<br>Hand, Wrist, Forearm,<br>Elbow, Arm and Shoulder | It was possible to examine<br>arms also before but it was<br>necessary to select <b>Other</b><br>and to introduce the Arm<br>data. Now arm is included<br>in the <b>standard</b> and most<br>common districts to be<br>examined with E-scan |
| Installation<br>area | 3.7 x 3.7 sq. m. | Integrated shielding: 3.7 x<br>4 mt<br><br>Shielding box: 4 x 4.5 sq.<br>m. (min height 2.4 m) | Variation due to the<br>industrialization of the<br>system. |
### Comparison Chart for Substantial Equivalence
{3}------------------------------------------------
510(k) Summary E-Scan Biosound Esaote
## Comparison Chart for Substantial Equivalence Continued
.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 27 1099
Colleen Hittle Official Correspondent Biosound Esaote, Inc. 8000 Castleway Drive Indianapolis, Indiana 46250 RE:
K990968 E-Scan Magnetic Resonance Diagnostic Device Dated: March 19, 1999 Received: March 23, 1999 Regulatory Class: II 21 CFR 892.1000/Procode: 90 LNH
Dear Ms. Hittle:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications For Use
510(k) Number (if known):
K 99 0968
Device Name:
E-Scan
Indications for Use:
The E-Scan is intended for diagnostic nuclear magnetic resonance imaging of the hip, leg, knee, ankle, foot, shoulder, elbow, forearm, wrist and hand. The device produces transverse, sagittal, coronal and oblique cross-sectional images, displaying the internal structure of the limbs and joints being imaged. The images that are produced correspond to the spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon the MR parameters (spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and chemical shift). If interpreted by a medical expert, these images can provide diagnostically useful information.
The system preliminarily named "Artoscan S" is going to be commercialized under the trade name "E-Scan". The changes performed, with respect to the cleared version, are due to the optimization process of the system, in particular in relation to an improvement of the accessibility inside the gantry and therefore toward a greater patient comfort. The intended use of the modified device has not changed as a result of the modifications nor such modifications alter its fundamental scientific technology.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
✓
OR
Over-The-Counter Use
David U. Bergman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological I 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.