NEWLIFE DISH (ETCHED OR NON-ETCHED, WITH OR WITHOUT LID), MODELS 14-52, 1452E, 1452EL, 1452L
Applicant
Humagen Fertility Diagnostics, Inc.
Product Code
MQK · Obstetrics/Gynecology
Decision Date
Jul 14, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6160
Device Class
Class 2
Indications for Use
The NewLife Dish's intended use is to hold zygotes/embryos during micromanipulation or other tissue culture procedures in the IVF laboratory. New Life Dishes are used in the tissue culture techniques performed by embryologists when injecting a single sperm into an egg, or assisting an embryo in hatching prior to reimplantation. The dishes are disposable tissue culture labware used in procedures that have been developed to aid infertile couples achieve pregnancy. Specifically, the ICSI procedure is beneficial in cases where the male infertility is impaired.
Device Story
NewLife Dish is a 60 mm polystyrene tissue culture dish used in IVF laboratories. Available in etched or unetched configurations to facilitate embryo tracking during micromanipulation. Operated by embryologists for procedures such as intracytoplasmic sperm injection (ICSI) and assisted hatching. Provides a controlled environment for zygotes/embryos during culture. Disposable, single-use design. Benefits patients by supporting assisted reproduction techniques to address infertility.
Clinical Evidence
Bench testing only. Toxicity assessed via two-cell mouse embryo bioassay; nontoxicity defined as treated group blastocyst development within 10% of control group. Endotoxin levels verified via Limulus Amebocyte Lysate assay (<20 EU/device). Sterility validated via gamma radiation (SAL 10^-6) and microbial barrier testing of packaging.
Technological Characteristics
Polystyrene construction; 60 mm diameter; etched or unetched surface patterns. Sterilized via gamma radiation (SAL 10^-6). Packaging consists of heat-sealed INTEGRA Peel K paper. Standalone labware.
Indications for Use
Indicated for holding zygotes/embryos during IVF laboratory micromanipulation or tissue culture procedures, including ICSI and assisted hatching, to aid infertile couples, particularly those with male factor infertility.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Submission Summary (Full Text)
{0}------------------------------------------------
# 12990941
### 510(k) SUMMARY
Submitter's Name: Humagen Fertility Diagnostics, Inc.
Address: 2400 Hunter's Way Charlottesville, VA 22911
Telephone #: (804) 979-4000
FAX #: (804) 295-5912
Contact person: Cindy Showalter
Date summary prepared: May 28, 1999
{1}------------------------------------------------
#### Device name:
Classification name: Assisted reproduction labware (per CFR# 884.6160)
Common/Usual name: ICSI Tissue Culture Dish
NewLife™ Dish Proprietary names:
#### Substantial Equivalence:
Substantial equivalence is being supported by the Federal Register Notice Final Rule entitled "Obstetric and Gynecologic Devices; Reclassification and Classification of Medical Devices Used for In Vitro Fertilization and Related Assisted Reproduction Procedures. This equivalence is supported by the SUMMARY statement:
> Upon the effective date, the Federal Register document may be cited in the absence of an existing predicate device which would be used to support substantial equivalence.
#### Description of Device:
NewLife Dishes are fabricated from polysterene and are 60 mm. diameter. Dishes are made with or without clear lids, etched or unetched. Etched dishes have a center circle with 8 smaller circles inside the larger circle, and a 9th circle inside the 8 circles. There are numbers 1 through 8 outside the large circle, with the number 1 at the top of the dish. Unetched dishes have center circles, but no other circles are etched on the dishes. Dishes are disposable, intended for one time use only.
#### Testing Procedures:
Each lot of New Life dishes is mouse embryo tested for toxicity. We use a two-cell mouse embryo bioassay. A two-cell mouse embryo test was performed as follows: 25 ul tissue culture medium containing serum albumin and penicillin/streptomycin was overlaid with 1 ml mineral oil and equilibrated overnight in 5.5% CO2. On the morning of the embryo harvest, medium was exposed to product and embryos were cultured in the media beginning at the two-cell stage. For NewLife Dishes, the MEA is performed in the test dish, providing continuous exposure. Control droplets of medium were not exposed to product before receiving embryos. The combined percent of embryos developing to expanded and/or hatching blastocysts were assessed at 72 hours of culture. Embryos exposed to the dishes are considered nontoxic if the number of expanded and/or hatching blastocysts in the treated group is within 10% of the control group, and no contamination of the culture
{2}------------------------------------------------
has occurred. Greater than 80% hatching/expanded blastoycysts in the control group indicates a valid assay.
Each lot of dishes is also tested for endotoxin levels using the Limulus Amebocyte Lysate assay. The level of endotoxin units per device must be less than 20 to be considered acceptable.
Each lot of dishes is sterilized by gamma radiation with a sterility assurance level of 106. Quarterly dose audits are performed to insure the sterility assurance level is maintained or following each batch sterilization if less often than quarterly.
The packaging to maintain device sterility consists of a pouch made from INTEGRA® Peel K packaging. This is a reinforced, heavy-weight 70 lb. paper coated with a sealant, designed for medical device packaging applications. The package is heat-sealed. This packaging has been tested to ensure a microbial barrier, and has been validated to maintain sterility for a shelf life of 2 years.
#### Intended use statement:
The NewLife Dish's intended use is to hold zygotes/embryos during micromanipulation or other tissue culture procedures in the IVF laboratory.
New Life Dishes are used in the tissue culture techniques performed by embryologists when injecting a single sperm into an egg, or assisting an embryo in hatching prior to reimplantation.
The dishes are disposable tissue culture labware used in procedures that have been developed to aid infertile couples achieve pregnancy. Specifically, the ICSI procedure is beneficial in cases where the male infertility is impaired.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol of medicine and health. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JUL 14 1999
Ms. Cindy Showalter Quality Assurance Manager Humagen Fertility Diagnostics, Inc. 2400 Hunter's Way Charlottesville, VA 22911
Re: K990941 Newlife Dish Dated: May 31, 1999 Received: June 2, 1999 Regulatory Class: II 21 CFR §884.6160/Procode: 85 MQK
Dear Ms. Showalter:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compilance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also. please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
| 510 (k) NUMBER (IF KNOWN) : | KG90941 |
|-----------------------------|--------------|
| DEVICE NAME: | NewLife Dish |
INDICATIONS FOR USE:
The NewLife Dish s indication for use is to hold zygotes/embryos during micromanipulation or other tissue culture procedures in the IVF laboratory.
NewLife Dishes are used in the tissue culture techniques performed by embryologists when injecting a single sperm into an egg, or assisting an embryo in hatching prior to re-implantation.
The dishes are disposable tissue culture labware used in procedures that have been developed to aid infertile couples achieve pregnancy. Specifically, the ICSI procedure is beneficial in cases where the male infertility is impaired.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
> Prescription Use (Per 21 CFR 801.109)
ાર
Over-The-Counter-Use (Optional Format 1-2-96)
Rati Palys
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.