K990852 · CooperSurgical, Inc. · KNS · Jun 11, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K990852
Device Name
COOPERSURGICAL INFRARED COAGULATOR
Applicant
CooperSurgical, Inc.
Product Code
KNS · Gastroenterology, Urology
Decision Date
Jun 11, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CooperSurgical IRC is indicated for the treatment of hemorrhoids, especially bleeding, non-prolapsed internal hemorrhoids, as a means of hemostasis in the treatment of hemorrhoids by routine excision, or to coagulate external thrombotic hemorrhoids that have been removed ("collectively referred to as hemorrhoids") through the coagulative necrosis of the mucosa proximal to the base of the hemorrhoid.
Device Story
The CooperSurgical InfraRed Coagulator (IRC) is a medical device used by physicians to treat hemorrhoids via thermal coagulation. The device consists of a console, handpiece, light guide, and disposable contact tip. The physician applies the contact tip to the target tissue with light mechanical pressure and activates the device. The system delivers short pulses of visible and infrared light to the tissue, inducing controlled coagulative necrosis. A built-in digital timer automatically deactivates the lamp based on a preselected pulse duration, which determines the depth of tissue necrosis. The device is used in a clinical setting to achieve hemostasis or treat hemorrhoids by routine excision. By providing precise, controlled thermal energy, the device facilitates tissue destruction and hemostasis, potentially reducing bleeding and improving patient outcomes during hemorrhoid treatment.
Clinical Evidence
No clinical data provided; substantial equivalence is based on the predicate devices' established safety and effectiveness and bench testing of the contact tips.
Technological Characteristics
System includes console, handpiece, light guide, and disposable PVC contact tips. Operates via visible and infrared light pulses for thermal coagulation. Features a digital timer for automatic pulse duration control. Device is a Class II medical device (21 CFR 876.4300).
Indications for Use
Indicated for treatment of hemorrhoids, including bleeding, non-prolapsed internal hemorrhoids, and external thrombotic hemorrhoids that have been removed, in patients requiring hemostasis or coagulative necrosis of mucosa.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
CooperSurgical's InfraRed Coagulator (for genital condylomas and general warts)
Submission Summary (Full Text)
{0}------------------------------------------------
## JUN 11 1999
### 510(K) SUMMARY FOR COOPERSURGICAL, INC.'S INFRARED COAGULATOR
K990852
# Submitter's Name, Address, Telephone Number, and Contact Person
CooperSurgical, Inc. 15 Forest Parkway Shelton, Connecticut 06484
| Contact: | Deb Pekar<br>Manager, Regulatory Affairs, CooperSurgical, Inc. |
|------------|----------------------------------------------------------------|
| Phone: | (203) 929-6321 |
| Facsimile: | (203) 925-0135 |
#### Date Prepared: March 15, 1999
#### Name of the Device
CooperSurgical InfraRed Coagulator
#### Common or Usual Name
Infrared Coagulator
#### Predicate Devices
- (1) Redfield Corporation's Redfield Infrared Coagulator for the treatment of hemorrhoids, especially bleeding, non-prolapsed internal hemorrhoids, as a means of hemostasis in the treatment of hemorrhoids by routine excision, or to coagulate external thrombotic hemorrhoids that have been removed; and
- (2) CooperSurgical's InfraRed Coagulator for the treatment of genital condylomas (condyloma accuminata) and general warts.
#### Intended Use
The CooperSurgical IRC is indicated for the treatment of hemorrhoids, especially bleeding, non-prolapsed internal hemorrhoids, as a means of hemostasis in the treatment of hemorrhoids by routine excision, or to coagulate external thrombotic hemorrhoids that have been removed ("collectively referred to as hemorrhoids") through the coagulative necrosis of the mucosa proximal to the base of the hemorrhoid.
{1}------------------------------------------------
#### Principles of Operation
The Cooper IRC delivers short pulses of visible and infrared light through a small contact tip applicator that is applied to the tissue. This light causes thermal coagulation that results in tissue necrosis. The user sets the pulse duration control knob depending on the depth of the tissue necrosis required. The depth of coagulation is directly related to the pulse length delivered to a given area of tissue. The area(s) to be treated may be locally anesthetized at the physician's discretion.
Page 2 of 2
K990852
Contact photocoagulation requires direct contact of the entire optical window of the disposable contact tip. The physician applies light mechanical pressure to the optical window against the tissue to be treated before depressing the activation trigger of the hand piece, waiting at least five seconds between exposures. To achieve coagulation, the user depresses the activation trigger. A built-in digital timer deactivates the lamp automatically according to the pulse duration setting preselected by the physician.
#### Technical Characteristics
The CooperSurgical IRC consists of the following main components: (1) a console unit; (2) a hand piece; (3) a removable light guide; and (4) a disposable contact tip. The CooperSurgical IRC is essentially the same device as Redfield Corporation's Infrared Coagulator that has already been cleared by FDA for the treatment of hemorrhoids. With the exception of four features, the CooperSurgical IRC is identical to the CooperSurgical IRC that is cleared for the treatment of genital condylomas and general warts.
#### Summary of the Basis for the Finding of Substantial Equivalence
The safety and effectiveness of the CooperSurgical IRC for the treatment of hemorrhoids is based on FDA's clearance of: (1) Redfield's Infrared Coagulator for the treatment of hemorrhoids; (2) CooperSurgical's InfraRed Coagulator for the treatment of genital condylomas and general warts; and (3) CooperSurgical's testing of the PVC contact tips.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 1 1999
CooperSurgical, Inc. c/o Mr. Jonathan S. Kahan Hogan & Hartson, L.L.P. 555 Thirteenth Street, N.W. Washington, D.C. 20004
Re: K990852 CooperSurgical InfraRed Coaqulator Dated: March 15, 1999 Received: March 15, 1999 Requiatory Class: II 21 CFR §876.4300/ Procode : 78 KNS
Dear Mr. Kahan:
We have reviewed vour Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regult in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compiance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: CooperSurgical InfraRed Coagulator
Indications for Use:
The CooperSurgical InfraRed Coagulator is indicated for the treatment of hemorrhoids, especially bleeding, non-prolapsed internal hemorrhoids, as a means of hemostasis in the treatment of hemorrhoids by routine excision, or to coagulate external thrombotic hemorrhoids that have been removed through coagulative necrosis of the mucosa proximal to the base of the hemorrhoid.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use | (Per 21 C.F.R. 801.109)
(Division Sign-Off)
and Radiological Dev 510(k) Number.
Division of Reproductive, Abdominal, B
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.