← Product Code [JMM](/productcode/JMM) · K990843

# SERUM CROSSLAPS ONE STEP ELISA (K990843)

_Osteometer Biotech A/S · JMM · Jul 9, 1999 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K990843

## Device Facts

- **Applicant:** Osteometer Biotech A/S
- **Product Code:** [JMM](/productcode/JMM.md)
- **Decision Date:** Jul 9, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1400
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Serum CrossLaps™ One Step ELISA assay is intended for in vitro diagnostic use as an indication of human bone resorption and may be used as an aid in : Monitoring bone resorption changes of A. 1.) Anti-resorptive therapies in postmenopausal women: a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs b) Bisphosphonate therapies 2. ) Anti-resorptive therapies in individuals diagnosed with osteopenia; a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs b) Bisphosphonate therapies B. Predicting skeletal Response (Bone Mineral Density) in postmenopausal woman under going anti-resorptive therapies a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs b) Bisphosphonate therapies

## Device Story

Serum CrossLaps™ One Step ELISA is an in vitro diagnostic immunoassay for quantitative measurement of Type I Collagen C-Telopeptide in human plasma and serum. Device utilizes non-competitive binding protein assay format. Samples, standards, and controls are pipetted into streptavidin-coated microtitre wells; mixture of biotinylated antibody and peroxidase-conjugated antibody is added. Complex of antigen, biotinylated antibody, and peroxidase-conjugated antibody binds to streptavidin surface. After incubation and washing, chromogenic substrate is added; color reaction stopped with sulfuric acid. Absorbance measured at 450 nm. Used in clinical laboratories; operated by laboratory technicians. Output provides quantitative measurement of bone resorption markers, aiding clinicians in monitoring therapy efficacy and predicting bone mineral density changes in patients on anti-resorptive treatments.

## Clinical Evidence

Clinical data included in the submission to support the intended use for monitoring bone resorption and predicting skeletal response in postmenopausal women and osteopenic patients undergoing anti-resorptive therapies.

## Technological Characteristics

Enzyme-linked immunosorbent assay (ELISA); non-competitive binding protein assay. Uses two specific monoclonal antibodies against ß-isomerized aspartic acid residue in EK AHD-B-GGR sequence. Requires cross-linked chains for signal. Microtitre wells coated with streptavidin. Detection via peroxidase-conjugated antibody and chromogenic substrate; absorbance measured at 450 nm.

## Regulatory Identification

A hydroxyproline test system is a device intended to measure the amino acid hydroxyproline in urine. Hydroxyproline measurements are used in the diagnosis and treatment of various collagen (connective tissue) diseases, bone disease such as Paget's disease, and endocrine disorders such as hyperparathyroidism and hyperthyroidism.

## Predicate Devices

- CrossLaps™ ELISA ([K972788](/device/K972788.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a sequence of handwritten characters and numbers. The sequence starts with the letter 'K', followed by the numbers '990843'. The handwriting style appears casual, with some variations in stroke thickness.

# Summery of Safety and Effectiveness:

| Name of Device:        | Serum CrossLaps™ One Step ELISA                                  |
|------------------------|------------------------------------------------------------------|
| Classification Name:   | Enzyme immunoassay, Type I Collagen C-Telopeptide                |
| Analyte Code and Name: | JMM                                                              |
| Submitter:             | Osteometer BioTech A/S<br>Herlev Hovedgade 207<br>DK-2730 Herlev |
| Date:                  | March 9th, 1999                                                  |

The Serum CrossLaps™ One Step ELISA kit was developed for the quantitative measurement of Type I Collagen C-Telopeptide (CrossLaps™) in human plasma and serum.

The EIA format is a non-competitive binding protein assay.

The Serum CrossLaps 114 One Step ELISA is based on two highly specific monoclonal antibodies against the amino acid sequence of EK AHD-B-GGR, where the aspartic acid residue (D) is ß-isomerized. In order to obtain a specific signal in the Serum CrossLaps 114 One Step ELISA, two chains of EK AHD-S-GGR must be cross-linked

Standards, control, or unknown serum samples are pipetted into the appropriate microtitre wells coated with streptavidin, followed by application of a mixture of a biotinylated antibody and a peroxidase-conjugated antibody. Then, a complex between CrossLaps 110 antigens, biotinylated antibody and peroxidase-conjugated antibody is generated, and this complex binds to the streptavidin surface via the biotinylated antibody. Following the one-step incubation at room temperature, the wells are emptied and washed. A chromogenic substrate is added and the colour reaction is stopped with sulfuric acid. Finally, the absorbance is measured at 450 nm.

{1}------------------------------------------------

This pre-market notification demonstrates that the Serum CrossLaps™ One Step ELISA for the quantitative measurement of Type I Collagen C-telopeptides in plasma and serum is substantially equivalent to CrossLaps™ ELISA that was cleared by FDA in a previous submission (#K972788).

This pre-market notification includes clinical data demonstrating that

The CrossLaps™ ELISA assay is intended for in vitro diagnostic use as an indication of human bone resorption and may be used as an aid in :

#### Monitoring bone resorption changes of A.

- 1.) Anti-resorptive therapies in postmenopausal women:
- a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs
- b) Bisphosphonate therapies

### 2.) Anti-resorptive therapies in individuals diagnosed with osteopenia;

- a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs
- b) Bisphosphonate therapies

### B. Predicting skeletal Response (Bone Mineral Density) in postmenopausal woman under going anti-resorptive therapies

- a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs
- b) Bisphosphonate therapies

...

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized symbol that resembles a triple helix or a series of flowing lines, often interpreted as representing the department's mission related to health and well-being.

JUL - 9 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Eva Gamwell Henriksen Head of Regulatory Affairs and Quality Assurance OSTEOMETER BIOTECH A/S HERLEV HOVEDGADE 207 DK-2730 HERLEV, DENMARK

Re: K990843

> Trade Name: Serum CrossLaps™ One Step ELISA Regulatory Class: I reserved Product Code: JMM Dated: May 25, 1999 Received: May 27, 1999

Dear Ms. Henriksen:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

### Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Gutman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## 510(k) Number (if known):_

Serum CrossLaps™ One Step ELISA Device Name:

### Indications for use:

The Serum CrossLaps™ One Step ELISA assay is intended for in vitro diagnostic use as an indication of human bone resorption and may be used as an aid in :

K990843

- Monitoring bone resorption changes of A.
	- 1.) Anti-resorptive therapies in postmenopausal women:
	- a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs
	- b) Bisphosphonate therapies
	- 2. ) Anti-resorptive therapies in individuals diagnosed with osteopenia;
	- a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs
	- b) Bisphosphonate therapies
- B.

Predicting skeletal Response (Bone Mineral Density) in postmenopausal woman under going anti-resorptive therapies

- a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs
- b) Bisphosphonate therapies

Sean Cooper

(Division Sign-Off) (Division of Clinical Laborator) 89908

1) Number PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

(Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

Prescription use
(Per 21 CFR 801.109)

✓

Over-The-Counter Use

(Optional Format 1-2-96)

0011

---

**Source:** [https://fda-staging.innolitics.com/device/K990843](https://fda-staging.innolitics.com/device/K990843)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
