Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal OB/GYN, Abdominal, Intraoperative (abdominal organs and vascular), Small Organs (breast, thyroid, testicle), Pediatric, Trans-vaginal, Peripheral Vessel, Cardiac, Musculo-skeletal (conventional), Neonatal Cephalic, Trans-Rectal. Typical examinations performed using C1 system are: General abdominal and pelvic studies including organ surveys, blood flow assessment, and retroperitoneal cavity studies. Study of small parts and superficial structures including breasts, shoulders, thyroid, and the abdominal wall. Pediatric scans of organs, superficial, and bony structures. Monitoring procedures for infertility studies (other than in vitro fertilization). First, second and third trimester pregnancy studies. Neonatal head studies. General cardiac studies in adults. Prostate, prostate biopsy guidance, rectal wall studies.
Device Story
C1 is a portable, software-controlled diagnostic ultrasound system; acquires ultrasound data via curved or linear transducers; displays 2D, M-mode, and Amplitude Doppler images. System automatically optimizes operating conditions (frame rate, line density, frequency) based on user inputs. Clinicians use the system for anatomical measurements and clinical diagnostic analysis. Output includes mechanical index (MI) and thermal index (TI) for acoustic monitoring. Device aids in clinical decision-making by providing real-time visualization of internal structures and blood flow; benefits patients through non-invasive diagnostic imaging.
Clinical Evidence
Bench testing only. The device conforms to electromechanical safety standards (EN 60601-1, UL 2601-1, IEC 1157) and acoustic output requirements (AIUM/NEMA 1992). No clinical data provided.
Technological Characteristics
Portable ultrasound system; piezoelectric transducer technology; 2.0-7.0 MHz frequency range; linear and curved array transducers. Modes: 2D, M-mode, Amplitude Doppler. Connectivity: Standalone. Safety standards: EN 60601-1, UL 2601-1, C22.2 No. 601.1, IEC 1157, EN 60601-1-2. Software-controlled optimization of frame rate, line density, and center frequency.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, intraoperative, small organ (breast, thyroid, testicle), pediatric, trans-vaginal, peripheral vessel, cardiac, musculoskeletal, neonatal cephalic, and trans-rectal applications. Contraindicated for in vitro fertilization and percutaneous umbilical blood sampling.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
3/26/99 510(k) Premarket Notification
Modified Ultrasound Generator - February 20, 1999
K990806
C1 Ultrasound System
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
#### 1) Submitter's name, address, telephone number, contact person:
Howard Holman Director, Program Management and Compliance SonoSite, Inc. 19807 North Creek Parkway, Suite 200 Bothell, WA 98011-8214 (425) 487-7602
Date prepared: February 16, 1999
- 2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
Common/ Usual Name
Diagnostic Ultrasound System with Accessories
# Proprietary Name
C1 Ultrasound System (subject to change)
Classification Names
| Ultrasonic Pulsed Doppler Imaging System | 90-IYN | 892.1550 |
|------------------------------------------|--------|----------|
| Diagnostic Ultrasonic Transducer | 90-ITX | 892.1570 |
| Ultrasonic Pulsed Echo Imaging System | 90-IYO | 892.1560 |
#### 3) Identification of the predicate or legally marketed device:
SonoSite, Inc. believes that C1 ultrasound system is substantially equivalent to the currently marketed ATL HDI® 5000 and Medison SA88 Plus diagnostic ultrasound systems and the previously cleared Cozumel ultrasound system.
{1}------------------------------------------------
#### 4) Device Description:
C1 is a general purpose, highly portable, software-controlled. diagnostic ultrasound system. Its function is to acquire ultrasound data and display it on a monitor in 2D, M-mode, and Amplitude Doppler or in a combination of modes. C1 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information used for clinical diagnostic purposes. C1 has an output display with two basic indices, a mechanical index (MI) and a thermal index (TI). One index is automatically displayed. MI is displayed in B-mode. TI is displayed in all other modes.
The C1 system is designed to accept a curved or linear transducer. All actions affecting the performance of the transducer are activated from the main system control panel.
The C1 system is designed to accept transducers of the following types and frequency:
> frequency range: 2.0 - 7.0 MHz transducer types: Linear array Curved linear array
Specific operating conditions (frame rate, line density, center frequency, number of active elements etc.) are automatically optimized by the system software in response to user inputs such as depth, exam type, transducer, and optimize.
C1 has been designed to meet the following electromechanical safety standards:
- EN 60601-1 (IEC 601-1,) European Norm, Medical Electrical . Equipment
- UL 2601-1, Underwriters Laboratories Standards, Medical Electrical . Equipment
- C22.2 No. 601.1, Canadian Standards Association, Medical Electrical . Equipment
- CEI/IEC 1157:1992, International Electrotechnical Commission, . Requirements for the declaration of the acoustic output of medical diagnostic ultrasonic equipment
- EN 60601-1-2 (IEC 601-1-2,) European Norm, Collateral Standard: . Electromagnetic Compatibility
{2}------------------------------------------------
#### 5) Intended Use:
C1 intended uses as defined FDA guidance documents are:
- Fetal OB/GYN .
- Abdominal .
- Intraoperative (abdominal organs and vascular) ●
- Small Organs (breast, thyroid, testicle) ●
- Pediatric .
- Trans-vaginal .
- Peripheral Vessel ●
- Cardiac ●
- Musculo-skeletal (conventional) ●
- Neonatal Cephalic ●
- Trans-Rectal
Typical examinations performed using C1 system are:
- . General abdominal and pelvic studies including organ surveys, blood flow assessment, and retroperitoneal cavity studies.
- . Study of small parts and superficial structures including breasts, shoulders, thyroid, and the abdominal wall.
- Pediatric scans of organs, superficial, and bony structures. ●
- Monitoring procedures for infertility studies (other than in vitro fertilization).
- First, second and third trimester pregnancy studies. ●
- Neonatal head studies. .
- General cardiac studies in adults. .
- Prostate, prostate biopsy guidance, rectal wall studies
#### 6) Technological Characteristics:
This device operates identical to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as a 2D and M-mode images. Doppler shift caused by blood flow is displayed as Color Flow, or as spectrum analysis. The modes of this device (2D, M-mode, and Amplitude Doppler) are the same as predicate devices identified in item 3. Transducer patient contact materials are biocompatible.
This device conforms to the Standard for Real-Time Display of Thermal
{3}------------------------------------------------
and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment (AIUM/NEMA, 1992) for an on-screen display feature that provides information on potential thermal and cavitation bioeffect mechanisms. A user education program provides additional information so users may moderate the system's acoustic output in accordance with the ALARA (as low as reasonably achievable) principle.
The device's acoustic output limits are:
All Applications:
| ISPTAd | 720 mW / cm2 | (Maximum) |
|-----------------------|-----------------|-----------|
| TIS/TIB/TIC | 0.1 - 4.0 | (Range) |
| Mechanical Index (MI) | 1.9 | (Maximum) |
| ISPPAd | 0 - 700 W / cm2 | (Range) |
The limits are same as predicate Track 3 devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 26 1999
Sonosite, Inc. c/o TUV Product Service, Inc. Carol Stamp 1775 Old Highway 8 NW Suite 104 New Brighton, MN 55112
Re: K990806 C1 Ultrasound System Regulatory Class: II/21 CFR 892.1550, 21 CFR 892.1570 and 21 CFR 892.1560 Product Code: 90 IYN, 90 ITX and 90 IYO Dated: March 4, 1999 Received: March 11, 1999
Dear Mr. Stamp:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivations for use stated in the enclosure and we have had
marketed predicate devices marketed in interests communications for use sta marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enclosure) to legally Medical Device Amendments, or to devices that bevelop reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing provisions of the Actuale requirements for annual registerion,
and authersi
This determination of substantial equivalence applies to the following transducers intended for use with the Cl Ultrasound System, as described in your premarket notifications
Transducer Model Number
C7 - 4MHz IVT L7 - 4MHz Linear Array C4 - 2MHz Curves Array
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirent detemination assumes
for Medical Devices: General (GMB) Practice requirement, as set forth in the Quality System Reg for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic (QS
FDA will verify such assumptions . Railyre to compluguid . (2) CRP Part 800) pe FDA will verify such as not (2) ceation (21 certain and through periodic QS inspections, the addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal the Federal Register. Please not to your premarks notification does not affect any obligation your device in may have under sections 53 l and 542 of the Act for devices under the Electronic of the may obligation
may have under sections 53 l and 542 of the Act for devices under the E provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Please be advised that the determination above Is based on the fact that no medical devices have been demonstrated to be safe and effective for in vitro fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for in vitro fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the investigational device exemptions (IDE) regulations, 21 CFR, Part 812. Therefore, your product labeling must be consistent with FDA's position on this use.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
{6}------------------------------------------------
Page 3 – Carol Stamp
If you have any questions regarding the content of this letter, please contact Robert Philllips, Ph.D.at (301) 594-1212.
Sincerely yours,
David A. Szyrman
for
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{7}------------------------------------------------
#### System: C1 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | | Mode of Operation | | | | | |
|---------------------------|-------------------------------------------------------|---|-------------------|-----|-----|------------------|---------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | | | | B+M | Note |
| | Abdominal | P | P | | | | B+M | Note |
| | Intra-operative<br>(Abdominal organs and<br>vascular) | P | P | | | | B+M | Note |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | P | P | | | | B+M | Note |
| | Small Organ (breast,<br>thyroid, testicles.) | P | P | | | | B+M | Note |
| | Neonatal Cephalic | P | P | | | | B+M | Note |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | N | N | | | | B+M | Note |
| | Trans-vaginal | P | P | | | | B+M | Note |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | | | | B+M | Note |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | P | P | | | | B+M | Note |
| | Cardiac Pediatric | N | N | | | | B+M | Note |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | | | | B+M | Note |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Note 1: Other includes Amplitude Doppler, combined B and Amplitude Doppler, 3-D Imaging, Harmonic Imaging, and imaging for guidance of biopsy previously cleared through 510(k) K981505.
Note 2: Other includes Amplitude Doppler, combined B and Amplitude Doppler, 3-D Imaging, Harmonic Imaging, and imaging for guidance of biopsy

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K990806
Prescription Use (Per 21 CFR 801.109)
Indications for Use
{8}------------------------------------------------
## System: C1 Ultrasound System Transducer: C7-4 MHz IVT
:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | Mode of Operation | | | | | | | |
|---------------------------|----------------------------------------------|-------------------|---|-----|-----|------------------|---------------------|------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | P | P | | | | B+M | Note | |
| | Abdominal | | | | | | | | |
| | Intra-operative | | | | | | | | |
| | (Abdominal organs and | | | | | | | | |
| | vascular) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging | Laparoscopic | | | | | | | | |
| & Other | Pediatric | | | | | | | | |
| | Small Organ (breast,<br>thyroid, testicles.) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | N | N | | | | B+M | Note | |
| | Trans-vaginal | P | P | | | | B+M | Note | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | | |
| Vessel | Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Note 1: Other includes Amplitude Doppler, combined B and Amplitude Doppler, 3-D Imaging, Harmonic Imaging, and imaging for guidance of biopsy previously cleared through 510(k) K981505.
Note 2: Other includes Amplitude Doppler, combined B and Amplitude Doppler, 3-D Imaging, Harmonic Imaging, and imaging for guidance of biopsy.
David A. Szyman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K990806
Prescription Use (Per 21 CFR 801.109)
Indications for Use
{9}------------------------------------------------
## System: C1 Ultrasound System Transducer: L7-4 MHz Linear Array
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|-------------------------------------------------------|-------------------|---|-----|-----|------------------|---------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | P | P | | | | B+M | Note |
| | Abdominal | P | P | | | | B+M | Note |
| | Intra-operative<br>(Abdominal organs and<br>vascular) | P | P | | | | B+M | Note |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | | | | B+M | Note |
| | Small Organ (breast,<br>thyroid, testicles.) | P | P | | | | B+M | Note |
| | Neonatal Cephalic | P | P | | | | B+M | Note |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | | | | B+M | Note |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | N | N | | | | B+M | Note |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | | | | B+M | Note |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Note 1: Other includes Amplitude Doppler, combined B and Amplitude Doppler, 3-D Imaging, Harmonic Imaging, and imaging for guidance of biopsy previously cleared through 510(k) K981505.
Note 2: Other includes Amplitude Doppler, combined B and Amplitude Doppler, 3-D Imaging, Harmonic Imaging, and imaging for guidance of biopsy
David h. Sezmm
Division Sign-Off Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K770206
Prescription Use (Per 21 CFR 801.109)
Indications for Use
{10}------------------------------------------------
### System: C1 Ultrasound System Transducer: C4-2 MHz Curved Array
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|---------------------------|-------------------------------------------------------|---|---|-----|-----|------------------|---------------------|--------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Oth<br>(Spec |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | | | | B+M | Note |
| | Abdominal | P | P | | | | B+M | Note |
| | Intra-operative<br>(Abdominal organs and<br>vascular) | P | P | | | | B+M | Note |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | P | P | | | | B+M | Note |
| | Small Organ (breast,<br>thyroid, testicles.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | P | P | | | | B+M | Note |
| Cardiac | Cardiac Pediatric | N | N | | | | B+M | Note |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Note 1: Other includes Amplitude Doppler, combined B and Amplitude Doppler, 3-D Imaging, Harmonic Imaging, and imaging for guidance of biopsy previously cleared through 510(k) K981505.
Note 2: Other includes Amplitude Doppler, combined B and Amplitude Doppler, 3-D Imaging, Harmonic Imaging, and imaging for guidance of biopsy
Yoril le. Sazhron
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K990806
Prescription Use (Per 21 CFR 801.109)
Indications for Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.