CANTOR & SILVER 5X TINTED (HIOXIFILCON A) SOFT (SPHERICAL & TORIC) DAILY WEAR CONTACT WEAR CONTACT LENS, TINTED (TRANSPA
Applicant
Cantor & Silver , Ltd.
Product Code
LPL · Ophthalmic
Decision Date
Apr 21, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The CANTOR & SILVER 5X Tinted (hioxifilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia. The lenses may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program. The CANTOR & SILVER 5X Tinted (hioxifilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism up to 10.00 diopters. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.
Device Story
Soft contact lens fabricated from hioxifilcon A (copolymer of 2-HEMA and GMA, cross-linked with EGDMA); hydrophilic material becomes pliable when hydrated. Lens covers cornea and extends beyond limbus to provide refractive surface. Tinted with reactive color additives on anterior surface corresponding to iris; tinting process does not alter lens performance. Available in spherical and toric designs for daily wear. Requires hydrogen peroxide-based disinfection. Used by patients for vision correction; prescribed by eye care professionals. Benefits include visual acuity correction and cosmetic enhancement via tinting.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on material chemistry, physical properties, and manufacturing process comparisons to predicate devices.
Technological Characteristics
Material: hioxifilcon A (42% polymer, 58% water). Refractive index: 1.515 (dry), 1.404 (hydrated). Oxygen permeability: 20 x 10^-11 (cm²/sec)(ml O₂/ml x mm Hg @ 35°C). Non-ionic copolymer of 2-HEMA and GMA, cross-linked with EGDMA. Lathe-cut manufacturing. Tinted with reactive color additives. Sterilization via standard procedures.
Indications for Use
Indicated for aphakic and non-aphakic persons with non-diseased eyes for correction of myopia, hyperopia, and refractive astigmatism (up to 1.50D for spherical, up to 10.00D for toric).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
4/21/99
# SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
K990757
Applicant information:
Date Prepared:
March 12, 1999
Name: Address Cantor & Silver Limited Manor Road, Brackley Northamptonshire England NN13 6ED
Contact Person:
Phone Number: Fax:
Mr. David Cantor Managing Director/President 011 44 1280 702002 011 44 1280 703003
Official Correspondent: And US Consultant
Med-Vice Consulting, Inc. Mr. Martin Dalsing President/CCO 623 Glacier Grand Junction, CO 81503 970.243.5490 970.243.5501
Phone Number: Fax Number:
Device Information:
Regulatory Classification: Class II
86 LPL Product Code:
Trade Name:
CANTOR & SILVER 5X Tinted (hioxifilcon A) Soft (Spherical & Toric) Daily Wear Contact Lens, Tinted.
Classification Name:
Lenses, Soft Contact, Daily Wear
{1}------------------------------------------------
#### Equivalent Device:
The CANTOR & SILVER 5X Tinted (hioxifilcon A) Soft (Spherical & Toric) Daily Wear Contact Lens is substantially equivalent to the predicate device in terms of intended use and design. The predicate device is the Durasoft 2 Colors manufactured by Wesley-Jessen.
### Device Description:
The CANTOR & SILVER 5X Tinted (hioxifilcon A) Soft (Spherical & Toric) Daily Wear Contact lens are fabricated from hioxifilcon A, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution. The color enhanced soft contact lenses are tinted with "listed" color additives. The color additives are used in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect. As part of the manufacturing process, the lenses containing the color additives are thoroughly washed to remove unbound reactive color additives. The manufacturing process alters and/or changes the specifications to the clear version of a contact lens by affixing a listed color reactive additive on that portion of the anterior (front) surface of the lens that corresponds to the iris. The finted contact lenses are available in a light, medium or dark shade of the following enhance colors; Blue, Green, Aqua, Yellow, Lavender, Brown, Ultra Violet, and Amber.
The color additive (timing) effect is formed by reacting one or more of the "listed" additives with HEMA based (poly hydroxyethy) methacrylate) soft contact lens. The color additives used are not removed by lens handling and cleaning/disinfecting procedures as recommended. Except for affecting the amount of light transmitted through the lens, the coloring process does not alter the original characteristics of the pre-tinted lens.
In the bydrated state, the lens conforms to the curvature of the eye covering the comea and extending slightly beyond the limbus forming a transparent optical surface. The (hioxifilcon A) soft hydrophilic contact lens a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
{2}------------------------------------------------
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 58% water by weight.
The non-ionic lens material, hioxifilcon A is an ultra high molecular weight copolymer of 2-bydroxyethyl methacrylate (2-HEMA) and 2,3-Dibydroxypropyl Methacrylate (Glycerol Methacrylate, GMA) and cross-linked with ethylene glycol dimethacrylate (EGDMA). It consists of 42% hioxifiloon A and 58% water by weight when immersed in normal saline solution buffered with sodium bicarbonate. The physical properties of the lens are:
| Refractive Index | 1.515 (dry) 1.404 (hydrated) |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Light Transmission (tinted) | greater than 70% |
| Color Additives | reactive black 5, reactive blue 21, reactive blue 19, reactive blue 4,<br>reactive blue 163, reactive red 11, reactive red 180, reactive yellow 15,<br>reactive yellow 86, or reactive orange 78 |
| Water Content | 58 % ±2% |
| Specific Gravity | 1.308 (dry) 1.136 (hydrated) |
| Oxygen Permeability | $20 X 10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg @ 35°C), (revised Fatt method) |
#### INDICATED USE:
The CANTOR & SILVER 5X Tinted (hioxifileon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia. The lenses may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the assigmatism does not interfere with visual acuity. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.
The CANTOR & SILVER 5X Tinted (hioxifilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism up to 10.00 diopters. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.
{3}------------------------------------------------
#### Substantial Equivalence:
The device will be manufactured according to specified process controls and a Quality Management System certified to ISO 9002 and EN46002 by the National Accreditation of Certification Bodies. The device will undergo manufacturing, packaging and sterilization procedures similar to devices currently marketed and distributed by Cantor & Silver Limited in the United States. The established safety profile (pre-clinical toxicology and manufacturing/chemistry data) of the device is equivalent to the BENZ-G 5X (hioxifilcon A), 510(k) #K952620. The tinting characteristics and parameters of the lens are substantially equivalent to the Durasoft 2 Colors tinted soft contact lens, manufactured Wesley-Jessen. Being similar with respect to indications for use, materials, physical construction and safety & effectiveness to the predicate device, this meets the requirements per section 510(k) of the act regarding substantial equivalence and does not raise different questions of safety and effectiveness than the predicate devices identified above.
The following matrix illustrates that the production method, lens function, and the "listed" color additives and timing characteristics of the CANTOR & SILVER 5X Tinted (hioxifilcon A) Soft (Spherical & Toric) Daily Wear Contact Lens are substantially equivalent to the predicate device.
| | Characteristic | CANTOR & SILVER 5X<br>TINTED | DURASOFT 2 COLORS |
|-----|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.) | PRODUCTION<br>METHOD | Lathe-Cut | Lathe-Cut |
| 2.) | LENS FUNCTION | Refractive medium that focuses<br>light rays from near and distant<br>objects on the retina, while<br>compensating for refractive error,<br>including (astigmatism) | Refractive medium that focuses<br>light rays from near and distant<br>objects on the retina, while<br>compensating for refractive error,<br>including (astigmatism) |
| 3.) | FDA "listed"<br>colored additives | The reactive colored additives<br>consist of reactive black 5,<br>reactive blue 4, reactive blue 19,<br>reactive 21, reactive blue 163,<br>reactive yellow 15, reactive yellow<br>86, reactive orange 78, reactive<br>red 11 and reactive red 180 | The colored pigments consist of<br>iron oxides, chromium oxide<br>greens, titanium dioxide, [phthalo-<br>cyaninato (2-)] copper, carbazole<br>violet and phthalocyanine green. |
| a. | Uses<br>and<br>restrictions | The color additives listed above<br>may be used to color contact<br>lenses in amounts not to exceed<br>the minimum reasonably required<br>to accomplish the intended<br>coloring effect. | The color additives listed above<br>may be used to color contact<br>lenses in amounts not to exceed<br>the minimum reasonably required<br>to accomplish the intended<br>coloring effect. |
| 4.) | Color Additive<br>Characteristics | The color additives used are not<br>removed by lens handling and<br>cleansing/disinfecting procedures.<br>The optical and performance<br>characteristics are not altered by<br>the lens coloring process. | The color additives used are not<br>removed by lens handling and<br>cleansing/disinfecting procedures.<br>The optical and performance<br>characteristics are not altered by<br>the lens coloring process. |
| 5.) | Colors Offered | Blue, Green, Aqua, Yellow,<br>Lavender, Brown, Ultra Violet<br>and Amber | Sky blue, Jade green, Aquamarine<br>and Violet blue |
#### Substantial Equivalence Matrix
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human figures or profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 1 1933
Mr. Martin Dalsing Med. Vice Consulting, Inc. 623 Glacier Drive Grand Junction, CO
Re: K990757 Trade Name: Cantor & Silver 5X Tinted (hioxifilcon A) Soft Contact Lens Regulatory Class: II Product Code: 86 LPL Dated: April 7, 1999 Received: April 12, 1999
Dear Mr. Dalsing:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{5}------------------------------------------------
Page 2 - Mr. Martin Dalsing
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on. the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
#### CANTOR & SILVER 5X TINTED (hioxifilcon A) Soft (Spherical & Toric) Device Name: Daily Wear Contact Lens, Tinted.
# INDICATIONS FOR USE:
The CANTOR & SILVER SX Tinted (hioxifilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia. The leases may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.
The CANTOR & SILVER 5X Tinted (hioxifilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-disessed eyes with myopia or hyperopia and/or possesses refractive astigmatism up to 10.00 diopters. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.
(PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V (Per 21 CFR 801.109)
or
Over-The-Counter Use
(Optional Format 1-2-96)
SRRell
(Division Sign-Off) Division of Ophthalmic Devices
510(k) Number K990757
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.