Olympus HX-5/6-1 Endoscopic Clip Fixing Device has been designed for endoscopic clip placement within the gastrointestinal (GI) tract for the purpose of endoscopic marking, hemostasis in the upper GI tract for mucosal/submucosal defects <3cm, bleeding ulcers and arteries <2mm, polyps<1.5cm in diameter, and anchoring to affix jejunal feeding tubes to the wall of the small bowel. This device is not intended for the repair of GI tract.
Device Story
Endoscopic clipping device used for GI tract procedures; consists of main body, control wire, and clips. Device inserted through endoscope instrument channel. Operation: clip attached to hook via wire advancement; tensioning wire seats clip onto stainless steel coil; FEP tube sheath covers distal end. Upon reaching target, sheath retracted; pulling control slider tensions wire, pulling clip into pipe; pipe forces clip arms to close on tissue for deployment. Used by physicians in clinical settings for hemostasis, marking, and feeding tube anchoring. Benefits include controlled tissue approximation and secure fixation.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Stainless steel coil; FEP (fluorinated ethylene propylene) tube sheath. Mechanical actuation via manual control wire tensioning. Dimensions/form factor compatible with standard endoscope instrument channels. Manual operation; no energy source or software.
Indications for Use
Indicated for patients requiring endoscopic clip placement in the GI tract for marking, hemostasis of upper GI mucosal/submucosal defects <3cm, bleeding ulcers/arteries <2mm, polyps <1.5cm, or anchoring jejunal feeding tubes to the small bowel wall. Not indicated for repair of GI tract lumenal perforations.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
3/26/99
t
\$\kappa\$ 990687 pg, /gr
# 510(k) SUMMARY OLYMPUS HX-5/6-1 Endoscopic Clipping Device
## A. Submitter's Name, Address, Phone and Fax Numbers
#### 1. Manufacturer of the subject devices
| Name & Address of manufacturer: | Olympus Optical Co., Ltd.<br>22-2 Nishi-Shinjuku, 1-Chome,<br>Shinjuku-ku, Tokyo 163-8610<br>Japan |
|---------------------------------|----------------------------------------------------------------------------------------------------|
| Registration No.: | 8010047 |
| Address, Phone and Fax Numbers: | 2951 Ishikawa-Cho, |
| of R&D Department, | Hachioji-shi, Tokyo 192-8507 |
| Endoscope Division | Japan |
| | TEL 0426-42-5101 |
| | FAX 0426-46-2786 |
## B. Name of Contact Person
Name: Address, Phone and Fax Numbers: Ms. Laura Storms-Tyler Olympus America Inc. Regulatory Affairs Two Corporate Center Drive Melville, New York 11747-3157 TEL: (516)-844-5688 FAX: (516)-844-5416
# C. Device Name, Common Name, Classification Name and Predicate Devices
| Device Name: | Olympus HX-5/6-1 Endoscopic Clipping Device |
|----------------------|--------------------------------------------------------|
| | Standard Clips MD-59/850 |
| | Long Clips MD-858 |
| | Short Clips MD-859 |
| | MAJ-459 Short Clip |
| Common Name: | Endoscopic Clipping Device |
| Classification Name: | Endoscope and accessories |
| Predicate Device: | Olympus HX-5/6-1 Endoscopic Clipping Device<br>K963160 |
{1}------------------------------------------------
K 99 0687 pg 2gr
#### D. Description of the Device(s)
The HX-5/6-1 Endoscopic Clipping Device is available as a set consisting of the HX-5/6-1 Endoscopic Clipping Device main body and clips.
These clips are attached to the hook when the wire is advanced out of the distal end of the device. Applying tension to the control wire will "seat" a step on the clip onto the distal end of the stainless steel coil. The FEP tube sheath may then be advanced to cover the distal end of the coil and the attached clip. The device may then be inserted through the instrument channel of the appropriate endoscope.
When the device has been advanced to the area of interest, the tube sheath is retracted by moving the tube joint distally until an audible "click" is heard. When the control section slider is pulled proximally, the control wire is tensioned, and the clip is pulled into the clip body (pipe). Due to the shape of the clip itself, when the clip is pulled into the clip pipe, it will initially open wider. As it is pulled in even further, the clip pipe will force the clip arms to close on the target tissue and deploy.
### E. Intended Use of the Device(s)
Olympus HX-5/6-1 Endoscopic Clip Fixing Device has been designed for endoscopic clip placement within the gastrointestinal (GI) tract for the purpose of endoscopic marking, hemostasis in the upper GI tract for mucosal/submucosal defects <3cm, bleeding ulcers and arteries <2mm, polyps<1.5cm in diameter, and anchoring to affix jejunal feeding tubes to the wall of the small bowel. This device is not intended for the repair of GI tract.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 25, 2015
Olympus America, Inc. Laura Storms-Tyler Director, Regulatory Affairs Two Corporate Center Drive Melville, NY 11747-3157
Re: K990687
> Trade/Device Name: Olympus HX-5/6-1 Endoscope Clipping Device Regulation Number: 21 CFR§ 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: II Product Code: PKL Dated (Date on orig SE Itr): February 25, 1999 Received (Date on orig SE Itr): March 1, 1999
Dear Laura Storms-Tyler,
This letter corrects our substantially equivalent letter of March 26, 1999.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{3}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
15 990687 510(k) Number(if known): __
Device Name: Olympus HX-5/6-1 Endoscopic Clipping Device Indications for Use:
> Olympus HX-5/6-1 Endoscopic Clip Fixing Device has been designed for endoscopic clip placement within the gastrointestinal(GI) tract for the purpose of endoscopic marking, hemostasis in the upper GI tract for mucosal/submucosal defects <3cm, bleeding ulcers and arteries <2mm, polyps <1.5cm in diameter, ' and anchoring to affix jejunal feeding tubes to the wall of the small bowel. This device is not intended for the repair of GI tract lumenal perforations.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| Division of | ctive, Abdominal, ENT, and Radiological Devices |
|-------------|-------------------------------------------------|
|-------------|-------------------------------------------------|
| 510(k) Number | K990687 |
|---------------|---------|
|---------------|---------|
| Prescription Use | <div></div> | OR | Over-The-Counter Use |
|------------------|-------------|----|----------------------|
|------------------|-------------|----|----------------------|
(Per 21 CFR 801.109)
(Optoinal Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.