QUANTASE TOTAL GALACTOSE SCREENING ASSAY

K990644 · Quantase , Ltd. · JIA · Jun 21, 1999 · Clinical Chemistry

Device Facts

Record IDK990644
Device NameQUANTASE TOTAL GALACTOSE SCREENING ASSAY
ApplicantQuantase , Ltd.
Product CodeJIA · Clinical Chemistry
Decision DateJun 21, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1310
Device ClassClass 1
AttributesPediatric

Indications for Use

The Quantase Total Galactose Screening Assay is an enzymatic colorimetric end-point method for the determination of total D(+)galactose (D(+)galactose + galactose-1-phosphate) in dried blood spot specimens taken from newborn human infants as part of a newborn screening programme. The test is intended as a screening the total D(+)galactose concentrations in newborn blood spot specimens. Elevated results are not diagnostic per se of galactosaemia, but indicate the urgent need for further study of the newborn from which the presumptive positive specimen was received. The kit is NOT intended for use in monitoring the circulating concentrations of total D(+)galactose in galactosaemic patients nor to detect ante-natal galactosaemia.

Device Story

Quantase Total Galactose Screening Assay is an in vitro diagnostic kit for newborn screening. Input: dried blood spot specimens from newborns. Principle: enzymatic colorimetric end-point method measuring total D(+)galactose and galactose-1-phosphate. Output: colorimetric concentration measurement. Used in clinical laboratory settings by trained laboratory personnel. Results indicate need for follow-up diagnostic testing for galactosaemia; not diagnostic on its own. Benefits: facilitates early detection of galactosemia in newborns.

Clinical Evidence

No clinical data provided in the document; bench testing only.

Technological Characteristics

Enzymatic colorimetric end-point assay. Form factor: 2000 test kit. Analyte: total D(+)galactose (D(+)galactose + galactose-1-phosphate).

Indications for Use

Indicated for newborn human infants as part of a newborn screening program to measure total D(+)galactose (D(+)galactose + galactose-1-phosphate) in dried blood spot specimens. Not for monitoring galactosaemic patients or detecting ante-natal galactosaemia.

Regulatory Classification

Identification

A galactose test system is a device intended to measure galactose in blood and urine. Galactose measurements are used in the diagnosis and treatment of the hereditary disease galactosemia (a disorder of galactose metabolism) in infants.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The logo is simple and recognizable, representing the department's role in providing health and human services to the United States. JUN 21 1999 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 R. Stewart Campbell Ph.D. Technical Director Quantase Ltd. 3 Riverview Business Park Friarton Road Perth, Scotland, United Kingdom K990644 Re: > Trade Name: Quantase™ Total Galactose Screening Assay 2000 Test Kit Regulatory Class: I Reserved Product Code: JIA Dated: April 29, 1999 Received: April 29, 1999 Dear Dr. Campbell: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1}------------------------------------------------ ## Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known): K 990644 Device Name: Quantase Total Galactose Screening Assay, 2000 test Kit Indications For Use: The Quantase Total Galactose Screening Assay is an enzymatic colorimetric end-point method for the determination of total D(+)galactose (D(+)galactose + galactose-1-phosphate) in dried blood spot specimens taken from newborn human infants as part of a newborn screening programme. The test is intended as a screening the total D(+)galactose concentrations in newborn blood spot specimens. Elevated results are not diagnostic per se of galactosaemia, but indicate the urgent need for further study of the newborn from which the presumptive positive specimen was received. The kit is NOT intended for use in monitoring the circulating concentrations of total D(+)galactose in galactosaemic patients nor to detect ante-natal galactosaemia. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...