K990573 · Estech, Inc. · DXC · Mar 11, 1999 · Cardiovascular
Device Facts
Record ID
K990573
Device Name
ESTECH ARTERIAL REMOTE ACCESS PERFUSION CANNULA
Applicant
Estech, Inc.
Product Code
DXC · Cardiovascular
Decision Date
Mar 11, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ESTECH Arterial Remote Access Perfusion Cannula is intended for use in arterial perfusion of the aorta, via a femoral artery, in cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass (CPB). The device may also be used to occlude the ascending aorta, deliver cardioplegia solution and vent the aortic root.
Device Story
Sterile, disposable, 21 French polyurethane cannula; features three integrated lumens and distal inflatable polyurethane balloon. Central lumen delivers arterial blood (up to 5 L/min) via multiple proximal outlet ports; secondary lumen delivers cardioplegia or vents aortic root; third lumen controls balloon inflation. Used in cardiovascular surgery by surgeons to establish extracorporeal cardiopulmonary bypass. Device provides antegrade flow pattern, mimicking physiological conditions. Radiopaque markers and depth marks assist positioning. Simplifies procedures by combining femoral artery cannulation and aortic occlusion functions into single device, replacing multi-component systems. Benefits include reduced procedural complexity and improved flow dynamics compared to retrograde-flow predicate systems.
Clinical Evidence
Evidence includes bench testing (balloon inflation/burst, repeat inflation, 6-hour hold, flow/pressure drop, hemolysis, biocompatibility), animal studies, and a clinical investigation. Bench testing demonstrated performance within specifications. Animal studies verified handling and functional modes. Clinical investigation conducted to verify safety, efficacy, and performance.
Indicated for patients undergoing cardiovascular surgery requiring extracorporeal cardiopulmonary bypass. Used for arterial perfusion of the aorta via femoral artery, aortic occlusion, cardioplegia delivery, and aortic root venting.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
Predicate Devices
Heartport Arterial Perfusion Systems
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
# ESTECH Arterial Remote Access Perfusion Cannula
Common/Classification Name: Cardiopulmonary Bypass vascular catheter, cannula or tubing as classified under 21 CFR 870.4210
ESTECH 4115 Blackhawk Plaza Circle, Suite 100 Danville, CA 94506 925-648-2033 (Tel), 925-648-2034 (Fax)
Prepared: February 19, 1999
#### Legally Marketed Predicate Device A.
The ESTECH Arterial Remote Access Perfusion Cannula is substantially equivalent to perfusion cannulae currently marketed in the U.S. The ESTECH cannula is specifically equivalent to the Heartport Arterial Perfusion Systems have the same intended use, which is to provide arterial perfusion of the aorta, via a femoral artery, in cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass. Both systems provide occlusion of the ascending aorta with an inflatable balloon and may be used for delivery of cardioplegia through an internal lumen of the device. Additional similarities between the ESTECH Arterial Perfusion Cannula and the Heartport Cannula are the ability to monitor Aortic Root Pressure and the capability to provide Left Ventricle Venting. The ESTECH cannula is 21 French in diameter and the Heartport system is offer in 21 and 23 French diameters.
The primary difference between the systems is that the Heartport Arterial Perfusion System requires the use of the Femoral Artery Cannula and an Aortic Occlusion cannula, which must be inserted through the Femoral Artery Cannula. These functions are combined in the single device ESTECH Arterial Remote Access Perfusion Cannula. The ESTECH device delivers blood into the Aorta "Antegrade Flow" and the Heartport device delivers blood into the femoral artery "Retrograde Flow".
#### B. Device Description
The ESTECH Arterial Remote Access Perfusion Cannula is a sterile, disposable, flexible polyurethane tube with three integrated lumens and an inflatable polyurethane balloon at the distal end of the cannula. The outside diameter of the cannula is 21 French. The cannula has a central lumen for delivery of arterial blood through multiple distal outlets at a flow rate up to 5 liters per minute, a small lumen for delivery of cardioplegia or venting at the aortic root, and an additional smaller lumen for control of the distal balloon. The blood outlet ports are multiple elongated openings along the cannula proximal to the occlusion balloon. Radiopaque balloon markers and insertion depth marks aid in positioning the device. The cannulas are provided sterile in individual packages for single use.
್ತು.
{1}------------------------------------------------
### C. Indications for Use
The ESTECH Arterial Remote Access Perfusion Cannula is intended for use in arterial perfusion of the aorta, via a femoral artery, in cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass (CPB). The device may also be used to occlude the ascending aorta, deliver cardioplegia solution and vent the aortic root.
### D. Substantial Equivalence Summary
The ESTECH Arterial Remote Access Perfusion Cannula has the same intended use as the 510(k) cleared Heartport Perfusion System. Both systems provide arterial perfusion of the aorta, via a femoral artery, in cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass.
The primary difference is that the Heartport Arterial Perfusion System requires the use of a Femoral Artery Cannula and an Aortic Occlusion cannula, which must be inserted through the Femoral Artery Cannula. These functions are combined in the single device ESTECH Perfusion Cannula to simplify the procedure.
Another difference is the perfusion flow pattern of each system. The multiple holes along the length of the ESTECH cannula produce a normal antegrade flow pattern similar to physiological conditions of flow. Heartport produces retrograde flow from a single outlet port of the Heartport femoral cannula. Testing was performed to demonstrate that the ESTECH cannula does not induce a higher degree of hemolysis or pressure drop than expected during bypass procedures.
All materials used in the ESTECH cannula have a safe history of use in medical devices and are used by the manufacturer to produce similar 510(k) cleared devices.
The ESTECH Arterial Remote Access Perfusion Cannula is therefore substantially equivalent to currently marketed devices.
## E. Technological Characteristics
See Device Description, above.
#### F. Testing
The laboratory testing adhered to Good Laboratory Practices guidelines.
The occlusion balloon has been tested to demonstrate that it meets performance and safety specifications as follows:
Occlusion Balloon Volume/Pressure/Diameter: This test was performed to document the inflation characteristics of the balloon up to burst. The pressure and diameter data as a function of inflation volume were collected.
Occlusion Balloon Repeat Inflation Test: This test was performed to demonstrate that the balloon can be inflated repeatedly if necessary.
{2}------------------------------------------------
Occlusion Balloon Inflation Hold Test: This test was performed to demonstrate that the balloon can withstand at least 6 hours of inflation.
Hemolysis Testing: This test was performed to evaluate the cellular damage produced by the system when used during cardiac bypass procedures. Since all cardiac bypass procedures produce some degree of hemolysis which may be at least partially compensated for in human or animal testing, an in vitro test was used to evaluate the relative effect of the ESTECH cannula in the circuit.
Flow Testing: The pressure drop was measured for a range of arterial perfusion flow rates of 0-5 liters per minute. In additional testing, the pressure/flow characteristics of the Heartport cannula were measured. While the pressure drop was determined to be within ESTECH specifications, the outlet blood velocity of the Heartport device was determined to be unacceptable.
Biocompatability Testing: Biocompatability Testing was performed on all critical components of the Cannula to verify biocompatability.
Animal Testing: Animal studies of the ESTECH Arterial Remote Access Cannula to evaluate the handling characteristics and to verify performance of all intended functions. The ESTECH perfusion cannula was used in all functional modes. The conclusion of this study was that complete function of the ESTECH Arterial Remote Access Perfusion Cannula and ease of use were demonstrated.
Clinical Investigation: A Clinical Investigation was conducted to verify the safety. efficacy, and performance of the device.
## G. Conclusions
ESTECH has demonstrated through its comparison of characteristics with predicate device and comparison of performance testing with the predicate device that the ESTECH Arterial Remote Access Perfusion Cannula is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned within a circle that contains the text "HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR I I 1999
Mr. Arthur A. Bertolero Estech, Inc. 4115 Blackhawk Plaza Circle, Suite 100 Danville, CA 94506
Re : K990573 Estech Arterial Remote Access Perfusion Cannula Requlatory Class: II (Two) Product Code: 74 DXC, 74 DWF Dated: February 22, 1999 Received: February 23, 1999
Dear Mr. Bertolero:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls The general controls provisions of the provisions of the Act. Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in In addition, FDA may publish further regulatory action. announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obliqation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
## Page 2 - Mr. Arthur A. Bertolero
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Also, please note the regulation Compliance at (301) 594-4639. entitled, "Misbranding by reference to premarket notification" Other general information on your (21CFR 807.97). responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html."
Sincerely yours,
Thomas J. Allelson
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number (if known): _ K 9005 73 ________________________________________________________________________________________________________________________________________
Device Name: ESTECH Arterial Remote Access Perfusion Cannula
Indications for Use:
The ESTECH Arterial Remote Access Perfusion Cannula is intended for use in arterial perfusion of the aorta, via a femoral artery, in cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass (CPB). The device may also be used to occlude the ascending aorta, deliver cardioplegia solution and vent the aortic root.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use __________________
(Optional Format 1-2-96)
M. Page.
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.