IS H. PYLORI IGG ELISA TEST SYSTEM
Device Facts
| Record ID | K990462 |
|---|---|
| Device Name | IS H. PYLORI IGG ELISA TEST SYSTEM |
| Applicant | Columbia Bioscience, Inc. |
| Product Code | LYR · Microbiology |
| Decision Date | Sep 3, 1999 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.3110 |
| Device Class | Class 1 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K990462 · Sep 3, 1999 | IS H. PYLORI IGG ELISA TEST SYSTEM | Columbia Bioscience, Inc. | Retrospective clinical sera samples; Routine clinical diagnostic results (biopsy, culture, stain, CLO) | Retrospective patient sera were used to determine the sensitivity and specificity of the H. pylori IgG ELISA kit by comparing test results against the patients' established clinical status from routine diagnostic procedures. | Retrospective sera; Clinical diagnostic status; Sensitivity and specificity |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Retrospective clinical performance evaluation | Adult patients with gastrointestinal symptoms; Sample Size: 249; Number of Sites: 1 (clinical commercial laboratory) | Biopsy with culture, stain, and CLO results | Sensitivity, specificity, and overall agreement |
Indications for Use
For the qualitative determination of IgG antibodies to Helicobactor pylori antigen in human sera by indirect enzyme immunoassay. The H.pylori IgG assay may be used as an aid in the diagnosis of H. pylori infection in adult patients with gastrointestinal symptoms. The test can be performed either manually or in conjunction with the MAGO® PLUS Automated EIA Processor.
Device Story
The H. pylori IgG ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for detecting IgG antibodies to H. pylori antigen in human serum. Purified H. pylori antigen is bound to microwells; patient sera, calibrators, and controls are added and incubated. If present, anti-H. pylori IgG antibodies bind to the antigen. After washing, horseradish peroxidase-labeled anti-human IgG conjugate is added, followed by substrate. The resulting end product absorbance is measured spectrophotometrically at 450 nm (reference 600-630 nm). Absorbance is directly proportional to IgG concentration. The test is performed manually or using the MAGO® PLUS Automated EIA Processor in clinical laboratory settings. Results assist healthcare providers in diagnosing H. pylori infection in symptomatic adults.
Clinical Evidence
Performance evaluated using 249 retrospective patient sera characterized by biopsy (culture, stain, CLO). Sensitivity was 94.9% (112/118; 95% CI: 89.3-98.1%) and specificity was 90.4% (113/125; 95% CI: 83.8-94.9%). Overall agreement with biopsy was 92.6%. Relative agreement with another commercial ELISA was 99.6% (238/239). Precision studies (intra- and inter-assay) were conducted at three sites; manual and MAGO® PLUS automated procedures showed high correlation (r=0.993).
Technological Characteristics
Enzyme-linked immunosorbent assay (ELISA) using purified H. pylori antigen bound to microwells. Detection via horseradish peroxidase-labeled anti-human IgG conjugate and spectrophotometric measurement at 450 nm. Compatible with manual or automated (MAGO® PLUS) processing.
Indications for Use
Indicated for qualitative detection of IgG antibodies to H. pylori in human serum as an aid in diagnosing H. pylori infection in adult patients with gastrointestinal symptoms.
Regulatory Classification
Identification
Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.
Predicate Devices
- Wampole H. pylori IgG ELISA