K990452 · Ybeam Technologies, Inc. · GEX · Apr 29, 1999 · General, Plastic Surgery
Device Facts
Record ID
K990452
Device Name
ORLIGHT 2000 SURGICAL LASER SYSTEM
Applicant
Ybeam Technologies, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 29, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OrLight 2000 CO2 Surgical Laser System is intended for use in the ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery, including aesthetic surgery and resurfacing. The specific indications are as follows: - Ablation, vaporization, excision, incision, and coagulation of soft tissue in laser skin resurfacing, laser derm-abrasion, and laser burn debridement. - Laser skin resurfacing (ablation and/or vaporization) for the treatment of wrinkles, rhytids, and furrows. - Laser skin resurfacing (ablation, vaporization, excision, incision, and coagulation) for the reduction, removal and/or treatment of actinic keratosis, solar/actinic elastosis, actinic cheilitis, lentinges, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas, tattoos, telangiectasia, squamous cell carcinoma, epidermal nevi, xanthelasma palpebrarum, syringoma, and verrucae vulgares (warts). - Laser incision and/or excision of soft tissue for the performance of laser blepharoplasty.
Device Story
OrLight 2000 is a carbon dioxide (CO2) surgical laser system utilizing a radio frequency (RF) amplifier. Device delivers laser energy to target tissue via an articulated arm or hollow wave guide. Used by physicians in clinical settings for dermatology and plastic surgery procedures, including skin resurfacing, derm-abrasion, and burn debridement. Laser energy enables precise ablation, vaporization, excision, incision, and coagulation of soft tissue. Clinical benefit includes reduction or removal of skin lesions, scars, and aesthetic imperfections. Physician operates system to achieve desired tissue effect based on visual assessment of surgical site.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance was validated through design testing and comparison of soft tissue ablation characteristics against predicate CO2 laser systems.
Technological Characteristics
Carbon dioxide (CO2) laser system; radio frequency (RF) amplifier; delivery via articulated arm or hollow wave guide. Class II device. No specific material standards or software algorithm details provided.
Indications for Use
Indicated for patients requiring soft tissue ablation, vaporization, excision, incision, or coagulation in dermatology and plastic surgery. Includes treatment of wrinkles, scars, keloids, hemangiomas, tattoos, various skin lesions (e.g., actinic keratosis, squamous cell carcinoma), and performance of blepharoplasty.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Coherent Ultrapulse Carbon Dioxide Lasers
Sharplan Carbon Dioxide SilkLaser
Submission Summary (Full Text)
{0}------------------------------------------------
Laser System
Premarket Notification February 2, 1999
K990452
## SECTION 7
# 5104 SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
- Submitter's name, address, telephone number, contact person, and date 1. summary prepared:
- a. Y-Beam Technologies, Inc. 216 Technology Drive, Suite L Irvine, CA 92618
- Joseph Neev, Ph.D. b. Contact Person: President
- January 20, 1999 Date Summary Prepared: C.
#### 2. Name of device, including trade name and classification name:
| a. Trade/Proprietary Name: | OrLight 2000 Surgical Laser System |
|----------------------------|----------------------------------------|
| b. Classification Name: | Surgical laser system (21 CFR 878.4810 |
### Identification of the predicate device or legally marketed device or devices to 3. which substantial equivalence is being claimed:
The OrLight 2000 Surgical Laser is substantially equivalent to the following predicate devices : Coherent Ultrapulse Carbon Dioxide Lasers and Sharplan Carbon Dioxide SilkLaser.
- A description of the device that is the subject of the 510(k), including 4. explanation of how the device functions, basic scientific concepts, significant physical and performance characteristics (design, material, physical properties):
{1}------------------------------------------------
The OrLight 2000 Surgical Laser system is a carbon dioxide laser that utilizes a radio frequency amplifier. Laser treatment is delivered to the target tissue through an articulated arm or a hollow wave guide.
#### 5. Statement of intended use:
The OrLight 2000 Surgical Laser System is intended for use in the ablation, vaporization, excision, incision and coagulation of soft tissue in dermatology and plastic surgery, including aesthetic surgery and resurfacing.
## Statement of how the technological characteristics of the device compare to 6. those of the predicate or legally marketed device:
The OrLight 2000 Surgical Laser System has similar design features, functional features, and laser parameters as the predicate, legally marketed devices (Sharplan and Coherent carbon dioxide lasers).
#### 7. Brief summary of nonclinical tests and results:
The OrLight 2000 Surgical Laser System has been designed and tested to applicable safety standards. In addition, the OrLight 2000 Surgical Laser System was found to perform equivalently to the predicate devices with respect to soft tissue ablation.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often used as a symbol of medicine. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 9 1000
Y-Beam Technologies, Inc. c/o Dr. Judy F. Gordon, D.V.M. Judy Gordon Consulting 18732 Saginaw Drive Irvine, California 92612
> K990452 Trade Name: OrLight 2000 Surgical Laser System Regulatory Class: II Product Code: GEX Dated: February 2, 1999 Received: February 12, 1999
Dear Dr. Gordon:
Re:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 – Dr. Judy F. Gordon
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
[signature]
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page *_ of _*
510(k) Number (if known): K990452
Device Name: Or Light 2000 Surgical Laser System
Indications For Use:
The OrLight 2000 CO2 Surgical Laser System is intended for use in the ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery, including aesthetic surgery and resurfacing. The specific indications are as follows:
- Ablation, vaporization, excision, incision, and coagulation of soft tissue in ● laser skin resurfacing, laser derm-abrasion, and laser burn debridement.
- Laser skin resurfacing (ablation and/or vaporization) for the treatment of . wrinkles, rhytids, and furrows.
- Laser skin resurfacing (ablation, vaporization, excision, incision, and ● coagulation) for the reduction, removal and/or treatment of actinic keratosis, solar/actinic elastosis, actinic cheilitis, lentinges, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas, tattoos, telangiectasia, squamous cell carcinoma, epidermal nevi, xanthelasma palpebrarum, syringoma, and verrucae vulgares (warts).
- Laser incision and/or excision of soft tissue for the performance of laser . blepharoplasty.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of General Restorative Devices | |
| 510(k) Number | K990452 |
| Prescription Use<br>(Per 21 CFR 801/109) | X | OR | Over-The-Counter Use |
|------------------------------------------|---|----|----------------------|
|------------------------------------------|---|----|----------------------|
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.