The SideKick Infusion Kit is intended to provide continuous infusion of a local anesthetic directly into an intraoperative (soft tissue / body cavity) site for general surgery for postoperative pain management. Additional routes of administration include percutaneous, subcutaneous, intramuscular and epidural infusion. The SideKick pump is re-usable. The disposable SideKick administration set is single patient use only. No testing has been conducted to determine the efficacy of the SideKick for the delivery of blood, blood products, lipids or fat emulsions. The SideKick is not intended for the delivery of blood, blood products, lipids or fat emulsions. The SideKick is suitable for use as an ambulatory device and is intended for use in the hospital, home environment or alternative care sites.
Device Story
SideKick Infusion Kit is a mechanical, spring-powered ambulatory infusion pump system. It delivers local anesthetics at a continuous, fixed flow rate (1 or 2 ml/hr) to intraoperative sites or via percutaneous, subcutaneous, intramuscular, or epidural routes for postoperative pain management. The system consists of a reusable mechanical pump and a disposable administration set, which includes a catheter, needle, syringe, and accessories. Used in hospitals, homes, or alternative care sites, the device is operated by clinicians or patients. The pump provides a fixed flow rate without electronic components, eliminating fluid runaway risks associated with electronic pumps. The healthcare provider sets up the device to provide consistent pain relief, benefiting the patient through ambulatory postoperative pain management.
Clinical Evidence
Bench testing only. Performance data confirmed flow rate accuracy within ±15% at a 95% confidence interval under standard operating conditions (31°C, 0.9% NaCl test solution). Biological testing of fluid path components conforms to ISO 10993 Part 1.
Technological Characteristics
Mechanical spring-powered infusion pump. Disposable administration set; reusable pump. Flow rates: 1 or 2 ml/hr. Accuracy: ±15%. Sterilization: ETO gas. Fluid path materials conform to ISO 10993. No external power source required.
Indications for Use
Indicated for continuous infusion of local anesthetics for postoperative pain management in patients undergoing general surgery. Routes of administration include intraoperative (soft tissue/body cavity), percutaneous, subcutaneous, intramuscular, and epidural. Not indicated for delivery of blood, blood products, lipids, or fat emulsions.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date 4/23/99 at the top left. Below the date is the I-FLOW logo, which is a stylized oval shape with the words "I-FLOW" inside. Underneath the logo, the words "I-FLOW CORPORATION" are printed in a simple, sans-serif font. The logo and text are centered.
20202 Windrow Drive Lake Forest. CA 92630 (800) 448-3569 (949) 206-2700 Fax (949) 206-2600
1990425
# SUMMARY OF SAFETY AND EFFECTIVENE
February 9, 1999
Trade Name: SideKick Infusion Kit
Common Name: Infusion Pump Kit
Classification Name: Pump, Infusion
All questions and/or comments concerning this document should be made to:
Robert J. Bard, Esq., R.A.C. Vice President of Regulatory and Legal Affairs
I-Flow Corporation 20202 Windrow Drive Lake Forest, CA 92630
Telephone: 949.206.2700 Fax: 949.206.2600
{1}------------------------------------------------
#### 1.0 GENERAL INFORMATION
#### 1.1 Purpose of Submission
- This submission is intended to notify the Federal Food and Drug 1.1.1 Administration that I-Flow Corporation intends to market a new kit, the SideKick Infusion Kit.
- Trade Name: SideKick Infusion Kit 1.1.2
- 1.1.3 Common Name: Infusion Pump Kit
- 1.1.4 Classification Name: Pump, Infusion
- Classification Panel: General Hospital and Personal Use Device 1.1.5
#### 1.2 Statement of Equivalence
- 1.2.1 The SideKick Infusion Kit includes components that are legally marketed (either pre-amendment devices or devices that have been granted permission to market via premarket notification regulation).
- 1.2.2 The SideKick Kit is substantially equivalent to the I-Flow Paragon Infusion Kit (K984146), the I-Flow Paragon Infusion System (K923875), the I-Flow PainBuster Infusion Kit (K980558, K982946), the Sgarlato Pain Control Infusion Pump (PCIP) (K896422), the I-Flow Homepump C-Series (K944692) and the McKinley Outbound Disposable Syringe Infuser (K982256).
#### 2.0 PHYSICAL SPECIFICATIONS AND DESCRIPTIONS
#### 2.1 Description of the SideKick Infusion Kit
- 2.1.1 The SideKick Infusion Kit is identical to the I-Flow Paragon Infusion Kit with the exception of the SideKick pump and administration set replacing the Paragon pump and administration set.
- 2.1.2 The kit is comprised of a SideKick pump and administration set and various kit components such as catheter, needle, syringe, Y adapter, dressing, tape, gauze and carry case.
- 2.1.2.1 The Paragon Infusion Kit contains all the above components except for a Paragon pump and administration set instead of the SideKick pump and administration set.
- 2.1.3 The SideKick administration set is intended to attach to the kit catheter at the distal end of the set to provide continuous infusion of a local anesthetic directly into the intraoperative site for general surgery for postoperative pain management.
- 2.1.4 The SideKick administration set is a disposable device intended for single patient use. The SideKick pump is reusable.
- 2.1.5 The SideKick is suitable for use as an ambulatory device and is intended for use in the hospital, home environment or alternative care sites.
#### 2.2 Product Configuration
- 2.2.1 The SideKick Infusion Kit models are available in 100 ml fill volumes with 1 or 2 ml/hr flow rates.
- 2.2.2 Each model consists of a SideKick administration set with the following optional components/accessories:
{2}------------------------------------------------
- SideKick pump, catheter, needle, syringe, dressing, carry case, 2.2.2.1 antiseptic skin swabs, tape, gauze and Y adapter.
#### 2.3 Components and Materials
All fluid path components of the SideKick administration set are identical to the fluid path components of the Paragon administration set.
#### 2.4 Power Requirements
- The SideKick pump is a mechanical pump that utilizes spring energy for 2.4.1 power. No additional external power source is required.
#### 3.0 OPERATIONAL SPECIFICATIONS AND DESCRIPTIONS
#### 3.1 Standard Operating Conditions:
| Residual Volume: | < 5 ml |
|------------------------|---------------------------------|
| Operating Temperature: | 31°C skin temperature (90°F) |
| Test Solution: | 0.9% NaCl |
| Operating Pressure: | 9 to 1 psi pressure source |
| Head Height: | 0" |
| Accuracy: | ±15% at 95% confidence interval |
- 3.2 Flow Rate Performance Data: Testing occurred at standard operating conditions. All models produced an average flow rate within the ±15% accuracy claim.
#### 3.3 Safety / Alarm Functions
- 3.3.1 The SideKick pump and administration set provide a continuous fixed flow and as such is not subject to fluid runaway conditions similar to that of some electronic pumps.
## 4.0 BIOLOGICAL SPECIFICATIONS
- 4.1 Biological testing is in conformance with ISO 10993 Part 1 for all fluid path components of the SideKick administration set.
## 5.0 CHEMICAL AND DRUG SPECIFICATIONS
- 5.1 Compatibility
- 5.1.1 There are no specific drugs referenced in the labeling for the SideKick Infusion Kit.
- 5.1.2 The SideKick Infusion Kit is intended for use with general local anesthetics and epidural medications.
## 6.0 INTENDED USE
- 6.1 The SideKick Infusion Kit is intended to provide continuous infusion of a local anesthetic directly into an intraoperative (soft tissue / body cavity) site for general surgery for postoperative pain management.
- 6.2 Additional routes of administration include percutaneous, subcutaneous, intramuscular and epidural infusion.
- 6.3 The SideKick pump is re-usable. The disposable SideKick administration set is single patient use only.
{3}------------------------------------------------
- 6.4 No testing has been conducted to determine the efficacy of the SideKick for the delivery of blood, blood products, lipids or fat emulsions. The SideKick is not intended for the delivery of blood, blood products, lipids or fat emulsions.
- The SideKick is suitable for use as an ambulatory device and is intended for use in the 6.5 hospital, home environment or alternative care sites.
#### 7.0 PACKAGING
- Packaging is suitable for either radiation or ETO sterilization. 7.1
#### 8.0 STERILIZATION INFORMATION
- 8.1 The method of sterilization is ETO gas.
#### 9.0 COMPARISON TO LEGALLY MARKETED DEVICES
- 9.1 The SideKick Infusion Kit has similar routes of administration and components as the following predicate devices: the Paragon Infusion Kit, the Paragon Infusion System, PainBuster Infusion Kit, Sgarlato Pain Control Infusion Pump (PCIP), Homepump C-Series and McKinley Outbound Syringe Infuser.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and head. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circle's perimeter. The text is in all capital letters and is arranged to follow the curve of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 3 1999
Robert J. Bard, Esq., R.A.C. Vice President Regulatory and Legal Affairs I-Flow Corporation 20202 Window Drive Lake Forest, California 92630
Re: K990425 SideKick Infusion Kit Trade Name: Regulatory Class: II Product Code: FRN Dated: February 9, 1999 February 11, 1999 Received:
Dear Mr. Bard:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Mr. Bard
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
sincerely yours
L. G. Whatrust
Timot Wy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K990425
Image /page/6/Picture/1 description: The image shows the logo for I-Flow Corporation. The logo consists of the words "I-Flow" inside of a rounded rectangle. The word "Corporation" is printed below the logo.
20202 Windrow Drive Lake Forest, CA 92630 (800) 448-3569 (949) 206-2700 Fax (949) 206-2600
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: __ SideKick Infusion Kit
# Indications for Use:
1. The SideKick Infusion Kit is intended to provide continuous infusion of a local anesthetic directly into an intraoperative (soft tissue / body cavity) site for general surgery for postoperative pain management. Additional routes of administration include percutaneous, subcutaneous, intramuscular and epidural infusion.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUED ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices
510(k) Number K990425
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use __
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.