← Product Code [MGQ](/productcode/MGQ) · K990368

# DUODERM HYDROACTIVE GEL WOUND DRESSING (K990368)

_Convatec, A Bristol-Myers Squibb Co. · MGQ · Mar 26, 1999 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K990368

## Device Facts

- **Applicant:** Convatec, A Bristol-Myers Squibb Co.
- **Product Code:** [MGQ](/productcode/MGQ.md)
- **Decision Date:** Mar 26, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

For Over-The Counter Use, DuoDERM Hydroactive Gel Wound Dressing may be used for abrasions, lacerations, minor cuts, minor scalds and burns, and skin tears. Under the supervision of a healthcare professional, DuoDERM Hydroactive Gel may be used for the hydration and management of partial and full-thickness wounds, such as pressure sores, leg ulcers, diabetic ulcers and skin conditions resulting from radiation dermatitis.

## Device Story

DuoDERM Hydroactive Gel is a topical wound dressing. It provides a moist wound healing environment and absorbs wound exudate. Used in clinical settings under professional supervision for chronic/complex wounds or by patients for minor injuries. It supports wound management by maintaining hydration. It is not indicated for third-degree burns, as a permanent dressing, or as a cure for wounds.

## Clinical Evidence

No clinical data provided. Safety and effectiveness supported by bench testing performed in accordance with Good Laboratory Practice Regulations and reference to previous premarket notification K931618.

## Technological Characteristics

Hydroactive gel wound dressing. Provides moist environment and exudate absorption. Non-implantable, topical application. No electronic components or software.

## Predicate Devices

- Carrasyn® Hydrogel Wound Dressing

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

3/26/99

K990368

510(k) Premarket Notification DuoDERM® Hydroactive® Gel Wound Dressing

## SUMMARY OF SAFETY AND EFFECTIVENESS

| Applicant:                            | ConvaTec, A Division of E.R. Squibb and Sons, Inc.<br>100 Headquarters Park Drive, Skillman, NJ 08558 |
|---------------------------------------|-------------------------------------------------------------------------------------------------------|
| Contact:                              | Adrienne McNally, Director, Regulatory Affairs<br>(908) 904-2630                                      |
| Device:                               | DuoDERM® Hydroactive Gel Wound Dressing                                                               |
| • Substantially<br>Equivalent Device: | Carrasyn® Hydrogel Wound Dressing                                                                     |

For Over-The Counter Use, DuoDERM Hydroactive Gel Wound Dressing may be used for abrasions, lacerations, minor cuts, minor scalds and burns, and skin tears. Under the supervision of a healthcare professional, DuoDERM Hydroactive Gel may be used for the hydration and management of partial and full-thickness wounds, such as pressure sores, leg ulcers, diabetic ulcers and skin conditions resulting from radiation dermatitis.

· DuoDERM Hydroactive Gel Wound Dressing should not be used on individuals who are sensitive to or who have had an allergic reaction to the gel or its components.

DuoDERM Hydroactive Gel Wound Dressing is substantially equivalent to Carrasyn Hydrogel Wound Dressing. The products are equivalent in intended use and function. Both products provide a moist wound healing environment. DuoDERM Hydroactive Gel Wound Dressing is also capable of absorbing exudate.

Data/information supporting the safety of DuoDERM Hydroactive Gel Wound Dressing was presented in Premarket Notification K931618. All testing was performed in accordance with Good Laboratory Practice Regulations.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 6 1999

Ms. Adrienne McNally Director, Regulatory Affairs ConvaTec A Division of E.R. Squibb & Sons, Inc. 100 Headquarters Park Drive Skillman, New Jersey 08558

Re: K990368

> Trade Name: DuoDERM® Hydroactive® Gel Wound Dressing Regulatory Class: Unclassifed Product Code: MGQ Dated: February 3, 1999 Received: February 8, 1999

Dear Ms. McNally:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market your device subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act) and the following limitations:

- 1. This device may not be labeled for use on third degree burns.
- This device may not be labeled as having any accelerating effect on the rate of 2. wound healing or epithelization.
- 3. This device may not be labeled as a long-term, permanent, or no-change dressing, or as an artificial (synthetic) skin.
- This device may not be labeled as a treatment or a cure for any type of wound. 4.

The labeling claims listed above would be considered a major modification in the intended use of the device and would require a premarket notification submission (21 CFR 807.81).

The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.

{2}------------------------------------------------

Page 2 - Ms. Adrienne McNally

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations (CFR), Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practices (GMP) for Medical Devices: General GMP regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or 301-443-6597 or at its internet address http://www.fda.gov/cdrh/dsmamain.html.

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K990368

510(k) Premarket Notification DuoDERM® Hydroactive® Gel Wound Dressing

## INDICATIONS FOR USE STATEMENT

510(k) Number (if known): Not Known

Device Name: DuoDERM Hydroactive Gel Wound Dressing

Indications for Use:

For Over-The Counter Use, DuoDERM Hydroactive Gel Wound Dressing may be used for abrasions, I of Over The Counter Coo, a scalds and burns, and skin tears. Under the supervision of a healthcare laccrations, inmor edus, minor boards mic bund, be used for the hydration and management provider, DaoDDANI 11, arounds such as pressure sores, leg ulcers, diabetic ulcers and skin conditions resulting from radiation dermatitis.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over the Counter Use
(Optimal Format 1-2-96)

neral Restorative Devices 990368

---

**Source:** [https://fda-staging.innolitics.com/device/K990368](https://fda-staging.innolitics.com/device/K990368)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
