← Product Code [GKZ](/productcode/GKZ) · K990352

# COULTER AC T DIFF 2 ANALYZER (K990352)

_Coulter Corp. · GKZ · Apr 1, 1999 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K990352

## Device Facts

- **Applicant:** Coulter Corp.
- **Product Code:** [GKZ](/productcode/GKZ.md)
- **Decision Date:** Apr 1, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The COULTER® AC T diff 2m analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories.

## Device Story

The COULTER® Ac·T diff 2™ Analyzer is an automated hematology analyzer for clinical laboratory use. It processes whole blood samples via manual or closed-vial (rotary cap piercing) modes, or pre-diluted samples. The device utilizes the Coulter principle (electrical impedance) for cell enumeration and sizing, and a single-beam photometer for hemoglobin measurement. It automatically dilutes and mixes samples; uses a lytic reagent to lyse red blood cells for WBC and hemoglobin analysis; and employs sweep flow to prevent cell re-entry into the sensing zone. The system calculates a three-population leukocyte differential (lymphocytes, mononuclear cells, granulocytes) based on cell size histograms. Output includes CBC parameters (WBC, RBC, Hgb, Hct, MCV, MCH, MCHC, Plt, RDW, MPV, and differential counts) displayed to the clinician. Results are flagged for values outside normal ranges. The device aids clinicians in hematological assessment and patient monitoring.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and performance comparisons to predicate devices.

## Technological Characteristics

Automated hematology analyzer using electrical impedance (Coulter principle) for cell counting/sizing and single-beam photometry for hemoglobin. Features include automatic diluting/mixing, closed-vial sampling, and sweep flow technology. Reagent system: isotonic diluent, lytic reagent, and cleaner. Software provides automated calibration, control calculations, and data storage. Interface uses universal icons.

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- COULTER® Ac-T diff™ Analyzer ([K973634](/device/K973634.md))
- COULTER® COUNTER® S-PLUS IV and LYSE S PLUS D ([K823355](/device/K823355.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

4/1/99

K990352

## Summary of Safety & Effectiveness COULTER® AC.T diff 2rw Analyzer

#### 1.0 Submitted By

Tom Enalish Manager, Premarket Product Regulatory Compliance Coulter Corporation 11800 SW 147 Avenue MC 31-B06 Miami, FL 33196 Telephone: (305) 380-4331 FAX: (305) 380-3618

#### 2.0 Date Submitted

January 27, 1999

#### 3.0 Device Name(s)

### 3.1 Proprietary Names

COULTER® Ac·T diff 2™ Analyzer

### 3.2 Classification Names

An automated differential cell counter is a device used to identify and classify one or more of the formed elements of blood. (CFR 864.5220)

#### 4.0 Predicate Device(s)

COULTER® Ac-T diff™ Analyzer (K973634) COULTER® COUNTER® S-PLUS IV and LYSE S PLUS D (K823355)

#### 5.0 Description

The COULTER® Ac·T diff 2™ Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. COULTER® Ac.T diff 2™ Analyzer has the same technological characteristics and is substantially equivalent to the COULTER® COUNTER® S-PLUS IV and LYSE S PLUS D (also called COULTER COUNTER® Model S PLUS IV with Three Population Differential and Model S PLUS IV Diff), which was cleared by 510(k) K823355 on Dec. 28, 1982 and is a modified version of the COULTER® Ac-T diff ™ Analyzer which were cleared by 510(k) K973634 on October 29, 1997.

As with the predicate devices, the COULTER® Ac-T diff 2™ Analyzer utilizes the Coulter principle for the enumeration and sizing of blood cells, in combination with an automatic diluting and mixing device for sample processing and a single beam photometer for the measurement of hemoglobin. The same reagent system, composed of an isotonic diluent, lyitc reagent to lyse red blood cells for WBC and hemoglobin measurement and instrument cleaner, is used on Model S PLUS IV Diff and Ac-T diff Analyzers.

{1}------------------------------------------------

All the devices determine the following CBC parameters:

| WBC  | (Leukocyte count)                           |
|------|---------------------------------------------|
| LY # | (Lymphocyte number)                         |
| LY%  | (Lymphocyte percent)                        |
| MO # | (Mononuclear number)                        |
| MO % | (Mononuclear percent)                       |
| GR # | (Granulocyte number)                        |
| GR % | (Granulocyte percent)                       |
| RBC  | (Erythrocyte count)                         |
| Hgb  | (Hemoglobin)                                |
| Hct  | (Hematocrit)                                |
| MCV  | (Mean Corpuscular Volume)                   |
| MCH  | (Mean Corpuscular Hemoglobin)               |
| MCHC | (Mean Corpuscular Hemoglobin Concentration) |
| Plt  | (Platelet or thrombocyte count)             |
| RDW  | (Red Cell Distribution Width)               |
| MPV  | Mean Platelet (thrombocyte) Volume          |
| Pct* | Plateletcrit                                |
| PDW* | Platelet Distribution Width                 |

*These parameters are not for diagnostic use but can be used as internal instrument checks on the platelet parameters.

Like the Ac T diff Analyzer, the Ac T diff 2 can analyze samples in either of two modes: whole blood and pre-dilute. Additionally, the Ac-T diff 2 can sample closed vial specimens by virtue of a rotary cap piercing functionality. The MODEL S PLUS IV Diff does not have a pre-dilute mode or a closed vial sampling mode.

#### 6.0 Intended Use:

The COULTER® Ac-T diff 2 analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter intended For In Vitro Diagnostic Use in clinical laboratories.

#### 7.0 Comparison to Predicate(s):

The following tables outline the basic similarities and differences between the Ac-T diff 2 analyzer and the predicate devices.

{2}------------------------------------------------

| COULTER® AC-T diff 2™ Analyzer                     | COULTER®AC-T diff™ Analyzer                        | COULTER® S-PLUS IV with Diff®                      |
|----------------------------------------------------|----------------------------------------------------|----------------------------------------------------|
| Utilizes the Coulter principle for enumeration and | Utilizes the Coulter principle for enumeration and | Utilizes the Coulter principle for enumeration and |
| sizing of blood cells.                             | sizing of blood cells.                             | sizing of blood cells.                             |
| Utilizes an automatic sampling, diluting and       | Utilizes an automatic sampling, diluting and       | Utilizes an automatic sampling, diluting and       |
| mixing device for sample processing.               | mixing device for sample processing.               | mixing device for sample processing.               |
| Reagent system includes isotonic diluent, lytic    | Reagent system includes isotonic diluent, lytic    | Reagent system includes isotonic diluent, lytic    |
| reagent and cleaning agent.                        | reagent and cleaning agent.                        | reagent and cleaning agent.                        |
| Simultaneous analysis of RBC and WBC in            | Simultaneous analysis of RBC and WBC in            | Simultaneous analysis of RBC and WBC in            |
| separate analysis vessels.                         | separate analysis vessels.                         | separate analysis vessels.                         |
| Utilizes sweep flow to prevent cells from          | Utilizes sweep flow to prevent cells from          | Utilizes sweep flow to prevent cells from          |
| re-entering the sensing zone and being counted as  | re-entering the sensing zone and being counted as  | re-entering the sensing zone and being counted as  |
| platelets.                                         | platelets.                                         | platelets.                                         |
| Ability to set normal patient ranges against which | Ability to set normal patient ranges against which | Ability to set normal patient ranges against which |
| sample results are compared. Sample results are    | sample results are compared. Sample results are    | sample results are compared. Sample results are    |
| flagged with "H" if the result is above the normal | flagged with "H" if the result is above the normal | flagged with "H" if the result is above the normal |
| range and "L" if below the normal range.           | range and "L" if below the normal range.           | range and "L" if below the normal range.           |
| Uses a single aperture each for WBC and RBC        | Uses a single aperture each for WBC and RBC        | Uses three apertures each for WBC and RBC          |
| counting and sizing.                               | counting and sizing.                               | counting and sizing.                               |
| Parameters: WBC, RBC, Hgb, Hct, MCV, MCH,          | Parameters: WBC, RBC, Hgb, Hct, MCV, MCH,          | Parameters: WBC, RBC, Hgb, Hct, MCV, MCH,          |
| MCHC, Plt, LY % & LY #, MO% & MO #, GR %           | MCHC, Plt, LY % & LY #, MO% & MO #, GR %           | MCHC, Plt, LY % & LY #, MO% & MO #, GR %           |
| & GR #, RDW, MPV. Pct and PDW are also             | & GR #, RDW, MPV. Pct and PDW are also             | & GR #, RDW, MPV. Pct and PDW are also             |
| measured but not intended for diagnostic use.      | measured but not intended for diagnostic use.      | measured but not intended for diagnostic use.      |
| Instrument calculates a three-population leukocyte | Instrument calculates a three-population leukocyte | Instrument calculates a three-population leukocyte |
| count (LY %, MO%, GR %) from the WBC               | count (LY %, MO%, GR %) from the WBC               | count (LY %, MO%, GR %) from the WBC               |
| histogram based on cell size. Cells between 35 fL  | histogram based on cell size. Cells between 35 fL  | histogram based on cell size. Cells between 35 fL  |
| and 90 fL are classified as lymphocytes. The       | and 90 fL are classified as lymphocytes. The       | and 90 fL are classified as lymphocytes. The       |
| absolute number for each population is then        | absolute number for each population is then        | absolute number for each population is then        |
| compared                                           | compared                                           | compared                                           |
| Ability to store data.                             | Ability to store data.                             | Ability to store data.                             |
| Automated Calibration and Control calculations,    | Automated Calibration and Control calculations.    | Automated Calibration and Control calculations.    |
| Automated, self-cleaning probe.                    | Automated, self-cleaning probe.                    | See DIFFERENCES table                              |
| Uses Universal Icons for operator interface.       | Uses Universal Icons for operator interface.       | See DIFFERENCES table.                             |

{3}------------------------------------------------

Public Health Service

Image /page/3/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is an abstract image of an eagle.

APR 1 1999 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Tom English Manager, Premarket Product Regulatory Compliance Coulter Corporation 11800 SW 147 Avenue MC 31-B06 Miami. Florida 33196

Re: K990352 Trade Name: COULTER® Ac-T diff 2™ Analyzer Regulatory Class: II Product Code: GKZ Dated: January 27, 1999 Received: February 5, 1999

Dear Mr. English:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# INDICATIONS FOR USE STATEMENT

K990352
TO BE ASSIGNED

510(k) Number (if known):

Device Name: COULTER® AC.T diff 2ru Analyzer

Indications For Use:

The COULTER® AC T diff 2m analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories.

## 21 CFR 864.5220 Automated differential cell counter

An automated differential cell counter is a device used to identify and classify one or more of the formed elements of blood.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

(Division Sign-Off)
Division of Clinical Laboratory Devices K990352
510(k) Number

**Prescription Use**
(Per 21 CFR 801.109)

OR

Over-The-Counter Use _

(Optional Form 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K990352](https://fda-staging.innolitics.com/device/K990352)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
