K990328 · Morning Star, Dong Bang Acupuncture U.S.A., Inc. · MQX · May 4, 1999 · General Hospital
Device Facts
Record ID
K990328
Device Name
DBC, DONG BANG ACUPUNCTURE NEEDLES
Applicant
Morning Star, Dong Bang Acupuncture U.S.A., Inc.
Product Code
MQX · General Hospital
Decision Date
May 4, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5580
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Acupuncture needles are defined as devices "intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the States." Single use only acupuncture needles have been used for "the general practice of acupuncture" in the United States. The proposed DBC, Dong Bang Acupuncture Needles have the same intended use as the DBC Acupuncture Needles which are currently being marketed through interstate distribution (K963300), because the two devices are manufactured by a same company.
Device Story
DBC Dong Bang Acupuncture Needles are sterile, single-use, stainless steel needles; used by qualified acupuncture practitioners to pierce skin for general acupuncture practice. Device features various handle types (pipe or spring); packaged in bulk or single sealed blisters. Sterilized via gamma-ray or EOG. Device is identical to previously cleared predicate; no clinical data or complex algorithms involved.
Clinical Evidence
No clinical data. Bench testing only; device relies on historical safety record since 1988 and equivalence to a legally marketed predicate.
Technological Characteristics
Stainless steel acupuncture needles; single-use; sterile (gamma-ray or EOG); various handle types (pipe or spring).
Indications for Use
Indicated for use by qualified practitioners of acupuncture as determined by the States to pierce the skin for the general practice of acupuncture.
Regulatory Classification
Identification
An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.
Special Controls
*Classification.* Class II (special controls). The device, when it is an acupuncture point locator or a single use acupuncture needle, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9. Acupuncture needles must comply with the following special controls:(1) Labeling for single use only and conformance to the requirements for prescription devices set out in 21 CFR 801.109,
(2) Device material biocompatibility, and
(3) Device sterility.
{0}------------------------------------------------
K990328
# PRE-MARKET NOTIFICATION 510(K) SUMMARY {As Requested by 21 CFR 807. 929 (c)}
Submitter: Ae-Hoe Kwon (President of Morning Star/Staff of Dong Bang Medical Co., LTD) Morning Star, Dong Bang Acupuncture U.S.A., Inc. 1429 Lyndon St. S. Pasadena, CA 91030-3381 Tel: 626) 403-5959, Fax: 626) 403-0128, E-mail: DBCacup@aol.com
Issued Date: January 28, 1999
Trade name: DBC, Dong Bang Acupuncture Needles, 510(k) Number: K990328 Classification: II Common name: Acupuncture Needles Product code: MQX Classification name: Needle, Acupuncture, Single Use The Legally Marketed Device: DBC Acupuncture Needles, 510(k) K963300
## Description of Device:
The acupuncture needles manufactured by Dong Bang Medical Co., LTD in Korea have been imported and sold through interstate commerce in the USA since 1988 under the FDA labeling restrictions of "Caution: Federal law restricts this device to sale by or on the order of qualified practitioners as determined by the State." The subject of this 510(k) application, "DBC, Dong Bang Acupuncture Needles," is a y-ray or EOG sterile, non-pyrogen, stainless, and single use only acupuncture needle and is identical to the DBC brand needles, 510(k) K963300. The DBC, Dong Bang acupuncture needles have various type (pipe or spring) needle handles, and are packaged by bulk or single sealed blister.
#### Intended Use:
Acupuncture needles are defined as devices "intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the States." Single use only acupuncture needles have been used for "the general practice of acupuncture" in the United States. The proposed DBC, Dong Bang Acupuncture Needles have the same intended use as the DBC Acupuncture Needles which are currently being marketed through interstate distribution (K963300), because the two devices are manufactured by a same company.
### Safety, Effectiveness, and Fundamental Scientific Technology:
Since 1988, no accidents or device failure claims have been reported as a result of using the acupuncture needles supplied by Dong Bang Medical Co., LTD in the U.S.A. Sterile, stainless, single use only acupuncture needles offer greater safety. The proposed DBC, Dong Bang acupuncture needles meet the general specifications and criterion for acupuncture needles and are effective for the practice of acupuncture. The differences in trade and distributor names in labeling do not alter safety, effectiveness, or device's fundamental scientific technology.
#### Substantial Equivalence:
In conclusion, based on the information provided with this 510(k) application, the DBC, Dong Bang brand acupuncture needle meets the criterion for 510 (k) acceptance. The subject of this application is the same safe and effective DBC acupuncture needle which has been legally marketed in commercial distribution. For this DBC, Dong Bang acupuncture needle is identical to the legally marketed DBC acupuncture needle, 510(k) K963300 submitted by an other importer (Lhasa Medical, Inc.) of Dong Bang Medical Co., LTD. The differences in trade and distributor names in labeling do not affect the device's intended use or alter the device's
fundamental scientific technology.
Ae-Hoe Kwon, contact person Date 1/24/99
1429 Lyndon St.
1. Pasadena,
CA 91030-3381
: (626) 403-5959
: 403-0128
ਤੇ
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is simple and recognizable, representing the department's mission to protect the health of all Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 1999 MAY
Mr. Ae-Hoe Kwon President Morning Star, Dong Bang Acupuncture U.S.A., Incorporated 1429 Lyndon Street South Pasadena, California 91030-3381
Re : K990328 DBC, Dong Bang Acupuncture Needles Trade Name: Regulatory Class: II Product Code: MQX Dated: March 4, 1999 Received: March 8, 1999
### Dear Mr. Ae-Hoe Kwon
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 - Mr. Ae-Hoe Kwon
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA debering of substantial equivalence of your device to a legally marketed predicate device results in a classification for your markets and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note ene Ories Or Compland, "Misbranding by reference to the regulacion encretica, "Mibbianding by Coronoveral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K990328
# 510(K) NUMBER (IF KNOWN): K990328 DEVICE NAME: DBC DONGBANG ACUPUNCTURE NEEDLES
# INDICATIONS FOR USE
DBC Dongbang Acupuncture Needles have been used for "the general practice of acupuncture without any other specific use or treatment" in the United States. These single-use-only acupuncture needles are "intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the States."
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRIH, Office of Device Evaluation (ODE)
(Division Sign-Off) Patricia Cucente
Division of Dental, Infection Control,
and General Hospital Devices
| 510(k) Number | K990328 |
|------------------------------------------|--------------------------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109) | OR Over-The-Counter-Use<br>(Optional Format 1-2) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.