The Respironics SmartRecorder is a multichannel recording system designed to record and store physiologic signals acquired from adult and infant patients during sleep. SmartRecorder is designed for use in the home or hospital. Physiological signals are acquired via transducers attached to the patient and directly connected to the recorder and/or monitoring or therapeutic device connected to the recorder. Use of the associated Synergy S (host) software facilitates retrieval of data recorded by SmartRecorder via direct PC connection; remotely via internal modem; or by transferring data directly from the removable PCMCIA memory card to a PC. Once downloaded, the recorded data can be analyzed for physiological events. Synergy S allows the clinician can display, review, edit and print analyzed data as well as configure / customize all reports to best meet individual needs.
Device Story
SmartRecorder (Model 2500) is a portable, software-controlled multichannel recording system for sleep diagnostics; records up to seven dedicated sensor-input channels and four auxiliary channels; includes integrated Masimo SET oximetry technology. Device interfaces with infant monitors and sleep therapy devices to capture ECG, respiration, pressure, and flow signals. Data stored on removable PCMCIA flash memory. Operated in home or hospital settings by healthcare professionals. Synergy S host software (Windows-based) retrieves data via direct PC connection, internal modem, or memory card transfer. Software allows clinicians to validate, edit, and report physiological events; supports remote titration of CPAP/BiPAP settings. Provides clinicians with diagnostic data to inform sleep therapy decisions.
Clinical Evidence
No clinical testing was performed. Evidence consists of bench testing, including operational, performance, environmental, and EMC testing. Hardware and software requirements were verified through system-level and module/integration-level testing using simulated data.
Technological Characteristics
Portable, software-controlled multichannel recorder; 7 sensor-input channels, 4 auxiliary channels; PCMCIA flash memory storage; integrated Masimo SET oximetry; wall-powered; Windows-based Synergy S host software for data analysis and remote modem connectivity.
Indications for Use
Indicated for adult and infant patients requiring recording and storage of physiologic signals during sleep in home or hospital settings.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
K990199
Image /page/0/Picture/1 description: The image shows the logo for Respironics. The logo consists of a stylized figure inside a triangle, with the word "RESPIRONICS" written below it in a bold, sans-serif font. There is a registered trademark symbol next to the figure and the word "RESPIRONICS".
1255 Kennestone Circle Marietta, Georgia 30066-6029 USA
# SmartRecorder™ MULTICHANNEL RECORDING SYSTEM
# 510(k) SUMMARY
(As required by 21 CFR 807.93)
| Establishment<br>Registration Number: | 1040777 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Betsy Cortelloni<br>Respironics Georgia<br>1255 Kennestone Circle<br>Marietta, GA 30066<br>Phone: 770-429-2894<br>Fax: 770-499-1418 |
| Name of Contact: | Betsy Cortelloni, Regulatory Affairs Manager |
| Device Name: | SmartRecorder Multichannel Recording System |
| Device Model Number: | 2500 |
| Classification Name: | Programmable Diagnostic Computer, 21 CFR 870.1425 |
| Device Classification: | DQK |
| Predicate Devices: | Healthdyne Technologies' Alice 4 System, K971867.<br>Masimo SET MS-1 Pulse Oximeter and Accessories,<br>K962603. Simon Multichannel Recording System, K983572. |
| Device Description: | SmartRecorderTM is a software-controlled device, which<br>utilizes the same "engine" as Healthdyne Technologies'<br>Alice®4 Sleep Diagnostic system. SmartRecorder is a scaled<br>down version of Alice 4, having fewer channels and weighing<br>less than three pounds. SmartRecorder acquires, records<br>and stores up to seven dedicated channels of sensor-input<br>physiologic data, and provides the capability to interface with<br>Respironics infant monitoring devices and sleep therapy<br>devices to record ECG and respiration signals, or pressure<br>and flow signals, respectively. Additionally, SmartRecorder<br>acquires analog data through four auxiliary channels.<br>SmartRecorder includes built-in Masimo SET® oximetry<br>technology. |
{1}------------------------------------------------
### 510(K) SUMMARY -- SMARTRECORDER™ MULTICHANNEL RECORDING SYSTEM Device Description, continued:
The acquired data is recorded and stored in non-volatile removable PCMCIA flash memory, and can be retrieved by direct PC connection: remotely via internal modem; or by transferring the data directly from the memory card to a PC.
A graphic LCD display provides feedback and instruction to the clinician to assist in setup of a patient.
The associated Host software, Synergy S™, allows the HealthCare Professional (user) to access the stored data for subsequent validation and editing of all events, and provides reporting capabilities that are fully user-configurable. Synergy S is a Windows-based application, which also facilitates remote titration of CPAP / BiPAP settings via modem.
#### Intended Use: The Respironics SmartRecorder is a multichannel recording system designed to record and store physiologic signals acquired from adult and infant patients during sleep. SmartRecorder is designed for use in the home or hospital.
Physiological signals are acquired via transducers attached to the patient and directly connected to the recorder and/or monitoring or therapeutic device connected to the recorder.
Use of the associated Synergy S (host) software facilitates retrieval of data recorded by SmartRecorder via direct PC connection: remotely via internal modem; or by transferring data directly from the removable PCMCIA memory card to a PC. Once downloaded. the recorded data can be analyzed for physiological events.
Synerqy S allows the clinician can display, review, edit and print analyzed data as well as configure / customize all reports to best meet individual needs.
### Comparison of Technological Characteristics:
SmartRecorder is essentially a line extension of the Healthdyne Technologies Alice 4 System and utilizes the same software "engine" to control device functions. SmartRecorder records seven sensor-input channels and four auxiliary channels, as compared to 20 channels on Alice 4. Signals are recorded in a similar manner.
SmartRecorder is small and lightweight, and is designed for home or hospital use. It is operated using a wall mounted power supply.
The Synerqy Software facilitates download of recorded data to a PC via internal modem or direct connection to the device.
{2}------------------------------------------------
## 510(K) SUMMARY – SmartRecorder™ MULTICHANNEL RECORDING SYSTEM Continued:
- The System Qualification Test Plan for SmartRecorder included Testing: Operational and Performance testing, Environmental testing and EMC testing. The Hardware Test Plan and Software Test Plan consisted of system level and module / integration level testing to verify all the defined hardware and software requirements, respectively. Bench testing utilized simulated data. No Clinical testing was performed.
- The cumulative test results demonstrated the functionality, safety Conclusion: and effectiveness of the SmartRecorder Multichannel Recording System and the associated Synergy Software, as well as their substantial equivalence to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 9 1999
Ms. Betsy Cortelloni Respironics, Inc. 1255 Kennestone Circle Marietta, GA 30066
K990199 Re: Smartrecorder, Model 2500 Requlatory Class: II (two) 73 MNR Product Code: January 19, 1999 Dated: Received: January 21, 1999
Dear Ms. Cortelloni:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Ms. Betsy Cortelloni
This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Cal ahan, Ph. D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Devices and Radiological Health
Page of of
| 510(k) Number (if known): | K990199 |
|---------------------------|---------|
|---------------------------|---------|
Device Name: SMARTRECORDER MULTICHANNEL RECORDING SYSTEM
Indications for Use:
The Respironics SmartRecorder is a multichannel recording system designed to record and store physiologic signals acquired from adult and infant patients during sleep. SmartRecorder is designed for use in the home or hospital.
Physiological signals are acquired via transducers attached to the patient and directly connected to the recorder and/or monitoring or therapeutic device connected to the recorder.
Use of the associated Synergy S (host) software facilitates retrieval of data recorded by SmartRecorder via direct PC connection; remotely via internal modem; or by transferring data directly from the removable PCMCIA memory card to a PC. Once downloaded, the recorded data can be analyzed for physiological events.
Synergy S allows the clinician can display, review, edit and print analyzed data as well as configure / customize all reports to best meet individual needs.
A.A. Calla.
(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurence of CDRH, Office of Device Evaulation (ODE)
PRESCRIPTION પર્ટદ
OR
OVER - THE - COUNTER USE
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.