HP VIRIDIA COMPONENT MONITORING SYSTEM, MODEL REV. K
Applicant
Hewlett-Packard GmbH
Product Code
DSI · Cardiovascular
Decision Date
Jan 29, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Hewlett-Packard Viridia Component Monitoring System is intended for use in monitoring, recording, and alarming of multiple physiological parameters. The devices are indicated for use in health care facilities by health care professionals whenever there is a need for monitoring the physiological parameters of adult, neonatal, and pediatric patients.
Device Story
The HP Viridia Component Monitoring System is a multi-parameter patient monitor used in hospital environments. It integrates software from the HP Clinical Monitoring System (M1175A, M1176A, M1177A, M100B/M1002B) and the HP Component Transport System Viridia (M1205A). The device collects physiological data—including ECG, respiration, oxygen saturation, blood pressure, and temperature—via connected sensors and transducers. It processes these signals to provide real-time monitoring, recording, and alarming. Healthcare professionals use the system to observe patient status and make clinical decisions based on the displayed physiological parameters. The system is designed to facilitate shared functionality across HP monitoring platforms.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and software integration of previously cleared predicate devices.
Technological Characteristics
Multi-parameter patient monitoring system; software-based integration of existing HP monitoring platforms. Supports ECG, respiration, SpO2, blood pressure, and temperature monitoring. Includes display, alarm, and recording capabilities. Connectivity via patient cables, transducers, and electrodes.
Indications for Use
Indicated for monitoring, recording, and alarming of physiological parameters in adult, neonatal, and pediatric patients in healthcare facilities by healthcare professionals.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
K990125
ir
JAN 2 9 1999
## 8.0 510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
1. The submitter of this premarket notification is: Egon Pfeil Regulatory Affairs Medical Products Group-Europe Hewlett-Packard GmbH Herrenberger Strasse 110-140 D-71034 Germany Tel: 011 49 (7031) 14-2243 Fax: 011 49 (7031) 14-4297 Email:egon pfeil@hp.com
This summary was prepared on December 30, 1998
- 2. The name of this device is Hewlett-Packard Viridia Component Monitoring System. The common name is Patient Monitor. Classification names are as follows:
| egulation<br>umber | Classification Name |
|--------------------|-------------------------------------------------------------------|
| 68.2375 | Monitor, Breathing (Ventilatory) Frequency |
| 68.1720 | Analyzer, Gas, Oxygen, Gaseous Phase |
| 70.2340 | Electrocardiograph |
| 70.2700 | Oximeter |
| 70.2300 | Monitor, Cardiac (including Cardiotachometer & Rate Alarm) |
| 70.2350 | Adapter, Lead Switching, Electrocardiograph |
| 70.2810 | Recorder Paper Chart 510(k) exempt |
| 70.1100 | Alarm, Blood Pressure |
| 70.1110 | Computer, Blood Pressure |
| 70.1435 | Computer, Diagnostic, Pre-Programmed, Single-Function |
| 70.2450 | Display, Cathode-Ray Tube, Medical |
| 70.1025 | Detector and Alarm, Arrythmia |
| 70.2900 | Cable, Transducer and Electrode, Patient (including<br>connector) |
| 70.1120 | Cuff, Blood-Pressure |
| 70.1130 | System, Measurement, Blood Pressure, Non-invasive |
| 80.2910 | Clinical Electronic Thermometer |
3. The new combination device is substantially equivalent to previously cleared HP devices marketed pursuant to K941811, K973437 and K981376
{1}------------------------------------------------
- 4. The modification is a software based change that combines software in the HP Clinical Monitoring System product line (M1175A, M1176A and M1177A, M100B/M1002B ECG/RESP) with the software in the HP Component Transport System Viridia, (M1205A), the devices are to be known collectively as the HP Viridia Component Monitoring System. The combination system will allow for shared future functionality capabilities in HP patient monitoring systems.
- 5. The new combination device has the same intended use as the legally marketed predicate devices. When used in the hospital environment, the HP Viridia Component Monitoring System is intended for monitoring, recording, and alarming of multiple physiological parameters in adult, pediatric and neonatal patients.
- 6. The new combination device has the same technological characteristics as the legally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three heads in profile, facing to the right. The bird is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" surrounding the top half of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 29 1999
Mr. Egon Pfeil Regulatory Affairs Medical Products Group Hewlett-Packard GmbH Herrenberger Strasse 110-140 Boeblingen, Germany 71034
Re: · · K990125 HP Viridia Component Monitoring System, Model Rev K Regulatory Class: III (three) Product Code: DSI Dated: December 30, 1998 Received: January 13, 1998
Dear Mr. Pfeil:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions.
{3}------------------------------------------------
requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, 2 Misbranding by reference to premarket notification? (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
3.1 ODE Indications Statement
## Indications for Use Statement
510 (k) Number (if known)
The Hewlett-Packard Viridia Component Monitoring Device Name System
- Indications for The Hewlett-Packard Viridia Component Monitoring System is intended for use in monitoring, Use recording, and alarming of multiple physiological parameters. The devices are indicated for use in health care facilities by health care professionals whenever there is a need for monitoring the physiological parameters of adult, neonatal, and pediatric patients.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
V. Calvillo
で
Division of Cardiovascular, Respiratory, and Neurological D 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.