Assessment of symptoms that may be related to rhythm disturbances of the heart: Patients with . palpitations; The evaluation of arrhythmia's in patients from pediatric to adult age. . Assessment of risk in patients with or without symptoms of arrhythmia. . . Assessment of efficacy of Antiarrhythmic therapy. . . Assessment of Pacemaker Function. . . Assessment of real time ST segment analysis . . Assessment of symptomatic or asymptomatic patients, to evaluate for, ischemic heart disease and arrhythmia analysis during exercise testing. . . Assessment of transferred SpO2 level date from an MDE Escort Bedside monitor to EasiView for display and alarm. . . Assessment of transferred Invasive and Non-Invasive blood pressure, both systolic and diastolic, data from an MDE Escort Bedside monitor to EasiView for display and alarm. . Assessment of transferred temperature parameter data from an MDE Escort Bedside monitor to ● EasiView for display and alarm. . Assessment of transferred respiration data from an MDE Escort Bedside monitor to Easi View for display. . Assessment is for single-hospital environment.
Device Story
Zymed EasiView Telemetry System provides real-time cardiac monitoring for up to eight patients; inputs include ECG signals from transmitters (VHF band 174-216 MHz) and multi-parameter data (SpO2, NIBP, IBP, temperature, respiration) transferred from MDE Escort bedside monitors. System performs real-time ECG waveform analysis to detect/classify abnormal beats and rhythm disturbances; features include automatic ST analysis, trend data, and full disclosure storage. Used in hospital settings by clinicians; output displayed on central station monitor with visual/audible alarms. Clinicians use data for diagnosis, treatment planning, and assessment of antiarrhythmic therapy efficacy. System benefits patients through continuous monitoring and automated detection of cardiac events.
Clinical Evidence
Bench testing only. Performance validated against industry-standard AHA, MIT, and European ST-T (EST) databases. Metrics included sensitivity (SE), positive predictivity (+P), and false positive rate (FPR) for QRS, ventricular beats, couplets, and runs. Episode sensitivity (ESE), episode positive predictivity (E+P), duration sensitivity (DSE), and durative predictivity (D+P) measured for ST analysis. System demonstrated performance equal to or better than the predicate in high heart rates (>300 bpm) and noisy environments (baseline, electrode, or muscle noise).
Technological Characteristics
IBM PC AT compatible platform; 200 MHz Pentium Pro CPU; 64 MB RAM; 1.6 GB hard disk; SVGA display. Transmitters operate on tunable VHF frequencies (174-216 MHz). Supports 4, 6, or 8 beds. Software runs on Windows NT/DOS. Oscillometric NIBP measurement. Connectivity via network/strip chart/laser printer. Sterilization not applicable (non-invasive).
Indications for Use
Indicated for pediatric to adult patients requiring cardiac rhythm monitoring, arrhythmia evaluation, pacemaker function assessment, ST segment analysis, and monitoring of SpO2, blood pressure, temperature, and respiration in a single-hospital environment.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Medical Data Electronics: Escort 100/300 Series B Patient Monitor System (K930004/A)
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 1 1 1999
7 vmed
K990093
## 1
Zymed Inc. 20 North Aviador Street Camarillo, California 93010-8348 800.235.5941 - 805.987.9611 Fax 805.987.9532
## 510(k) Summary
Submitter:
Dudley Harris, Director of Regulatory Affairs/QA Zymed Medical Instrumentation 20 North Aviador Street Camarillo, CA 93010 805/987-9532 Fax: 800/235-5941 (401) Phone: Date of Summary: 1-8-99 D. Harris - see above Contact:
Trade Name: Common Name: Classification Name: (per 21 CFR 870.1025)
Zymed Telemetry System: Model EasiView Telemetry Central Station Monitor Detection and Alarm, Arrhythmia
Legally marketed device to which S.E. is claimed. Zymed Telemetry System: Model EasiView - 510(k) K 980186 Medical Data Electronics: Escort 100/300 Series B Patient Monitor System - 510(k) K930004/A
Description: A Zymed Telemetry monitoring system consists of a series of interface devices to include ECG transmitters, multi-parameter transmitters, a central Telemetry monitoring computerized unit with a strip chart recorder, Easi 5 lead and laser printer. The Zymed central monitor supports up to eight patients for real time cardiac monitoring. The system displays each patient's ECG continuously on the screen while performing real time ECG waveform analysis for all eight patients. This analysis permits immediate detection and classification of abnormal beats, cardiac rhythm disturbances and variations. The system will also display each patient's Oxygen Saturation, Respiration, Non-Invasive Blood Pressure, Invasive Blood Pressure and Temperature.
Each ECG transmitter's frequency can be programmed to operate at any frequency within the entire VHF band. For US domestic sites, the transmitters will comply with FCC band allocations (174-216 Mhz). In addition to ECG data, the transmitters also detect and transmit cardiac pacemaker information. Other information including transmitter status and individual lead impedance is also transmitted to the Zymed system for overall system safety and efficacy.
The Zymed system presents the user with a number of clinical tools such as visual and audible alarms for the diagnosis of patients with various heart conditions. The system also provides tools to review a patient's
{1}------------------------------------------------
cardiac performance. On-line review mechanisms as well as detailed analysis screens have been designed into the system to facilitate and to enhance the patient's diagnosis and treatment. Features such as individual ECG printouts, multi-channel automatic ST analysis, trend data analysis, and Full Disclosure data further enhance the system's qualities as a valuable and practical clinical tool.
The Zymed system presents the user with a number of clinical tools such as visual and audible alarms for multi-parameter monitoring of the patient. The Zymed system provides a means to monitor multiple patients from a central monitoring station for ECG, Oxygen Saturation, Non-Invasive Blood Pressure, Invasive Blood Pressure, Temperature, and Respiration.
The system has the following options available:
Choice of 4, 6, or 8 bed central monitor Full disclosure screen and printout (full resolution programmable from 0 to 168 hours) Choice of 6 lead sets, based on transmitter capability Full arrhythmia analysis to include multi-channel automatic ST Analysis Laser Printer, print server options Strip Chart Networking
Intended use:
- Assessment of symptoms that may be related to rhythm disturbances of the heart: Patients with . palpitations; The evaluation of arrhythmia's in patients from pediatric to adult age.
- . Assessment of risk in patients with or without symptoms of arrhythmia.
- . Assessment of efficacy of Antiarrhythmic therapy.
- . Assessment of Pacemaker Function.
- . Assessment of real time ST segment analysis
- . Assessment of symptomatic or asymptomatic patients, to evaluate for, ischemic heart disease and arrhythmia analysis during exercise testing.
- . Assessment of transferred SpO2 level date from an MDE Escort Bedside monitor to EasiView for display and alarm.
- . Assessment of transferred Invasive and Non-Invasive blood pressure, both systolic and diastolic, data from an MDE Escort Bedside monitor to EasiView for display and alarm.
- Assessment of transferred temperature parameter data from an MDE Escort Bedside monitor to ● EasiView for display and alarm.
- Assessment of transferred respiration data from an MDE Escort Bedside monitor to Easi View for display.
- Assessment is for single-hospital environment.
A review of the technological characteristics compared to the predicate devices are:
{2}------------------------------------------------
| Platform | EasiView (New)<br>telemetry system | EasiView (Old)<br>telemetry system |
|-------------------------------|------------------------------------|------------------------------------------------------|
| Type | IBM PC AT Compatible | Same |
| CPU | 200 Mhz Pentium Pro | Same |
| RAM | 64 M Bytes | Same |
| Hard Disk | 1.6 G Bytes | Same |
| Display | SVGA | Same |
| Transmitters: | | |
| 3 channel | yes | yes |
| Tunable frequencies | 174-216 Mhz | Same |
| Software: | | |
| Number of patients | 4, 6, or 8 | Same |
| Operating System | Windows NT<br>DOS Compatible | Same |
| Data Storage | 24 hrs of EKG/Channel | Same |
| Number of leads for Analysis: | 3 | Same |
| Automatic ST Analysis | yes | Same |
| Manual ST Analysis | no | no |
| Platform | EasiView (New)<br>telemetry system | MDE Escort Monitor System (Old)<br>monitoring system |
| Temperature: | | |
| Range | 20-50 degrees C | Same |
| Alarms | 0-150 F (0-65 C) | 0.1-44.0 C |
| Non-Invasive BP | | |
| Measurement Method | Oscillometric | Same |
| Alarms | | |
| Systolic | 0-300 mmHg | 25-240 mmHg |
| Mean | NA | 15-200 mmHg |
| Diastolic | 0-200 mmHg | 10-180 mmHg |
| Invasive BP | | |
| Number of Channels | 1 or 2 | Same |
| Alarms | | |
| Systolic | 0-300 mmHg | 0-240 mmHg |
| Mean | NA | 0-200 mmHg |
| Diastolic | 0-200 mmHg | 0-180 mmHg |
| SpO2 | | |
| Range | 20- 100% | Same |
| Alarms | 0-100% by 1%<br>0-100% by 1 % | 70-100% high limit<br>50-99% low limit |
| Respiration | | |
| Range | 4-200 | Same |
Performance was measured against industry accepted AHA (AHA), MIT (MIT) and European ST-T (EST) databases. Results were typical for the real time monitoring environment for the Easi View as targeted. Separate sensitivities (SE), positive predictivity (+P), and false positive rate (FPR) were examined for each database and measured for QRS, Ventricular, Couplets, Short runs and Long runs. Separate Episode
{3}------------------------------------------------
Sensitivities (ESE), Episode Positive Predictivity (E+P), Duration Sensitivity (DSE) and Durative Predictivity (D+P) were examined for the European ST-T(EST) database and measured for ST analysis. High heart rates to include pediatric patients were demonstrated to be within recommended guidelines in excess of 300 bpm, and performance in the presence of noise indicates the new system is equal to or better than the old system when looking at baseline, electrode or muscle as the cause of noise.
In summary, performance data between the two systems shows nearly identical data, and therefore, supports a claim of Substantial Equivalence.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 1 1999
Mr. Dudley Harris Director of RA/QA Zymed, Inc. 20 North Aviador Street Camarillo, CA 93010-8348
Re: K990093 Easiview Telemetry Central Station Monitor Regulatory Class: III (three) Product Code: DSI Dated: April 20, 1999 Received: April 21, 1999
Dear Mr. Harris:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food Drug and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this
{5}------------------------------------------------
response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, 2Misbranding by reference to premarket notification² (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J.Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## 510(k) Number: K990093
Device Name: EasiView Telemetry Monitoring System
Indications for Use:
- . Assessment of symptoms that may be related to rhythm disturbances of the heart; Patients with palpitations: The evaluations of arrhythmia's in patients from pediatric to adult age.
- Assessment of risk in patients with or without symptoms of arrhythmia. .
- Assessment of efficacy of Antiarrhythmia therapy. .
- Assessment of Pacemaker Function.
- . Assessment of real time ST segment analysis.
- Assessment of symptomatic or asymptomatic patients, to evaluate for, ischemic heart disease . and arrhythmias analysis during exercise testing.
- Assessment of transferred SpO2 level data from an MDE Escort Bedside monitor to ● EasiView for display and alarm.
- Assessment of transferred invasive and Non-invasive blood pressure, both systolic and . diastolic, data from an MDE Escort Bedside monitor to EasiView for display and alarm.
- Assessment of transferred temperature parameter data from an MDE Escort Bedside monitor . to EasiView for display and alarm.
- Assessment of transferred respiration data from an MDE Escort Bedside monitor to EasiView . for display.
- Assessment is indicated for single-hospital environment.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
F.K. Jones for the
Prescription Use
(CFR21 CFR 801.109)
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.