# MILLENNIUM MLC -120 (K990085)

_Varian Assoc., Inc. · IYE · Feb 8, 1999 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K990085

## Device Facts

- **Applicant:** Varian Assoc., Inc.
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** Feb 8, 1999
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The Varian Millennium Multileaf 120 Collimator (MLC) is an x-ray collimator designed to be mounted on a C or EX Series Varian Clinac® radiation therapy linear accelerator and is intended to shape the x-ray field perimeter. Field shape can either be static (fixed) or dynamic. Dynamic field shapes are controlled as a function of either Clinac® dose fraction or gantry angle. The intended use to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.

## Device Story

Millennium MLC-120 is an x-ray collimator accessory for Varian Clinac® C or EX Series linear accelerators. It mounts beneath standard field-defining jaws to shape radiation beams. Operates in static mode (replacing custom shadow blocks) or dynamic mode (leaf positions indexed to dose fraction or gantry angle). Used by radiation oncologists in clinical settings to deliver precise radiation to target volumes while sparing healthy tissue/organs. Enables 3D dose distribution through dynamic beam shaping during treatment.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

X-ray collimator accessory for Varian Clinac C or EX Series linear accelerators. Features 120-leaf design for field shaping. Operates via mechanical leaf positioning indexed to gantry angle or dose fraction. Connectivity is integrated with the host linear accelerator system.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Varian Multileaf Collimator with Dynamic Arc Therapy ([K973889](/device/K973889.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

FEB 8 1559 KG90085

### varian@ oncology systems

# Premarket Notification [510K] Summary as required by 21 CFR 807.92

Date Summary was prepared; January 8, 1999

Submitter's Name: Varian Oncology Systems 3045 Hanover Street Palo Alto, CA 94304

#### Contact Person:

Linda S. Nash Regulatory Compliance & Radiation Safety Manager Phone (650) 424-6990 (650) 424-4830 FAX E-Mail linda.nash@os.varian.com

Device Name: Millennium MLC - 120

Classification Name: Medical Linear Accelerator Accessory, 90 IYE (per 21 CFR section 892.5050)

Predicate Device: Varian Multileaf Collimator with Dynamic Arc Therapy, K973889

#### Product Description:

The Varian Millennium Multileaf Collimator (MLC) is an x-ray collimator designed to be mounted on a C or EX Series Varian Clinac® radiation therapy linear accelerator beneath the standard field defining the collimator jaw.

Image /page/0/Picture/13 description: The image shows a logo for "Partners for Life". The words "Partners" and "For" are stacked on top of each other in a bold, sans-serif font. The word "Life" is written in a cursive script and extends to the right of the other words. The logo is in black and white.

3 printed on recycled paper

{1}------------------------------------------------

### Intended Use:

The Varian Millennium Multileaf 120 Collimator (MLC) is an x-ray collimator designed to be mounted on a C or EX Series Varian Clinac® radiation therapy linear accelerator and is intended to shape the x-ray field perimeter. Field shape can either be static (fixed) or dynamic. Dynamic field shapes are controlled as a function of either Clinac® dose fraction or gantry angle. The intended use to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.

### Technological Characteristics:

See the attached "Specification Comparison Chart" (Tab F).

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

FEB 8 1999

Linda Nash Regulatory Compliance And Radiation Safety Manager Varian Associates, Inc. 3045 Hanover Street Palo Alto, CA 94304-1129

Re:

K990085 Millenium MLC-120 (Multileaf Collimator) Dated: January 8, 1999 Received: January 11, 1999 Regulatory class: II 21 CFR 892.5050/Procode: 90 IYE

Dear Mr. Nash:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have We neve reviewed your oodlor of the indications for use stated in the enclosure) to legally decemined the devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the manced production as noo manova lices that have been reclassified in accordance with the provisions of the Mcaral Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may ir your additional controls. Existing major regulations affecting your device can be found in the Code of De Subject to Julianal Controlly - Extentially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Mourden of the arreral voyalerily such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal rogatory dollor. In adallion, 1971 hay paur premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

~~~~

Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Varian Oncology Systems. The word "varian" is in bold, black letters, with a circled "VA" symbol to the right of the word. Below the word "varian" is the phrase "oncology systems" in a smaller, non-bold font.

# Indications For Use

510(k) Number (if known): _K990085

Device Name: Millennium MLC - 120

Indications for Use: The Varian Millennium 120 Multileaf Collimator (MLC) is provided to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation. In a static mode, the MLC performs the same function as the customized shadow blocks. In a dynamic mode, a series of MLC leaf positions can be indexed to either Clinac® dose fraction or gantry angle to create a changing beam shape while the radiation beam is on to create a three dimensional dose distribution.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH/Office of Device Evaluation (ODE)              |                                |
|--------------------------------------------------------------------|--------------------------------|
| (Division Sign-Off)                                                |                                |
| Division of Reproductive, Abdominal, ENT, and Radiological Devices |                                |
| 510(k) Number                                                      | K990085                        |
| Prescription Use (per 21 CFR 801.109)                              | OR Over-The-Counter Use ______ |

Varian Associates, Inc. 3045 Hanover Street Palo Alto, California 94304-1129 U.S.A. 650.493.4000 FAX: 650.424.4830

25 printed on recycled pa

---

**Source:** [https://fda-staging.innolitics.com/device/K990085](https://fda-staging.innolitics.com/device/K990085)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).