K990068 · General Electric Medical Systems Information Techn · MHX · Jul 14, 1999 · Cardiovascular
Device Facts
Record ID
K990068
Device Name
SOLAR 9500 INFORMATION MONITOR
Applicant
General Electric Medical Systems Information Techn
Product Code
MHX · Cardiovascular
Decision Date
Jul 14, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Solar 9500 Information Monitoring System is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients, in high acuity areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care. Physiologic data includes electrocardiogram, invasive blood pressure, noninvasive blood pressure, pulse, temperature, cardiac output, respiration, pulse oximetry, carbon dioxide, oxygen, anesthetic gas concentrations and mixed venous oxygen saturation, as summarized in the Solar 9500 Operator's Manual. The monitoring parameters are contained in Appendix B under Technical Specifications. The Solar 9500 Information Monitoring System is also intended to provide physiologic data over the Unity network to clinical information systems and allow the user to access hospital INTRAnet data via a Web Browser at the point-of-care.
Device Story
Solar 9500 is a patient monitoring system for high-acuity clinical environments. It receives physiological data from Tram-net network modules and multiparameter acquisition modules. Components include a processing unit, color display, Tram modules, and Tram-rac housing. The system displays, trends, stores, and prints patient data. It utilizes a Unity Ethernet LAN for communication with clinical information systems and cardiology products. An integrated web browser allows clinicians to access hospital intranet data (patient history, lab results, cath reports) at the point-of-care. The device supports clinical decision-making by providing real-time physiological monitoring and centralized access to patient information. It is operated by healthcare practitioners.
Clinical Evidence
Bench testing only. Quality assurance measures included requirements specification review, code inspections, software and hardware testing, safety testing, environmental testing, and final validation.
Technological Characteristics
Patient monitoring system with processing unit, color display, and modular data acquisition (Tram modules). Connectivity via Unity Ethernet LAN and hospital Enterprise Network. Includes embedded web browser for intranet access. Operates on networked architecture. Software-based data processing and display.
Indications for Use
Indicated for monitoring physiologic parameter data (ECG, blood pressure, pulse, temperature, cardiac output, respiration, pulse oximetry, CO2, O2, anesthetic gases, mixed venous O2 saturation) in adult, pediatric, and neonatal patients in high-acuity clinical settings (OR, PARR, critical care, ICU).
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Solar 9000 Anesthesia Information Monitor (K954852)
Marquette SL Series Transport Remote Acquisition Module (K921669)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a sequence of alphanumeric characters. The sequence starts with the letter 'K', followed by the numbers '9900', and then the Greek letter 'varphi', and ends with the number '8'. The characters are handwritten and have a slightly irregular appearance.
# 510(k) Summary of Safety and Effectiveness
Date:
January 7, 1999
GE Marquette Medical Systems 8200 West Tower Avenue Milwaukee, WI 53223 USA
Contact Person:
Submitter:
Karen Webb Requiatory Affairs Specialist GE Marquette Medical Systems Phone: (414) 362-3329 Fax: (414) 355-3790
Trade Name: Device:
Solar 9500 Information Monitor
## Common/Usual Name:
Patient monitor
Classification Names:
21 CFR 868.1400 Analyzer, Gas, Carbon Dioxide, Gaseous-Phase
21 CFR 868.1500 Analyzer, Gas, Enflurane, Gaseous-Phase (Anesthetic Conc.) 21 CFR 868.1620 Analyzer, Gas, Halothane, Gaseous-Phase (Anesthetic Conc.)
21 CFR 868.1690 Analyzer, Gas, Nitrogen, Gaseous-Phase (Anesthetic Conc.)
21 CFR 868.1700 Analyzer, Gas, Nitrous Oxide, Gaseous-Phase, (Anesthetic Conc.)
21 CFR 868.1720 Analyzer, Gas, Oxygen, Gaseous-Phase
21 CFR 868.2375 Breathing Frequency Monitor
21 CFR 870.1025 Detector and Alarm, Arrhythmia
21 CFR 870.1100 Monitor, Blood Pressure, Indwelling
21 CFR 870.1130 Noninvasive Blood Pressure Measurement System
21 CFR 870.1100 Blood Pressure Alarm
21 CFR 870.1425 Programmable Diagnostic Computer
21 CFR 870.2340 Electrocardiograph
21 CFR 870.1435 Monitor, Cardiac Output, Thermal (Balloon Type Catheter)
21 CFR 880.2910 Monitor, Temperature (with probe)
21 CFR 870.2300 Monitor, Cardiac (Incl. cardiotachometer & rate alarm)
- 21 CFR 870.2700 Oximeter, Pulse
Predicate Devices:
K954852 Solar 9000 Anesthesia Information Monitor
K921669 Marquette SL Series Transport Remote Acquisition Module
{1}------------------------------------------------
#### Device Description:
The Solar 9500 is a patient monitoring system that is designed to display patient physiological data that is received from the GE Marquette Medical Systems' Tram-net network and individual and multiparameter data acquisition modules.
The Solar 9500 Information Monitoring System is comprised of four basic components: the processing unit, color display, Tram modules(s), and Tram-rac housing. Optional components include a remote display.
The Solar 9500 utilizes the GE Marquette's Unity Ethernet LAN allowing communication with monitoring, clinical information and cardiology products. An additional Ethernet connection is provided for connection to the hospital Enterprise Network. The Enterprise network connection allows the user access to the hospital INTRAnet, through an embedded Web Browser on the Solar 9500. This web browsing capabliity enables the user to log on to the hospital INTRAnet directly from the monitor allowing access to information such as patient history, up-to-the-minute lab results and cath reports. Data can also be accessed from non-Marquette platforms via the Enterprise network.
#### Intended Use:
The Solar 9500 Information Monitoring System is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients, in high acuity areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive Physiologic data includes electrocardiogram, invasive blood care, care, pressure, noninvasive blood pressure, pulse, temperature, cardiac output, respiration, pulse oximetry, carbon dioxide, oxygen, anesthetic gas concentrations and mixed venous oxygen saturation, as summarized in the Solar 9500 Operator's Manual. The monitoring parameters are contained in Appendix B under Technical Specifications.
The Solar 9500 Information Monitoring System is also intended to provide physiologic data over the Unity network to clinical information systems and allow the user to access hospital INTRAnet data via a Web Browser at the point-of-care.
This information can be displayed, trended, stored, and printed.
#### Technology:
The Solar 9500 employs the same functional technology as the predicate devices.
{2}------------------------------------------------
## Test Summary:
The Solar 9500 complies with the voluntary standards as detailed in The following quality assurance Section 9 of this submission. measures were applied to the development of the Solar 9500:
- Requirements specification review �
- Code inspections .
- Software and hardware testing .
- Safety testing .
- Environmental testing .
- Final validation .
### Conclusion:
The results of these measurements demonstrated that the Solar 9500 ls as safe, as effective, and performs as well as the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird, with three curved lines representing its wings or feathers.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 4 1999
Ms. Karen Webb GE Marquette Medical Systems, Inc. 8200 West Tower Avenue Milwaukee, WI 53223
K990068 Re: Solar 9500 Information Monitor Regulatory Class: III (three) Product Code: 74 MHX Dated: January 7, 1999 Received: January 8, 1999
Dear Ms. Webb:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Ms. Karen Webb
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if Known): Unknown; 510(k) filed on January 7, 1999
Solar® 9500 Information Monitor Device Name:
Indications For Use:
The Solar@ 9500 Information Monitoring System is intended for use under the direct supervision of a licensed healthcare practitioner. The Intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients, in high acuity areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal Physiologic data includes electrocardiogram, invasive blood pressure, intensive care. noninvasive blood pressure, pulse, temperature, cardiac output, respiration, pulse oximetry, carbon dioxide, oxygen, anesthetic gas concentrations and mixed venous oxygen saturation, as summarized in the Solar® 9500 Operator's Manual. The monitoring parameters are contained in Appendix B under Technical Specifications.
The Solar® 9500 Information Monitoring System is also intended to provide physiologic data over the Unity network to clinical information systems and allow the user to access hospital INTRAnet data via a Web Browser at the point-of-care.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Daula Tiller
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.