K990025 · Careside, Inc. · JFJ · Feb 8, 1999 · Clinical Chemistry
Device Facts
Record ID
K990025
Device Name
CARESIDE AMYLASE
Applicant
Careside, Inc.
Product Code
JFJ · Clinical Chemistry
Decision Date
Feb 8, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1070
Device Class
Class 2
Indications for Use
The CARESIDE™ Amylase cartridge is intended for in vitro diagnostic use in conjunction with the CARESIDE Analyzer™ to quantitatively measure amylase activity in whole blood, serum or plasma.
Device Story
CARESIDE™ Amylase is a single-use disposable cartridge used with the CARESIDE Analyzer™ to measure amylase activity in whole blood, serum, or plasma. The user loads the specimen into the cartridge, which contains a multi-layer reagent film. The analyzer scans the barcode, heats the cartridge to 37°C, and spins it to distribute the sample. Amylase in the sample reacts with a synthetic substrate (PNP-Gm) to produce a yellow dye. The analyzer uses reflectance photometry at 425 nm to measure color intensity, which correlates to amylase activity. The system calculates results using lot-specific standard curves. Used in professional laboratory settings to aid in diagnosing pancreatic, salivary gland, and kidney diseases. Benefits include automated processing, eliminating manual specimen dilution, pipetting, and reagent pre-warming.
Clinical Evidence
Bench testing only. Performance evaluated via method comparison against Trace Scientific Amylase DST (r=0.97) and Vitros AMYL DT (r=0.95). Precision reported as total CV of 13% at 75 U/L. Detection limit is 15 U/L with a reportable range of 15-600 U/L. Interference testing showed no significant impact from ascorbic acid (20 mg/dL), hemoglobin (100 mg/dL), or triglycerides (1500 mg/dL).
Technological Characteristics
Dry film reagent cartridge; sensing via reflectance photometry at 425 nm; reaction temperature 37°C. Reagents: p-nitrophenyl-α-D-maltopentaoside and α-glucosidase. Connectivity: Barcode-based lot-specific calibration. System performs automated sample metering and distribution via centrifugal force. No manual dilution or pre-warming required.
Indications for Use
Indicated for use in the diagnosis and treatment of patients with disease of the pancreas, salivary gland, and kidney. For professional laboratory use; not for point-of-care or physician office use.
Regulatory Classification
Identification
An amylase test system is a device intended to measure the activity of the enzyme amylase in serum and urine. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).
Predicate Devices
Vitros AMYL Slides for Johnson and Johnson's Vitros DT 60 II system (K912844/A)
Submission Summary (Full Text)
{0}------------------------------------------------
K990035
CARESIDE, Inc. Page 10
## 510(K) SUMMARY: CARESIDE™ AMYLASE SAFETY V. AND EFFECTIVENESS
#### I. Applicant Information
- Applicant Name A.
- Applicant/Manufacturer Address B.
- ட் Telephone Number
- Contact Person D.
- FAX Number E.
- F. e-Mail Address
- G. Date 510(k) Summary prepared
#### II. Device Information
- Device Name (Trade) A.
- B. Device Name (Classification)
- C. Device Classification
## CARESIDE, Inc.
6100 Bristol Parkway Culver City, CA 90230 310-338-6767 Kenneth B. Asarch, Pharm.D., Ph.D. 310-338-6789 AsarchK@CARESIDE.com December 30, 1998
### CARESIDE™ Amylase
Amylase test system Clinical chemistry panel Amylase test system Regulation Number: 21 CFR 862.1070 Regulatory Class 2 Classification Number: 75JFJ None applicable
D. Special controls and performance standards
#### III. Substantial Equivalence Claim
- A. General equivalency claim
The ability to monitor analyte-specific biochemical reactions in dry film and other formats is widely recognized and has gained widespread acceptance for use in chemistry assays.
Amylase in vitro diagnostic products, in both dry film and other formats, are already on Amylase products include those that use amyloclastic methods the U.S. market. (measure the disappearance of substrate and include the iodine-starch method), saccharogenic methods (measure the production of sugars such as maltose and glucose), and chromogenic methods (measure a colored product).
#### B. Specific equivalency claim
This CARESIDE™ Amylase test is substantially equivalent in intended use and clinical performance to the currently marketed Vitros slides for the quantitative measurement of amylase on the Vitros DT 60 II system with the DTSC II module. Both are based on the principle of dry film and are read by reflectance photometry; however, the Vitros method is based upon the conversion of dyed starch to dyed saccharides while the CARESIDE method is based upon the conversion of a synthetic substrate to a colored product.
Name of Predicate Device: Johnson and Johnson's (formerly Eastman Kodak, Inc.) Vitros AMYL Slides for Johnson and Johnson's Vitros DT 60 II system with the DTSC II module (formerly Eastman Kodak's DT 60 II). Predicate Device 510K number: K912844/A Product Code: 75CIW
FEB
8
{1}------------------------------------------------
#### IV. Device Description
CARESIDE™ Amylase cartridges are used with the CARESIDE Analyzer™ to measure amylase activity in whole blood, serum or plasma specimens. The CARESIDE™ Amylase cartridge, a single use disposable in vitro diagnostic test cartridge, delivers a measured volume of serum or plasma to a dry film to initiate the measurement of amylase activity. The film cartridge (patent pending) contains all reagents necessary to measure amylase activity.
#### Explanation of Device Function A.
Each CARESIDE™ Amylase cartridge consists of an amylase-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the specimen into the cartridge Sample Well, closes the lid and inserts the cartridge into the CARESIDE Analyzer™.
Once loaded, the CARESIDE Analyzer™ scans the cartridge barcode, brings the cartridge and the contained specimen to 37℃, and spins the cartridge to move the sample from the Sample Well into the cartridge channels and chambers. 8.5 microliters of sample remains in the metering passage. Any excess sample flows into an overflow well.
The 8.5 microliters of sample is automatically dispensed onto the multi-layer reagent film. The spreading and substrate layer distributes the specimen uniformly. Amylase p-nitrophenyl-α-D-maltopentaoside to p-nitrophenyl-oligosaccharides converts (PNP-Gm) and free oligosaccharides (Gn). The sample mixture diffuses into the underlying detection layer where PNP-Gm is converted by a-glucosidase to generate Gm and p-nitrophenol which converts to a vellow dye. The color intensity of the resulting yellow dye, as measured by the amount of reflected light at 425 nanometers, directly relates to the amylase activity of the specimen.
Test Reaction Sequence:
amylase > PNP-Gm + Gn (m+n=5) PNP-Gs a-glucosidase > PNP + Gm PNP-Gm -
As the cartridges spin. a photodiode measures reflectance of light emitted by a wavelength-specific light emitting diode (LED) over a fixed time period. The analyzer uses the reflectance measurements and the lot-specific standard curve to calculate amylase activity.
#### B. Test Summary
Amylase is an enzyme produced by the pancreas and parotid glands that occurs in blood. Amylase catalyzes the hydrolysis of starch and related polysaccharide to yield maltose and other oligosaccharides. Amylase activity in the blood is elevated in inflammation and hemorrhage of the pancreas and is decreased in pancreatic insufficiency. Amylase measurement is useful in the differential diagnosis of the digestive system. Various other causes of hyperamylasemia include salivary lesions, biliary tract disorders, trauma, neoplastic diseases, ruptured ectopic pregnancy, pulmonary disease, and alcoholic intoxication. Certain drugs can also increase amylase activity in the blood in vivo.
#### V. Intended Use
#### A. Intended Use
The CARESIDE™ Amylase cartridge is intended for in vitro diagnostic use in conjunction with the CARESIDE Analyzer™ to quantitatively measure amylase activity in whole blood, serum or plasma.
- B. Indications for Use
This product is indicated for use in the diagnosis and treatment of patients with disease of the pancreas, salivary gland, and kidney.
{2}------------------------------------------------
#### VI. Technological Characteristics
- A. Similarities
| | CARESIDE™ Amylase | Vitros AMYL DT Slides |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Intended Use | Primarily to aid in the diagnosis and treatment of patients with disease of the pancreas, salivary gland, and kidney.<br>For <i>in vitro</i> diagnostic use. | Primarily to aid in the diagnosis of diseases of the pancreas<br>For <i>in vitro</i> diagnostic use |
| Indications | For professional laboratory:<br>not for point of care or physician office laboratory use. | |
| Measurement | Quantitative | Same |
| Method Principle | Dry film, p-nitrophenyl-α-D-maltopentaoside substrate reaction coupled with α-glucosidase | Dry film conversion by amylase of high molecular weight dyed starch into low molecular weight dyed saccharides |
| Specimen dilution | Not required | Same |
| Materials | p-nitrophenyl-α-<br>D-maltopentaoside,<br>α-glucosidase | Dyed amylopectin |
| Detector | Reflectance (425 nm) | Reflectance (555 nm) |
| Test time | Approx. 4 minute warm-up (on-board) plus approximately 6 minute test time. | 15 minutes slide warm-up (off-line) plus 5 minutes test time. |
| Reference Method | Rate determination using ethylidene-pNP-G7 substrate, spectrophotometric detection, 37 °C | PNP Maltopentaoside Method, 37 °C |
| Sample Type | Anti-coagulated whole blood, serum, or heparinized plasma. | Serum or heparinized plasma |
| Specimen volume | 8.5 µl test volume<br>(85 ± 15 µl applied volume) | 10 µl |
| Calibration | Calibration information bar-coded on each cartridge.<br>Calibration information may change with each lot. | Run Vitros DT II calibrators whenever a new slide lot is used or when necessary. |
| Quality Control | 2 levels | Same |
| Reporting Units | U/L | Same |
| Reaction Temp. | 37 °C | Same |
#### B. Differences
| | CARESIDE™ Amylase | Vitros AMYL DT Slides |
|-------------------------|-------------------|-----------------------|
| Specimen<br>Processing | Not required | Required |
| Accurate<br>pipetting | Not required | Required |
| Reagent pre-<br>warming | Not required | Required |
-
-
.
{3}------------------------------------------------
-
-
#### Comparative Performance Characteristics C.
| | CARESIDE™ Amylase | Vitros AMYL DT Slides |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Detection limit | 15 U/L | 5 U/L |
| Reportable range | 15 to 600 U/L | 5 to 900 U/L |
| Accuracy | Mean recovery 95% | Not provided |
| Precision | Total CV, 75 U/L, 13% | Total CV, 51 U/L, 13% |
| Method<br>comparison | CARESIDE™ = 0.94 (Trace Sci. Amylase DST) + 11.5 U/L, r = 0.97<br>CARESIDE™ = 0.76 (Vitros AMYL DT) + 14.6 U/L, r = 0.95 | |
| Linearity | Linearity yielded slope and<br>correlation coefficient within<br>acceptable limits. | Not provided |
| Interference | No significant interference observed at tested concentration of interferent:<br>Ascorbic Acid,........20 mg/dL<br>Hemoglobin,......... 100 mg/dL<br>Triglycerides ..... 1500 mg/dL | No reported interference |
#### D. Conclusion
The nonclinical and clinical data provided demonstrate that the CARESIDE™ Amylase product is as safe, effective, and performs as well as or better than the legally marketed predicate device
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the text "Public Health Service". The text is written in a simple, sans-serif font. The words are arranged horizontally, with "Public" and "Health" on the first line and "Service" on the second line.
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized design of three overlapping human figures.
FEB 8 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Kenneth B. Asarch, Pharm. D., Ph.D. Vice President, Quality Systems/ Regulatory Affairs Careside Inc. 6100 Bristol Parkway Culver City, California 90230
Re: K990025 Trade Name: CARESIDE™ Amylase Regulatory Class: II Product Code: JFJ Dated: December 30, 1998 Received: January 5, 1999
Dear Dr. Asarch:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
CARESIDE, Inc. Page 15
# VII. INDICATIONS FOR USE
510(k) Number:
K 990025
Device Name:
CARESIDE™ Amylase
Indications for use:
For in vitro diagnostic use with the CARESIDE Analyzer™ to measure amylase from whole blood, serum or plasma specimens to aid in the diagnosis and treatment of patients with disease of the pancreas, salivary gland, and kidney.
Alan Cooper
orv Devices
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.