← Product Code [MOS](/productcode/MOS) · K990007

# OPART, MODEL MRT-600 (K990007)

_Toshiba America Mri, Inc. · MOS · Apr 1, 1999 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K990007

## Device Facts

- **Applicant:** Toshiba America Mri, Inc.
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Apr 1, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Diagnostic imaging of the human body (including head, abdomen, breast, heart, pelvis, spine, blood vessels, limbs, and extremities), fluid visualization, 2D and 3D imaging, MR angiography and MR fluoroscopy.

## Device Story

Large and Medium Flexible QD Body coils are quadrature (QD) RF coil accessories for the OPART MRI system. Device utilizes two orthogonal RF traces (saddle and solenoid) to capture NMR signals, which are amplified and summed to improve signal-to-noise ratio. Flexible design allows placement directly on patient, eliminating need for patient pallet to pass through coil, enabling smaller coil size. Operated by trained MRI technicians/physicians in clinical settings. Output is raw RF signal processed by the MRI system into diagnostic images. Benefits include improved signal-to-noise ratio and patient positioning flexibility.

## Clinical Evidence

Bench testing only. Sample phantom images and clinical images provided to demonstrate imaging performance.

## Technological Characteristics

Quadrature (QD) RF coil; flexible form factor; variable tune and variable match circuitry; 50-ohm signal lines; operates at 15MHz; static field strength 0.35T; SAR <0.4 W/kg; acoustic noise 98.4 dB(A).

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Matrix 3000 Flexible Spine Coil ([K964753](/device/K964753.md))

## Reference Devices

- OPART diagnostic resonance system ([K962933](/device/K962933.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

000000 241 1000 1

Kcj9007

## 510(k) Premarket Notification

# SUMMARY OF SAFETY AND EFFECTIVENESS

| 1. | DEVICE NAME:                 | Magnetic Resonance Diagnostic Device Accessory |  |  |
|----|------------------------------|------------------------------------------------|--|--|
|    | Model Number:                | MRT-600                                        |  |  |
|    | Trade/Proprietary Name:      | OPART™                                         |  |  |
| 2. | ESTABLISHMENT REGISTRATION:  | 2636923                                        |  |  |
| 3. | U.S. AGENT NAME AND ADDRESS: | Toshiba America MRI, Inc.<br>280 Utah Avenue   |  |  |
|    |                              | South San Francisco, CA 94080                  |  |  |
|    | CONTACT PERSON:              | Ken Nehmer                                     |  |  |
|    |                              | (650)872-2722 ext. 6083                        |  |  |

- MANUFACTURING SITE: USA Instruments, Inc. 4. 675-B Alpha Drive Highland Heights, Ohio 44143, USA
Establishment Registration: 1529041

- DATE OF SUBMISSION: December 31, 1998 5.
- DEVICE DESCRIPTION: The Large and Medium Flexible QD Body coils are essentially a QD 6. (quadrature) extension and a flexible version of the standard Body Coil. The extension consists of adding a saddle trace to the existing solenoid trace. The two independent RF traces have RF magnetic fields which are orthogonal (oriented at 90 degrees with respect to each other) to create a quadrature coil. The NMR signals from the two independent loops are sent to an RF front end where they are then amplified and combined (summed) to provide a resultant signal with improved signal-to-noise ratio.

With the coil's flexibility. RF connectors are added which enable the coil to open for patient loading. This enables the coil to be placed directly on the patient pallet. The standard body coil requires the patient to lay on the patient pallet then the patient with pallet is slid through the body coil. An advantage of placing the patient directly on the coil is that the pallet does not go through the coil. Hence the pallet is not filling the coil. This enables the coil to be physically smaller, for improved signal to noise, without losing space for patient loading.

{1}------------------------------------------------

7.

## SAFETY PARAMETERS:

| Maximum static field strength:                  | 0.35 Tesla   |
|-------------------------------------------------|--------------|
| Rate of change of magnetic field:               | 19T/second   |
| Maximum radio frequency power deposition (SAR): | <0.4 Watt/kg |
| Acoustic noise levels (maximum):                | 98.4 dB (A)  |

## IMAGING PERFORMANCE PARAMETERS: 8.

| Specification volume: | Head: | 10cm dsv |
|-----------------------|-------|----------|
|                       | Body: | 20cm dsv |

Sample phantom images and clinical images are presented from both the Large Flexible QD Body coil (Appendix D & E) and the Medium Flexible QD Body coil (Add to file appendix 2 and 3).

### INTENDED USE 9.

| Anatomical regions: | Head, body, extremity, spine, neck, TMJ, breast, and heart                                                                                                                                                               |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nuclei excited:     | Hydrogen                                                                                                                                                                                                                 |
| Diagnostic use:     | Diagnostic imaging of the human body (including head, abdomen,<br>breast, heart, pelvis, spine, blood vessels, limbs, and extremities),<br>fluid visualization, 2D and 3D imaging, MR angiography and MR<br>fluoroscopy. |

### 10. EQUIVALENCY INFORMATION:

Toshiba America MRI, Inc., believes that both the Large and Medium Flexible QD Body Coil options for OPART™ system are substantially equivalent to the Matrix 3000 Flexible Spine Coil (K964753). The Large and Medium Flexible QD Body Coils are manufactured by USA Instruments, Inc. and is a modified version of USA Instruments Matrix 3000 Flexible Spine Coil.

The Large and Medium Flexible QD Body Coils are identical to the Matrix 3000 Flexible Spine Coil with some minor modifications added to allow for use with the OPART™ system. The modifications include changing capacitor values in order to shift the frequency to that used for the OPART™ imaging system (15Mhz). Additional tuning varactors were added for both tuning and matching. The original (Matrix 3000) design is variable tune with a fixed match scheme. For OPART™, the Large and Medium Flexible QD Body coils have variable tune and variable match. The interfacing cable was modified to accommodate more control lines for the varactors and coil ID. The signal lines of the coax portion of the cable is 50 ohms for both the Matrix 3000 and Large Flexible and Medium QD Body coils. Mechanically, both coils are identical. The modifications added to the coil do not raise new questions of safety or efficacy.

This optional coil does not introduce any new indications for use from those cleared in the Premarket Notification for OPART™ diagnostic resonance system 510(k) number K962933.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized symbol resembling three overlapping human profiles or abstract shapes, positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

1 1999 APR

Ken Nehmer Quality Engineer Toshiba America MRI, Inc. 280 Utah Avenue South San Francisco, CA 94080 Re:

K990007 Large and Medium Flexible OD Body Coil for Opart MRI System Dated: December 31, 1998 Received: January 4, 1999 21 CFR 892.1000/Procode: 90 MOS

Dear Mr. Nehmer:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marked predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Madical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

- Sincerely yours,

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page __ 1__ of __ 1__

了 ...

| 510(k) Number (if known): | K990007                                                  |
|---------------------------|----------------------------------------------------------|
| Device Name:              | OPART™ (MRT-600) Large and Medium Flexible QD Body Coils |
| Indications for Use:      |                                                          |
| Imaging of:               |                                                          |

The Whole Body (including head, abdomen, breast, pelvis, limbs and extremities, spine, neck, TMJ, heart, blood vessels). [Application terms include MRCP (MR Cholangiopancreatography), MR Urography, MR Myelography, MR Fluoroscopy, SAS (Surface Anatomy Scan), Dynamic Scan and Cine Imaging.]

Fluid Visualization

- Fluid Visualization -
- 2D/3D Imaging .

- MR Angiography/MR Vascular Imaging -
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | <div style="display:inline-block;">✓</div> |
|----------------------|--------------------------------------------|
| (Per 21 CFR§801.109) |                                            |

OR

| Over-The-Counter Use     |  |
|--------------------------|--|
| (Optional Format 1-2-96) |  |

| (Division Sign-Off)                                                |  |
|--------------------------------------------------------------------|--|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices |  |

| 510(k) Number | K990007 |
|---------------|---------|
|---------------|---------|

CONFIDENTIAL

---

**Source:** [https://fda-staging.innolitics.com/device/K990007](https://fda-staging.innolitics.com/device/K990007)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
